In vitro human skin irritation test for evaluation of medical device extracts.
Toxicol In Vitro 2013;
27:2175-83. [PMID:
23999410 DOI:
10.1016/j.tiv.2013.08.006]
[Citation(s) in RCA: 30] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/13/2013] [Revised: 08/18/2013] [Accepted: 08/19/2013] [Indexed: 11/16/2022]
Abstract
The aim of this study was to determine if the EpiDerm™ reconstructed human skin model (MatTek Corp.) could be an acceptable alternative to the ISO 10993-required rabbit skin irritation test for assessing medical device biocompatibility. Eleven medical device polymers were tested. Four extracts were prepared per polymer, two each with saline and sesame oil; half were spiked with two R-38 irritants, lactic acid for saline extracts and heptanoic acid for the sesame oil extracts. Tissue viability was assessed by MTT reduction and the proinflammatory response was assessed by IL-1α release. LOAELs of 2% for lactic acid in saline and 0.7% for heptanoic acid in sesame oil were determined. A cell viability reduction of >50% was indicative of skin irritation. Cells exposed to saline extracts spiked with 3.25% lactic acid had significantly reduced mean cell viabilities (12.6-17.2%). Cells exposed to sesame oil extracts spiked with 1.25% heptanoic acid also exhibited reduced mean cell viabilities (25.5%-41.7%). All spiked cells released substantial amounts of IL-1α (253.5-387.4pg/ml) signifying a proinflammatory response. These results indicate that the EpiDerm™ model may be a suitable in vitro replacement for the assessment of the irritation potential of medical device extracts.