Sutton DW, Mead EH, Schmid-Schönbein GW. A high precision dual feedback pump for unsteady perfusion of small organs.
Ann Biomed Eng 1989;
17:269-78. [PMID:
2735584 DOI:
10.1007/bf02368046]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/02/2023]
Abstract
A dynamic pump system is described for perfusion of small organs with whole blood. The pump system was designed with the following aims: Very low flowrates to perfuse single organs in small rodents; high dynamic response for pressure or flow to permit experimenting with a harmonic signal at frequencies up to 20 Hz or by way of sharp step transients in less than 10 msec; high precision to allow detection of fine physiological details, and minimum blood cell trauma or cell activation by use of a piston principle. Representative pressure-flow curves are shown for the rat gracilis muscle after vasodilation. The curves are highly reproducible and serve as a complimentary dataset for microvascular observations in the same organ.
Collapse