CT fluoroscopy-guided placement of inferior vena cava filters: feasibility study in pigs.
J Vasc Interv Radiol 2011;
22:1531-4. [PMID:
21856172 DOI:
10.1016/j.jvir.2011.07.007]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/06/2011] [Revised: 07/14/2011] [Accepted: 07/15/2011] [Indexed: 11/20/2022] Open
Abstract
PURPOSE
To evaluate the feasibility of computed tomography (CT)-guided placement of inferior vena cava (IVC) filters in a swine model.
MATERIALS AND METHODS
Five domestic pigs (60-70 kg) underwent transfemoral and transjugular IVC filter placement under real-time CT fluoroscopic guidance. Filter position was confirmed by contrast-enhanced CT and digital subtraction angiography. Filter tilt, distance to target position, and fluoroscopy time were analyzed.
RESULTS
A total of 10 filters were successfully implanted (five via transfemoral approach, five via transjugular approach) without complications. The mean distance to the target position was 0.3 cm ± 0.2. Mean filter tilt was 3.2° ± 2.3 (range, 0°-7°), without differences between deployment techniques (P = .8486). Average fluoroscopy time was 25.9 s ± 6.9 per procedure.
CONCLUSIONS
CT fluoroscopy-guided placement of IVC filters is safely feasible. Use of this technique may avoid the need to move critically ill patients.