Callaud A, Metrard G, Bailly M. 3D-Ring CZT System With New Low- and Medium-Energy Range: Ultrafast Dual-Isotope Lung SPECT/CT Improvement.
Clin Nucl Med 2023;
48:716-718. [PMID:
37200461 PMCID:
PMC10317294 DOI:
10.1097/rlu.0000000000004706]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/13/2023] [Accepted: 04/01/2023] [Indexed: 05/20/2023]
Abstract
ABSTRACT
New 3D-ring CZT systems with low- and medium-energy-range detectors allow for simultaneous dual-isotope lung scintigraphy. We compared 10-, 7-, 5-, and 3-minute acquisitions for 99m Tc and 81m Kr performed simultaneously on StarGuide CZT-SPECT/CT and reframed in 50 patients. Ventilation/perfusion mismatches were calculated (mean, 15.6% ± 28%), and Spearman correlation coefficients of mismatches were 0.994, 0.994, and 0.984 between 10- and 7-, 5-, and 3-minute acquisitions, respectively. No visual difference in image quality or final diagnosis was found. 3D-ring CZT-SPECT with low and medium energy range detectors allows ultrafast dual-isotope lung scintigraphy up to 3 minutes.