Flanagan JPM, Udovenya WHF, Astrahan MA, McKay D, Phillips C, McKenzie JD, O’Day R, Fog LS. Quantifying the effect of eccentric ruthenium plaque placement on tumor volume dose.
J Contemp Brachytherapy 2023;
15:442-447. [PMID:
38230400 PMCID:
PMC10789163 DOI:
10.5114/jcb.2023.133614]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/30/2023] [Accepted: 10/30/2023] [Indexed: 01/18/2024] Open
Abstract
Purpose
Ruthenium-106 brachytherapy is a common treatment for small to medium-sized uveal melanomas. In certain clinical contexts, plaques may be placed eccentrically to tumor center. The effect of plaque decentration, a common radiation dose measurement in radiotherapy: D98%, the percentage of the tumor volume receiving at least 98% of the prescribed dose (a commonly used term in radiation oncology), is unknown. We investigated this using two commonly used plaques (CCA and CCB; Eckert & Ziegler, BEBIG GmbH) in silico.
Material and methods
Using a Plaque Simulator™ (Eye Physics) plaque modelling software, treatment time required to deliver 100 Gy D98% with central plaque placement was calculated for both plaque models, treating tumors with basal dimensions of 10 mm (CCB plaque only) and 7 mm (CCA and CCB plaques), and a range of thicknesses. D98% was calculated for plaque-tumor edge distances of 0-5 mm. Additionally, we defined minimum plaque-tumor edge distances, at which D98% fell by 10% and 5% (safety margins).
Results
D98% decreased as plaque-tumor edge distance decreased, i.e. as plaque eccentricity increased. Minor (< 1 mm) plaque decentration caused minimal D98% changes across tumor thicknesses. Safety margins did not follow a consistent pattern.
Conclusions
Eccentric plaque placement reduces the radiation dose delivered to choroidal tumors. Both tumor (thickness, diameter) and plaque (size, location) characteristics are important D98% modulators. Further investigation of the effect of these characteristics and dose to organs at risk is essential.