Kretschmar O, Sglimbea A, Corti R, Knirsch W. Shunt reduction with a fenestrated Amplatzer device.
Catheter Cardiovasc Interv 2011;
76:564-71. [PMID:
20882662 DOI:
10.1002/ccd.22556]
[Citation(s) in RCA: 25] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
BACKGROUND
In specific high-risk patients with congenital heart disease (CHD), a complete closure of an intracardiac defect/shunt is not possible for a variety of reasons. We report our experiences with an interventional approach for shunt-reduction using various modifications of a self-fabricated Amplatzer device in our institution.
METHODS
Retrospective analysis of patients with CHD having received an interventional partial shunt occlusion since 09/2005.
RESULTS
Five patients, mean age 18.6(3.4-66) years, mean weight 36.4(14-102) kg, have been treated. In three patients (3.4, 3.9, 66 years) with an atrial septal defect (ASD) and a restrictive left ventricle (LV) (n = 1) or pulmonary arterial hypertension (PAH) (n = 2), respectively, an Amplatzer Septal Occluder (ASO) with a predilated (n = 2) or a presutured (n = 1) central hole was implanted. After successful immediate volume release in all, the balloon-dilated holes closed spontaneously during mid-term follow-up, pulmonary artery (PA) pressure and LV function remained normal. Two patients (2.7 and 17 years) with a Fontan circulation and severe cyanosis (saturation ≤80%) due to a large fenestration and elevated PA pressures received a partial occlusion of their shunt by implanting a centrally stented ASO or Amplatzer Vascular plug. After a follow-up of 31 and 39 months both stents remained patent under oral anticoagulation, oxygen saturation remained >85% with PA pressures unchanged, and both patients were in good clinical conditions.
CONCLUSIONS
In patients with an ASD and significant PAH and/or restrictive LV physiology as well as in Fontan patients with a large surgically created fenestration but failing Fontan circulation, a partial closure with a self-fenestrated Amplatzer device can be a feasible and successful therapeutic option. Balloon-dilated fenestrations in the Amplatzer device tend to close spontaneously during follow-up. Nonresorbable sutures or stenting can ensure patency of the created holes.
Collapse