1
|
Kim JH, Sadri V, Chen H, Bhat S, Kohli K, Makkar R, Babaliaros VC, Sharma RP, Yoganathan AP. Effect of Ascending Aortic Curvature on Flow in the Sinus and Neo-sinus Following TAVR: A Patient-Specific Study. Ann Biomed Eng 2024; 52:425-439. [PMID: 37922056 DOI: 10.1007/s10439-023-03392-x] [Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/04/2023] [Accepted: 10/23/2023] [Indexed: 11/05/2023]
Abstract
Patient-specific aortic geometry and its influence on the flow in the vicinity of Transcatheter Aortic Valve (TAV) has been highlighted in numerous studies using both in silico and in vitro experiments. However, there has not yet been a detailed Particle Image Velocimetry (PIV) experiment conducted to quantify the relationship between the geometry, flow downstream of TAV, and the flow in the sinus and the neo-sinus. We tested six different patient-specific aorta models with a 26-mm SAPIEN 3 valve (Edwards Lifesciences, Irvine, CA, USA) in a left heart simulator with coronary flow. Velocities in all three cusps and circulation downstream of TAV were computed to evaluate the influence of the ascending aorta curvature on the flow field. The in vitro analysis showed that the patient-specific aortic curvature had positive correlation to the circulation in the ascending aorta (p = 0.036) and circulation had negative correlation to the particle washout time in the cusps (p = 0.011). These results showed that distinct vortical flow patterns in the ascending aorta as the main jet impinges on the aortic wall causes a recirculation region that facilitates the flow back into the sinus and the neo-sinus, thus reducing the risk of flow stagnation and washout time.
Collapse
Affiliation(s)
- Jae Hyun Kim
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA
| | - Vahid Sadri
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA
| | - Huang Chen
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA
| | - Sanchita Bhat
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA
| | - Keshav Kohli
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA
| | - Raj Makkar
- Cedars-Sinai Medical Center, Smidt Heart Institute, Los Angeles, CA, USA
| | | | - Rahul P Sharma
- Division of Cardiovascular Medicine, Stanford University, Stanford, CA, USA
| | - Ajit P Yoganathan
- Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology and Emory University, Atlanta, GA, USA.
| |
Collapse
|
2
|
Gutierrez MA. Computational Analysis of Fluid Dynamics in the Transcatheter Aortic Valve Replacement. Arq Bras Cardiol 2020; 115:688-689. [PMID: 33111870 PMCID: PMC8386972 DOI: 10.36660/abc.20201002] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022] Open
Affiliation(s)
- Marco A Gutierrez
- Instituto do Coração (InCor), Faculdade de Medicina da Universidade de São Paulo (FMUSP), São Paulo, SP - Brasil
| |
Collapse
|