Alternative method for intramuscular fat analysis using common laboratory equipment.
Meat Sci 2015;
103:24-7. [PMID:
25591172 DOI:
10.1016/j.meatsci.2014.12.011]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/14/2014] [Revised: 11/11/2014] [Accepted: 12/31/2014] [Indexed: 10/24/2022]
Abstract
A procedure to quantify intramuscular fat was developed using common inexpensive laboratory equipment. Three homogenization methods of lyophilized muscle samples (Ball-mill, Grinder and Mortar) and two extraction methods (Ball-mill or Vortex) were used in turkey meat and pork. Two-hundred mg of lyophilized and homogenized samples were accurately weighed and mixed with 1.5 mL of dichloromethane-methanol (8:2) and shaken either in a Mixer Mill (MM400, Retsch Technology) or in a Vortex. The final mixture was separated by centrifugation. Solvent was evaporated under a nitrogen stream and lipid content was gravimetrically determined. Besides, it was checked that the fatty acid profile was not altered by the protocol used. Moreover, the analysis of 4 replicas from the same sample showed different variation coefficients (16-29%) for the new procedures proposed over a wide range of IMF content. The combination of Grinder and Vortex methodologies can be proposed as a simple and inexpensive alternative to previous ones.
Collapse