Kim JS, Kang MJ, Jeong KD, Hwang JH. Haptic fracture and dislocation of polyimide intraocular lens after neodymium: YAG laser capsulotomy: A case report.
Medicine (Baltimore) 2019;
98:e15720. [PMID:
31096526 PMCID:
PMC6531260 DOI:
10.1097/md.0000000000015720]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/03/2022] Open
Abstract
RATIONALE
We report a case of haptic fracture of polyimide intraocular lens (IOL) and its dislocation after neodymium:YAG (Nd:YAG) laser capsulotomy. To the best of our knowledge, this is the first report of both the fracture and dislocation of the IOL after laser posterior capsulotomy.
PATIENT CONCERNS
A 78-year-old male was referred to our clinic with a history of decreased visual acuity after Nd:YAG laser posterior capsulotomy in the right eye. A dislocated IOL was observed on slit lamp examination.
DIAGNOSIS
On slit-lamp examination, dislocated IOL with intact anterior capsulorrhexis margin was observed in the right eye.
INTERVENTIONS
We removed the dislocated IOL and fixed a 3-piece IOL (Acrysof MN60AC, Alcon Laboratories, Inc) in the ciliary sulcus under retrobulbar anesthesia.
OUTCOMES
After the operation, the observed uncorrected distance visual acuity (UCVA) was 20/20 and the measured intraocular pressure (IOP) was 18 mmHg; the IOL was fixed in the ciliary sulcus.
LESSONS
Three-piece, silicone-polyimide haptics in IOLs can get fractured after Nd:YAG laser capsulotomy. Thus, it is suggested that sufficient haptic dilation should be ensured while performing Nd:YAG laser capsulotomy in patients with polyimide haptics. Moreover, a small-sized capsulotomy is likely to be helpful in such patients.
Collapse