Manchanda V, Siddiqui O, Meena K, Sharma A, Saxena S. Quality assurance of SARS-CoV-2 testing laboratories during the pandemic period in India - An experience from a designated provider laboratory.
Indian J Med Microbiol 2024;
49:100577. [PMID:
38588879 DOI:
10.1016/j.ijmmb.2024.100577]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/28/2023] [Revised: 04/01/2024] [Accepted: 04/01/2024] [Indexed: 04/10/2024]
Abstract
PURPOSE
Indian Council of Medical Research (ICMR) initiated an Inter-Laboratory Quality Control testing (ILQC) program for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) testing. Under this program, SARS-CoV-2 testing laboratories across the country submit specimens to the assigned State Quality Control (SQCs) laboratories for ILQC testing. This study aimed to investigate the performance of public and private SARS-CoV-2 testing laboratories in Delhi and highlights the country's effort in ramping up testing facility with close monitoring of the quality of Covid-19 testing results.
METHODS
In the present study, two-years of SARS-CoV-2 testing data is included. During July 2020 through February 2022, a total of 1791 anonymised specimens were received from 56 public and private laboratories. These specimens were processed by reverse transcriptase - polymerase chain reaction (RT-PCR) tests as per National Institute of Virology (NIV) protocol and the results were uploaded on the ICMR quality control/quality assurance (QC/QA) portal without directly conveying the results to respective participating laboratories. This portal generated a final report stating concordance and intimate results to individual laboratories.
RESULTS
Among the 1791 specimens, 25 were rejected and the remaining 1766 were tested. Among these specimens 1691 (95.75%) revealed concordance, and 75 (4.24%) were discordant. A total of 29 laboratories had 100% concordance, 21 laboratories had over 90% concordance and six laboratories had over 80% concordance.
CONCLUSIONS
The study demonstrates that the establishment of an inter-laboratory comparison program for SARS-CoV-2 testing helped in monitoring quality of SARS-CoV-2 testing in the country.