In vivo validation of Diagnodent and Vista proof devices vs ICDAS clinical criteria on incipient carious lesions in adults.
Photodiagnosis Photodyn Ther 2021;
34:102252. [PMID:
33711531 DOI:
10.1016/j.pdpdt.2021.102252]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/09/2020] [Revised: 02/25/2021] [Accepted: 03/05/2021] [Indexed: 10/21/2022]
Abstract
AIM
This study aimed to validate the in vivo performance of Diagnodent and Vista proof devices with ICDAS clinical criteria on incipient carious lesions in adults.
METHODS
A total of 44 adult patients with 230 incipient occlusal caries took part in the present study. These patients were assessed for caries with ICDAS clinical criteria, and then they were examined with Diagnodent pen™ (DP) and Vista proof™ (VP) fluorescence devices. Sensitivity, specificity accuracy, and ICC agreement between devices with ICDAS criteria, which served as a gold standard, were evaluated.
RESULTS
Regarding the caries diagnostic devices, sensitivity and specificity found 0,61 and 0,51 for DP, and 0,64 and 0,54 for VP, respectively. The different detection methods showed no differences in diagnostic capacity (Az values) each other, and ICC values with ICDAS criteria were calculated low.
CONCLUSIONS
DP and VP do not contribute to incipient occlusal carious lesions' better detective ability compared with visual ICDAS clinical criteria. The DP and VP devices presented no differences in diagnostic ability and measured lesion depth concerning the visual examination.