Bioimpedance-Based Heart Failure Deterioration Prediction Using a Prototype Fluid Accumulation Vest-Mobile Phone Dyad: An Observational Study.
JMIR Cardio 2017;
1:e1. [PMID:
31758769 PMCID:
PMC6832026 DOI:
10.2196/cardio.6057]
[Citation(s) in RCA: 20] [Impact Index Per Article: 2.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/06/2016] [Revised: 12/19/2016] [Accepted: 01/28/2017] [Indexed: 12/21/2022] Open
Abstract
Background
Recurrent heart failure (HF) events are common in patients discharged after acute decompensated heart failure (ADHF). New patient-centered technologies are needed to aid in detecting HF decompensation. Transthoracic bioimpedance noninvasively measures pulmonary fluid retention.
Objective
The objectives of our study were to (1) determine whether transthoracic bioimpedance can be measured daily with a novel, noninvasive, wearable fluid accumulation vest (FAV) and transmitted using a mobile phone and (2) establish whether an automated algorithm analyzing daily thoracic bioimpedance values would predict recurrent HF events.
Methods
We prospectively enrolled patients admitted for ADHF. Participants were trained to use a FAV–mobile phone dyad and asked to transmit bioimpedance measurements for 45 consecutive days. We examined the performance of an algorithm analyzing changes in transthoracic bioimpedance as a predictor of HF events (HF readmission, diuretic uptitration) over a 75-day follow-up.
Results
We observed 64 HF events (18 HF readmissions and 46 diuretic uptitrations) in the 106 participants (67 years; 63.2%, 67/106, male; 48.1%, 51/106, with prior HF) who completed follow-up. History of HF was the only clinical or laboratory factor related to recurrent HF events (P=.04). Among study participants with sufficient FAV data (n=57), an algorithm analyzing thoracic bioimpedance showed 87% sensitivity (95% CI 82-92), 70% specificity (95% CI 68-72), and 72% accuracy (95% CI 70-74) for identifying recurrent HF events.
Conclusions
Patients discharged after ADHF can measure and transmit daily transthoracic bioimpedance using a FAV–mobile phone dyad. Algorithms analyzing thoracic bioimpedance may help identify patients at risk for recurrent HF events after hospital discharge.
Collapse