[Citation(s) in RCA: 8] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/10/2020] [Revised: 03/18/2021] [Accepted: 04/14/2021] [Indexed: 12/31/2022] Open
Abstract
PURPOSE/OBJECTIVES
This work seeks to evaluate the plan quality, treatment delivery efficiency, and accuracy of single-isocenter volumetric modulated arc therapy (VMAT) of abdominal/pelvic oligometastatic lymph nodes (LNs) stereotactic body radiation therapy (SBRT) on Halcyon Linac.
MATERIALS AND METHODS
After completing the in-house multitarget end-to-end phantom testing and independent dose verification using MD Anderson's single-isocenter/multi-target (lung and spine target inserts) thorax phantom, eight patients with two to three abdominal/pelvic oligometastatic LNs underwent highly conformal single-isocenter VMAT-SBRT treatment using the Halcyon Linac 6MV flattening filter free (FFF) beam. Targets were identified using an Axumin PET/CT scan co-registered with planning CT images and a single-isocenter was placed between/among the targets. Doses between 25 and 36.25 Gy in 5 fractions were delivered. Patients were treated every other day. Plans were calculated in Eclipse with advanced AcurosXB algorithm for heterogeneity corrections. For comparison, Halcyon VMAT-SBRT plans were retrospectively generated for SBRT-dedicated TrueBeam with a 6MV-FFF beam using identical planning geometry and objectives. Target coverage, conformity index (CI), dose to 2 cm away from each target (D2cm) and dose to adjacent organs-at-risk (OAR) were evaluated. Additionally, various treatment delivery parameters including beam-on time were recorded.
RESULTS
Phantom measurements showed acceptable spatial accuracy of conebeam CT-guided Halcyon SBRT treatments including compliance with MD Anderson's single-isocenter/multi-targets phantom credentialing results. For patients, the mean isocenter to tumor center distance was 3.4 ± 1.2 cm (range, 1.5-4.8 cm). The mean combined PTV was 18.9 ± 10.9 cc (range, 5.6-39.5 cc). There was no clinically significant difference in dose to LNs, CI, D2cm and maximal doses to OAR between single-isocenter Halcyon and Truebeam VMAT-SBRT plans, although, Halcyon plans provided preferably lower maximal dose to adjacent OAR. Additionally, total monitor units, beam-on time and overall treatment time was lower with Halcyon plans. Halcyon's portal dosimetry demonstrated a high pass rate of 98.1 ± 1.6% for clinical gamma passing criteria of 2%/2 mm.
CONCLUSION
SBRT treatment of abdominal/pelvic oligometastatic LNs with single-isocenter VMAT on Halcyon was dosimetrically equivalent to TrueBeam. Faster treatment delivery to oligometastatic LNs via single-isocenter Halcyon VMAT can improve clinic workflow and patient compliance, potentially reducing intrafraction motion errors for well-suited patients. Clinical follow-up of these patients is ongoing.