Cambruzzi M, Macfarlane P. Variation in syringes and needles dead space compared to the International Organization for Standardization standard 7886-1:2018.
Vet Anaesth Analg 2021;
48:532-536. [PMID:
34059461 DOI:
10.1016/j.vaa.2021.01.008]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/09/2020] [Revised: 11/17/2020] [Accepted: 01/09/2021] [Indexed: 10/21/2022]
Abstract
OBJECTIVE
To measure the dead space of various syringe volumes and brands and a range of needles gauges commonly used in clinical practice, and to compare the results to the International Organization for Standardization (ISO) standard 7886-1:2018.
STUDY DESIGN
Prospective observational study.
METHODS
Syringes of five brands and seven volumes: 1, 2.5, 5, 10, 20, 30, and 60 mL, 10 of each, 10 1 mL low dead space syringes and 10 23, 21 and 18 gauge needles were analysed. Syringe dead space was estimated by weighing each syringe, drawing up and expelling its rated volume of water for injection and re-weighing the syringe. The difference in mass between the two measurements was calculated and converted to a volume based on the density of water. The dead space of the needles was estimated using a similar technique: each needle was attached to a 1 mL syringe of known dead space. A Mettler electronic balance was used for the measurements. Mean dead space was calculated for each syringe volume and needle gauge. Data were compared to the ISO standard.
RESULTS
Syringe dead space for 1 and 2.5 mL was less than 0.07 mL. For 5 mL syringes four of five brands, and for 10 mL syringes one of five brands failed to comply with the ISO; the dead space was greater than 0.075 and 0.1 mL, respectively. For the 20, 30, 60 mL syringes the dead space was less than 0.2 mL. Needle dead space was 0.05 mL for 23 and 21 gauge, and 0.07 mL for 18 gauge, similar in order of magnitude to syringe dead space.
CONCLUSIONS
The dead space of syringes differs between brands, and some do not meet the ISO standard. When calculating the amount of drug lost due to dead space, both that of the syringe and needle used should be considered.