Analysis of implant stability changes in immediate loading using a laser displacement sensor in vivo and comparison of its sensitivity with that of resonance frequency analysis.
Clin Oral Implants Res 2021;
32:1341-1356. [PMID:
34403162 DOI:
10.1111/clr.13835]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/06/2021] [Revised: 07/07/2021] [Accepted: 08/05/2021] [Indexed: 11/30/2022]
Abstract
OBJECTIVE
The aim of this study was to analyze the stability changes in immediately loaded implants by using an in vivo quantitative measurement of micromotion under functional dynamic loading and to verify the sensitivity of Resonance Frequency Analysis (RFA) as compared to that of actual micromotion.
MATERIALS AND METHODS
The micromotions of immediately loaded implants placed in the tibia of 11 rabbits were monitored using a laser displacement sensor. Functional dynamic loading forces were applied 5 days a week for 6 weeks. The implant stability quotient (ISQ) was monitored using RFA.
RESULTS
The micromotion of the almost-loaded implants increased to peak values the day after loading was started and subsequently reached a plateau gradually. The ISQ changes in the loaded implants closely correlated with the alterations of the actual micromotion (r = -0.98, p < .01). Although the ISQ value itself correlated with the measured micromotion at the time of initial fixation (r = 0.73, p < .05), it did not correlate with the micromotion of the implant that acquired integration. No close correlation was observed between the ISQ and the histomorphometrical data.
CONCLUSION
The immediately loaded implants showed the lowest stability immediately after the start of loading, which gradually increased thereafter. RFA is considered a useful method for examining stability changes and initial stability; however, it cannot determine the absolute magnitude of the stability after integration.