Purpose
To compare corneal flaps created in rabbits with a 60 kHz femtosecond (FS) laser using different levels of raster energy and to measure early inflammation, corneal stromal cell death, and late postoperative adhesion strength.
Methods
Sixty rabbits were divided into three groups of 20 each. A flap 110 µm thick and 9.0 mm in diameter was made in one eye of each rabbit at raster energies of 0.7 µJ, 1.1 µJ, and 2.4 µJ. Histopathological evaluation for inflammation and apoptosis using terminal deoxynucleotidyl transferase dUTP nick end labeling (TUNEL) staining was performed at 4 and 24 hours after flap creation. The adhesion strength of the flaps was measured with a tension meter at 1 and 3 months.
Results
Twenty four hours after flap creation, the 2.4 µJ group had more inflammatory and CD11b-positive cells than the 0.7 and 1.1 µJ groups. The number of TUNEL-positive cells increased with raster energy at 4 and 24 hours. The grams of force (gf) needed to detach the flaps at 3 months was significantly higher in 2.4 µJ group (170 gf) than in 0.7 µJ group (97.5 gf) and 1.1 µJ group (100 gf, p = 0.03).
Conclusions
Using raster energy lower than 1.1 µJ to make a flap with a 60 kHz FS laser decreases inflammatory cell infiltration and corneal stromal cell death in the central cornea but may result in a weaker flap than using higher raster energy (2.4 µJ).