Cederberg KLJ, Jeng B, Sasaki JE, Lai B, Bamman M, Motl RW. Accuracy and precision of wrist-worn actigraphy for measuring steps taken during over-ground and treadmill walking in adults with Parkinson's disease.
Parkinsonism Relat Disord 2021;
88:102-107. [PMID:
34171566 DOI:
10.1016/j.parkreldis.2021.06.009]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/31/2021] [Revised: 06/14/2021] [Accepted: 06/15/2021] [Indexed: 02/08/2023]
Abstract
INTRODUCTION
Step counts represent a straight-forward method of measuring physical activity in adults with Parkinson's disease (PD). The present study examined the absolute and relative accuracy and precision of a wrist-worn research-grade accelerometer (i.e., ActiGraph GT3X+) for measuring step counts during over-ground and treadmill walking in adults with PD and controls without PD.
METHODS
Participants (PD: n = 29; controls: n = 31) wore two ActiGraph GT3X + accelerometers, one on each wrist, and completed an over-ground walking bout followed by a treadmill walking bout at the same speed. Step counts were measured manually using a hand-held tally counter. Accuracy and precision were based on absolute and relative metrics.
RESULTS
The ActiGraph GT3X + underestimated step counts in both participants with PD (4.7-11% error) and controls without PD (8.8-17% error), with a greater discrepancy in controls. The ActiGraph GT3X + provided more accurate and precise estimates of step counts when placed on the more affected wrist and non-dominant wrist for participants with PD and controls, respectively, and was more accurate and precise during over-ground walking compared with treadmill walking for both groups.
CONCLUSIONS
Our results suggest that placement of the device (i.e., dominant vs. non-dominant), type of activity (i.e., over-ground vs. treadmill walking), and presence of clinical conditions may impact the accuracy and precision of data when using the research-grade ActiGraph GT3X + accelerometer for measuring step counts.
Collapse