Barrow AE, Pilia M, Guda T, Kadrmas WR, Burns TC. Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen?
Am J Sports Med 2014;
42:343-9. [PMID:
24158183 DOI:
10.1177/0363546513507769]
[Citation(s) in RCA: 116] [Impact Index Per Article: 11.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 01/31/2023]
Abstract
BACKGROUND
Cortical suspension devices are commonly used for femoral graft fixation during anterior cruciate ligament (ACL) reconstructive surgery. Adjustable-length fixation devices provide technical advantages over fixed-length loops but may be more susceptible to lengthening during cyclic loading.
HYPOTHESIS
Both fixed-length and adjustable-length femoral cortical suspension devices would withstand ultimate loads greater than those normally experienced by the native ACL and would prevent clinically significant lengthening during prolonged cyclic loading.
STUDY DESIGN
Controlled laboratory study.
METHODS
Mechanical testing was performed on 3 ACL graft cortical suspensory devices by use of an extended cyclic loading (4500 cycles at 10-250 N) and pull-to-failure protocol. Two adjustable-length devices were additionally tested with the free suture ends tied.
RESULTS
Total displacement after 4500 cycles of tensioning at variable loads (expressed as mean ± SD) was 42.45 mm (±7.01 mm) for the Arthrex TightRope RT, 5.76 mm (±0.35 mm) for the Biomet ToggleLoc, and 1.34 mm (±0.03 mm) for the Smith & Nephew EndoButton CL Ultra (P < .001). The Arthrex TightRope reached clinical failure of 3 mm lengthening after fewer cycles (1349 ± 316) than the Biomet ToggleLoc (2576 ± 73) (P < .001). The Smith & Nephew EndoButton did not reach clinical failure during cyclic testing. With the free suture ends tied, after 4500 cycles, the Arthrex TightRope had a significant decrease in lengthening to 13.36 ± 1.86 mm (P < .037) There was also a significant difference in ultimate load between the TightRope (809.11 ± 52.94 N) and the other 2 constructs (P < .001).
CONCLUSION
The ultimate load of all graft-fixation devices exceeded the forces likely to be experienced in a patient's knee during the early postoperative rehabilitation period. However, the adjustable-length fixation devices experienced a clinically significant increase in loop lengthening during cyclic testing. This lengthening is partially caused by suture slippage into the adjustable-length loop.
CLINICAL RELEVANCE
Adjustable-length ACL graft cortical suspension devices lengthen under cyclic loads because free suture ends are pulled into the adjustable loop. This may allow for graft-fixation device lengthening during the acute postoperative period.