Ricci C, Velimirovic BM, Fitzgerald TN. Case report of migration of 2 ventriculoperitoneal shunt catheters to the scrotum: Use of an inguinal incision for retrieval, diagnostic laparoscopy and hernia repair.
Int J Surg Case Rep 2016;
29:219-222. [PMID:
27883967 PMCID:
PMC5122702 DOI:
10.1016/j.ijscr.2016.11.002]
[Citation(s) in RCA: 11] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/21/2016] [Revised: 10/31/2016] [Accepted: 11/01/2016] [Indexed: 11/27/2022] Open
Abstract
A ventriculoperitoneal shunt catheter migrated through an inguinal hernia into the scrotum.
A single incision was used to for diagnostic laparoscopy, catheter removal and hernia repair.
Prompt surgical removal of fractured shunt catheters is recommended to prevent organ injury.
Backgroud
Ventriculoperitoneal shunts are commonly used in the treatment of hydrocephalus, and catheter migration to various body sites has been reported. Pediatric and general surgeons are asked on occasion to assist with intraabdominal access for these shunts, particularly when there may be extensive adhesions or other complicating factors.
Methods
We describe a case in which an old shunt catheter was never removed from the abdomen, and it migrated through an inguinal hernia into the scrotum. The catheter became entangled and fibrosed to the testicle. A second and more recent shunt catheter was also in the scrotum. A single incision in the inguinal region was used to remove both shunt catheters, repair the inguinal hernia and perform diagnostic laparoscopy to assist in placing a new ventriculoperitoneal shunt.
Results
Prompt surgical removal is recommended for catheters remaining in the abdomen after ventriculoperitoneal shunt malfunction. These catheters may cause injury to the testicle, or possibly other intraabdominal organs. General or pediatric surgical consultation should be obtained for lost catheters or inguinal hernias.
Conclusion
In the case of an inguinal hernia containing a fractured shunt catheter, the hernia sac can be used to remove the catheter, repair the hernia and gain laparoscopic access to the abdomen to assist with shunt placement.