1
|
Abdel-Sayed S, Abdel-Sayed P, Berdaj D, Ferrari E, Halbe M, von Segesser LK, Maisano F. Effect of blood viscosity on the performance of virtually wall-less venous cannulas. Perfusion 2019; 35:393-396. [PMID: 31696777 DOI: 10.1177/0267659119885514] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
AIM This study was designed to quantify the influence of blood as test medium compared to water in cannula bench performance assessment. METHODS An in vitro circuit was set-up with silicone tubing between two reservoirs. The test medium was pumped from the lower reservoir by centrifugal pump to the upper reservoir. The test-cannula was inserted in a silicone tube connected between the lower reservoir and the centrifugal pump. Flow rate and pump inlet-pressure were measured for wall-less versus thin-wall cannula using a centrifugal pump in a dynamic bench-test for an afterload of 40-60 mmHg using two media: blood 10 g/dL and 5.6 g/dL and water 0 g/dL. RESULTS The wall-less cannula showed significantly higher flows rates as compared to the thin-wall cannula (control), with both hemoglobin concentrations and water. Indeed, for a target volume of 200-250 mL of blood (Hg 10 g/dL) in the upper reservoir, the cannula outlet pressure (P) was -14 ± 14 mmHg versus -18 ± 11 mmHg for the wall-less and control respectively; the cannula outlet flow rate (Q) was 3.91 ± 0.41 versus 3.67 ± 0.45 L/min, respectively. At the same target volume but with a Hg of 5.7 g/dL, P was -16 ± 12 mmHg versus -19 ± 12 mmHg and Q was 4 ± 0.1 versus 4 ± 0.4 L/min for the wall-less cannula and control respectively. Likewise, P and Q values with water were -1 mmHg versus -0.67 ± 0.58 mmHg and 4.17 ± 0.45 L/min versus 4.08 ± 0.47 L/min for the wall-less and control respectively. CONCLUSION Walls-less cannula showed 5.6% less pump inlet-pressure differences calculated between blood and water, as compared to that of thin-wall cannula (-21 times). Flow differences were 6% and 10% for the walls-less and thin-wall cannula respectively. We conclude that testing the cannula performance with water is a good scenario and can overestimate the flow by a 10%. However, superiority for wall-less is preserved with both water and blood.
Collapse
Affiliation(s)
- Saad Abdel-Sayed
- Department of Surgery and Anesthesiology, CHUV, Lausanne, Switzerland
| | | | - Denis Berdaj
- Division of Cardiac Surgery, University Hospital of Basel, Basel, Switzerland
| | | | - Maximilian Halbe
- Department of Cardiovascular Surgery, University Hospital of Zurich, Zurich, Switzerland
| | | | - Francesco Maisano
- Department of Cardiovascular Surgery, University Hospital of Zurich, Zurich, Switzerland
| |
Collapse
|
2
|
Performance Increase in Venous Drainage for Mini-Invasive Heart Surgery: Superiority of Self-Expanding Cannulas. INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY 2019; 9:297-301. [DOI: 10.1097/imi.0000000000000083] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
Objective Originally, the Smartcanula principle (collapsed insertion and expansion in situ) was developed for venous drainage by gravity. However, in minimally invasive surgery, augmentation with either constrained force vortex pumps or vacuum is often used. The current study was set up to assess whether smaller diameters of self-expanding venous cannulas are sufficient in conjunction with venous drainage augmentation resulting in smaller access orifices. Methods To evaluate cannulas intended for cardiopulmonary bypass, an in vitro circuit was set up with silicone tubing between the test cannula encased in a lower reservoir, the centrifugal pump, and after an upper reservoir. Afterload was set arbitrarily at 60 mm Hg using a centrifugal pump. The pressure value was measured using Millar pressure transducers. Flow rate (Q) was measured using an ultrasonic flow meter calibrated with volume tank and timer. Revolutions per minute of the centrifugal pump were calibrated with a stroboscope. Data display and data recording were controlled using a Lab View application. Self-expanding (24F Smartcanula) and control (25F Biomedicus) cannulas were used. Results Sixty measurements were recorded. At pump speed of 1500, 1570, 2000, 2500, and 3000 rpm, the Q values were 3.6, 5.2, 6.6, 9.3, and 11.8 L/min for the 24F self-expanding cannula and 3, 4.3, 5.4, 7.5, and 9.3 L/min for the control cannula. The pressure values were 3.6, −5.4, −15.9, −45.3, and 80.6 mm Hg. Biomedicus 25F showed Q values from 16% to 19% less as compared with 24F Smartcanula. The pressure values were 6, 7, 4, 2, and 2 times more as compared with 24F Smartcanula. Conclusions Our experimental evaluation demonstrated the superior performance of the Smartcanula with its self-expanding design in comparison with the reference commercially available standard cannulas. The Smartcanula with its small diameter is particularly welcome for minimally invasive surgery.
Collapse
|
3
|
Şahin T, Tezer M, Cerit L. Hydrodynamic Evaluations of Four Mock Femoral Venous Cannulas. Braz J Cardiovasc Surg 2018; 33:435-442. [PMID: 30517250 PMCID: PMC6257525 DOI: 10.21470/1678-9741-2018-0036] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/12/2018] [Accepted: 05/09/2018] [Indexed: 11/21/2022] Open
Abstract
Objective To report the results of four mock femoral venous cannulas and the
hydrodynamical superiority of one of them, which is the completely punched
(CP) model, upon the other three. Methods Four simulated femoral venous cannulas (single-stage, two-stage, multi-stage,
and CP model) were designed from a 1/4” x 1/16” x 68 cm polyvinyl chloride
(PVC) tubing line for testing. Holes on the PVC tubes were opened by a 5 mm
aortic punch. In order to evaluate the cannulas' drainage performance,
gelofusine was used as fluid. The fluid was drained for 60 seconds by
gravitation and then measured for each model separately. Results Mean drained volumes of single-stage, two-stage, and multi-stage cannulas
were 2.483, 2.561, and 2.603 mL, respectively. However, the CP cannula
provided us a mean drained volume of 2.988 mL. There were significant
differences among the variables of the CP cannula and the other three mock
cannulas concerning the drained fluid flow (P<0.01). Conclusion In our study, the measured mean volumes showed us that more drainage surface
area provides better fluid drainage.
Collapse
Affiliation(s)
- Türker Şahin
- Department of Cardiovascular Surgery, Near East University, Nicosia, Cyprus
| | - Murat Tezer
- Faculty of Educational Science, Near East University, Nicosia, Cyprus
| | - Levent Cerit
- Department of Cardiology, Near East University, Nicosia, Cyprus
| |
Collapse
|
4
|
Goto T, Tanabe T, Inamura T, Shirota M, Fumoto K, Saito Y, Fukuda W, Fukuda I, Daitoku K, Minakawa M. Effect of inflow cannula side-hole number on drainage flow characteristics: flow dynamic analysis using numerical simulation. Perfusion 2018; 33:649-655. [DOI: 10.1177/0267659118782246] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
Background: Venous drainage in cardiopulmonary bypass is a very important factor for safe cardiac surgery. However, the ideal shape of venous drainage cannula has not been determined. In the present study, we evaluated the effect of side-hole number under fixed total area and venous drainage flow to elucidate the effect of increasing the side-hole numbers. Method: Computed simulation of venous drainage was performed. Cannulas were divided into six models: an end-hole model (EH) and models containing four (4SH), six (6SH), eight (8SH), 10 (10SH) or 12 side-holes (12SH). Total orifice area of the side-holes was fixed to 120 mm2 on each side-hole cannula. The end-hole orifice area was 36.3 mm2. The total area of the side-holes was kept constant when the number of side-holes was increased. Result: The mean venous drainage flow rate of the EH, 4SH, 6SH, 8SH, 10SH and 12SH was 2.57, 2.52, 2.51, 2.50, 2.49, 2.41 L/min, respectively. The mean flow rate decreased in accordance with the increased number of side-holes. Conclusion: We speculate that flow separation at the most proximal site of the side-hole induces stagnation of flow and induces energy loss. This flow separation may hamper the main stream from the end-hole inlet, which is most effective with low shear stress. The EH cannula was associated with the best flow rate and flow profile. However, by increasing side-hole numbers, flow separation occurs on each side-hole, resulting in more energy loss than the EH cannula and flow rate reduction.
Collapse
Affiliation(s)
- Takeshi Goto
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| | - Tsubasa Tanabe
- Department of Intelligent Machines and System Engineering, Faculty of Science and Technology, Hirosaki University, Hirosaki, Japan
| | - Takao Inamura
- Department of Intelligent Machines and System Engineering, Faculty of Science and Technology, Hirosaki University, Hirosaki, Japan
| | - Minori Shirota
- Department of Intelligent Machines and System Engineering, Faculty of Science and Technology, Hirosaki University, Hirosaki, Japan
| | - Koji Fumoto
- Department of Mechanical Engineering, College of Science and Engineering, Aoyama Gakuin University, Kanagawa, Japan
| | - Yoshiaki Saito
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| | - Wakako Fukuda
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| | - Ikuo Fukuda
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| | - Kazuyuki Daitoku
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| | - Masahito Minakawa
- Department of Thoracic and Cardiovascular Surgery, Hirosaki University Graduate School of Medicine, Hirosaki, Japan
| |
Collapse
|
5
|
Chin JH, Lee EH, Kim JI, Choi IC. Prediction of the optimal depth for superior vena cava cannulae with cardiac computed tomography during minimally invasive cardiac surgery: a prospective observational cohort study. BMC Anesthesiol 2017; 17:56. [PMID: 28388941 PMCID: PMC5383961 DOI: 10.1186/s12871-017-0347-x] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/14/2016] [Accepted: 03/29/2017] [Indexed: 11/10/2022] Open
Abstract
BACKGROUND The determination of the adequate depth of superior vena cava cannulae during minimally invasive cardiac surgery is important for warranting venous drainage and preventing complications during cardiopulmonary bypass. We investigated whether preoperative cardiac computed tomography might be useful for predicting the optimal depth of superior vena cava cannulae. METHODS The patients who required superior vena cava cannulation and had cardiac tomographic image among those scheduled to undergo a minimally invasive cardiac surgery were evaluated. The distance between the upper border of the clavicular sternal head and the superior vena cava-right atrium junction was measured on cardiac computed tomography. Equivalence test for the difference between the distance measured on cardiac computed tomography and the distance verified by surgeon's direct inspection in the surgical field was performed. The range -1 cm to 1 cm was predefined as an equivalence region. In addition, the distances between the upper border of the clavicular sternal head and the carina level on chest radiography were measured to compare the relative position of carina with regard to the superior vena cava-right atrium junction. RESULTS A total of 46 patients were evaluated. The distance from the upper border of the clavicular sternal head to the superior vena cava-right atrium junction measured on cardiac computed tomography and the distance verified by surgeon's inspection was equivalent, with the 95% confidence interval for the mean difference within the equivalence region (0.05-0.52, P < 0.0001). The carina level on chest radiography was found at least 2 cm above the superior vena cava-right atrium junction in all patients. CONCLUSIONS Preoperative cardiac computed tomography might be valuable for predicting the adequate depth of superior vena cava cannulae. Additionally, the carina on chest radiography might indicate a useful landmark for proper position of central venous catheter. TRIAL REGISTRATION This study has been registered at Clinical Research Information Service on 6 July 2012 (KCT0000477) .
Collapse
Affiliation(s)
- Ji-Hyun Chin
- Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea
| | - Eun-Ho Lee
- Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea.
| | - Jong-Il Kim
- Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea
| | - In-Cheol Choi
- Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea
| |
Collapse
|
6
|
von Segesser LK, Berdajs D, Abdel-Sayed S, Tozzi P, Ferrari E, Maisano F. New, Virtually Wall-Less Cannulas Designed for Augmented Venous Drainage in Minimally Invasive Cardiac Surgery. INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY 2016. [DOI: 10.1177/155698451601100408] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Affiliation(s)
| | | | | | | | | | - Francesco Maisano
- Department of Cardiovascular Surgery, University Hospital, Zurich, Switzerland
| |
Collapse
|
7
|
New, Virtually Wall-Less Cannulas Designed for Augmented Venous Drainage in Minimally Invasive Cardiac Surgery. INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY 2016; 11:278-81. [DOI: 10.1097/imi.0000000000000283] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/27/2022]
Abstract
Objective Inadequate venous drainage during minimally invasive cardiac surgery becomes most evident when the blood trapped in the pulmonary circulation floods the surgical field. The present study was designed to assess the in vivo performance of new, thinner, virtually wall-less, venous cannulas designed for augmented venous drainage in comparison to traditional thin-wall cannulas. Methods Remote cannulation was realized in 5 bovine experiments (74.0 ± 2.4 kg) with percutaneous venous access over the wire, serial dilation up to 18 F and insertion of either traditional 19 F thin wall, wire-wound cannulas, or through the same access channel, new, thinner, virtually wall-less, braided cannulas designed for augmented venous drainage. A standard minimal extracorporeal circuit set with a centrifugal pump and a hollow fiber membrane oxygenator, but no inline reservoir was used. One hundred fifty pairs of pump-flow and required pump inlet pressure values were recorded with calibrated pressure transducers and a flowmeter calibrated by a volumetric tank and timer at increasing pump speed from 1500 RPM to 3500 RPM (500-RPM increments). Results Pump flow accounted for 1.73 ± 0.85 l/min for wall-less versus 1.17 ± 0.45 l/min for thin wall at 1500 RPM, 3.91 ± 0.86 versus 3.23 ± 0.66 at 2500 RPM, 5.82 ± 1.05 versus 4.96 ± 0.81 at 3500 RPM. Pump inlet pressure accounted for 9.6 ± 9.7 mm Hg versus 4.2 ± 18.8 mm Hg for 1500 RPM, −42.4 ± 26.7 versus −123 ± 51.1 at 2500 RPM, and −126.7 ± 55.3 versus −313 ±116.7 for 3500 RPM. Conclusions At the well-accepted pump inlet pressure of −80 mm Hg, the new, thinner, virtually wall-less, braided cannulas provide unmatched venous drainage in vivo. Early clinical analyses have confirmed these findings.
Collapse
|
8
|
von Segesser LK, Tozzi P, Mallbiabrrena I, Jegger D, Horisberger J, Corno A. Miniaturization in cardiopulmonary bypass. Perfusion 2016; 18:219-24. [PMID: 14575410 DOI: 10.1191/0267659103pf676oa] [Citation(s) in RCA: 31] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
Cardiopulmonary bypass (CPB) remains the key technology for more complex cardiac operations. The perfusion equipment used nowadays has seen tremendous progress since its introduction into clinical practice 50 years ago. However, overall, CPB is still far from perfect. Major haemodilution is not only a problem for red cell-dependent gas transport, but also for the platelet and humeral factor-dependent coagulation, the protein-dependent intravascular oncotic pressure and so forth. Reduction of the priming volume through further miniaturization of CPB equipment is, therefore, the most obvious next step. A systematic approach needs to optimize all CPB components, including pumps, oxygenator/heat exchanger structures, filters, reservoirs, cardiotomy suction, tubings and cannulas. This report provides an update of already commercially available low prime perfusion devices (e.g., the CORx integrated pump-oxygenator) as well as promising prototypes like the smart suction system and the smartcanula.™
Collapse
Affiliation(s)
- Ludwig K von Segesser
- Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, CHUV, Lausanne, Switzerland.
| | | | | | | | | | | |
Collapse
|
9
|
Bainbridge DT, Chu MWA, Kiaii B, Cleland A, Murkin J. Percutaneous superior vena cava drainage during minimally invasive mitral valve surgery: a randomized, crossover study. J Cardiothorac Vasc Anesth 2014; 29:101-6. [PMID: 25440652 DOI: 10.1053/j.jvca.2014.07.020] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/19/2014] [Indexed: 11/11/2022]
Abstract
OBJECTIVE Minimally invasive techniques commonly are applied to mitral valve surgery; however, there has been little research investigating the optimal methods of cardiopulmonary bypass for the right minithoracotomy approach. Controversy exists as to whether a percutaneous superior vena cava drainage cannula (PSVC) is necessary during these operations. The authors, therefore, sought to determine the effect of using a percutaneous superior vena cava catheter on brain near-infrared spectroscopy, blood lactate levels, hemodynamics and surgical parameters. DESIGN Randomized, blinded, crossover trial. SETTING Tertiary care university hospital. PARTICIPANTS Patients undergoing minimally invasive mitral valve surgery via a right minithoracotomy. INTERVENTIONS Twenty minutes of either clamped or unclamped percutaneous superior vena cava neck catheter drainage, during mitral valve repair. MEASUREMENT AND MAIN RESULTS For the primary outcome of brain near-infrared spectroscopy, there were no differences between the two groups (percutaneous superior vena cava clamped 55.0%±11.6% versus unclamped 56.1%±10.2%) (p = 0.283). For the secondary outcomes pH (clamped 7.35±0.05 versus unclamped 7.37±0.05 p = 0.015), surgical score (clamped 1.96±1.14 versus unclamped 1.22±0.51 p = 0.002) and CVP (clamped 11.6 mmHg±4.8 mmHg versus unclamped 6.1 mmHg±6.1 mmHg p<0.001) were significantly different. CONCLUSIONS The use of a percutaneous superior vena cava drainage improved surgical visualization and lowered CVP, but had no effect on brain near infrared spectroscopy during minimally invasive mitral valve surgery. (ClinicalTrials.gov Identifier: NCT01166841).
Collapse
Affiliation(s)
- Daniel T Bainbridge
- Department of Anesthesiology and Perioperative Medicine, University of Western Ontario, London, Ontario Canada.
| | - Michael W A Chu
- Division of Cardiac Surgery, Department of Surgery, University of Western Ontario, London, Ontario Canada
| | - Bob Kiaii
- Division of Cardiac Surgery, Department of Surgery, University of Western Ontario, London, Ontario Canada
| | - Andrew Cleland
- Department of Perfusion Services, London Health Sciences Centre, London, Ontario, Canada
| | - John Murkin
- Department of Anesthesiology and Perioperative Medicine, University of Western Ontario, London, Ontario Canada
| |
Collapse
|
10
|
Abdel-Sayed S, Favre J, von Segesser LK. How to prevent venous cannula orifice obstruction during extracorporeal circulation. Perfusion 2014; 30:187-94. [DOI: 10.1177/0267659114560042] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
Venous cannula orifice obstruction is an underestimated problem during augmented cardiopulmonary bypass (CPB), which can potentially be reduced with redesigned, virtually wall-less cannula designs versus traditional percutaneous control venous cannulas. A bench model, allowing for simulation of the vena cava with various affluent orifices, venous collapse and a worst case scenario with regard to cannula position, was developed. Flow (Q) was measured sequentially for right atrial + hepatic + renal + iliac drainage scenarios, using a centrifugal pump and an experimental bench set-up (afterload 60 mmHg). At 1500, 2000 and 2500 RPM and atrial position, the Q values were 3.4, 6.03 and 8.01 versus 0.77*, 0.43* and 0.58* l/min: p<0.05* for wall-less and the Biomedicus® cannula, respectively. The corresponding pressure values were -15.18, -31.62 and -74.53 versus -46.0*, -119.94* and -228.13* mmHg. At the hepatic position, the Q values were 3.34, 6.67 and 9.26 versus 2.3*, 0.42* and 0.18* l/min; and the pressure values were -10.32, -20.25 and -42.83 versus -23.35*, -119.09* and -239.38* mmHg. At the renal position, the Q values were 3.43, 6.56 and 8.64 versus 2.48*, 0.41* and 0.22* l/min and the pressure values were -9.64, -20.98 and -63.41 versus -20.87 -127.68* and -239* mmHg, respectively. At the iliac position, the Q values were 3.43, 6.01 and 9.25 versus 1.62*, 0.55* and 0.58* l/min; the pressure values were -9.36, -33.57 and -44.18 versus -30.6*, -120.27* and -228* mmHg, respectivly. Our experimental evaluation demonstrates that the redesigned, virtually wall-less cannulas, allowing for direct venous drainage at practically all intra-venous orifices, outperform the commercially available control cannula, with superior flow at reduced suction levels for all scenarios tested.
Collapse
Affiliation(s)
- S Abdel-Sayed
- Cardiovascular Research, Department of Cardio-Vascular Surgery, University Hospital-CHUV, Lausanne, Switzerland
| | - J Favre
- Laboratory of Movement Analysis and Measurement, Ecole Polytechnique Fédéral de Lausanne, Lausanne, Switzerland
| | - LK von Segesser
- Cardiovascular Research, Department of Cardio-Vascular Surgery, University Hospital-CHUV, Lausanne, Switzerland
| |
Collapse
|
11
|
Performance Increase in Venous Drainage for Mini-Invasive Heart Surgery: Superiority of Self-Expanding Cannulas. INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY 2014. [DOI: 10.1177/155698451400900407] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
|
12
|
von Segesser LK. Peripheral cannulation for cardiopulmonary bypass. Multimed Man Cardiothorac Surg 2014; 2006:mmcts.2005.001610. [PMID: 24413332 DOI: 10.1510/mmcts.2005.001610] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
Peripheral cannulation is the historical route for connecting the pump-oxygenator to the vasculature of the patient in order to establish partial or complete cardiopulmonary bypass. Although most open heart procedures are nowadays realized with central cannulation, there is renewed interest in remote cannulation through the femoral, iliac, axillary, subclavian and jugular vessels. Remote cannulation is not only of interest in hemodynamically unstable patients who can be put on cardiopulmonary bypass in local anesthesia, and stabilized prior to intubation, but also for complex procedures like replacement of the thoracoabdominal aorta, acute type A aortic dissections, complex redo open heart surgery, extracorporeal membrane oxygenation, and more recently, small access open heart surgery, robotic surgery, and others. In the following shall be described femoral arterial cannulation with standard cannulas, femoral arterial cannulation with percutaneous cannulas, femoral venous cannulation using standard cannulas, femoral venous cannulation using percutaneous cannulas, as well as optimized venous cannulation relying on more advanced cannula designs.
Collapse
Affiliation(s)
- Ludwig K von Segesser
- Department of Cardio-vascular Surgery, Centre Hospitalier Universitaire Vaudois, CHUV, CCV, BH 10-275, Rue du Bugnon 46, CH-1011 Lausanne, Switzerland
| |
Collapse
|
13
|
Lei Q, Zeng QS, Zhang XS, Xie B, Luo ZC, Guo HM, Chen JM, Zhuang J. Superior vena cava drainage during thoracoscopic cardiac surgery: bilateral internal jugular vein sheaths versus one percutaneous superior vena cava cannula. J Cardiothorac Vasc Anesth 2013; 28:914-8. [PMID: 24139456 DOI: 10.1053/j.jvca.2013.05.043] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/25/2013] [Indexed: 11/11/2022]
Abstract
OBJECTIVE To evaluate bilateral internal jugular vein sheaths as a replacement of one percutaneous superior vena cava cannula for superior vena cava drainage during thoracoscopic cardiac surgery. DESIGN A prospective and randomized study. SETTING Single cardiovascular institute. PARTICIPANTS Adults undergoing thoracoscopic cardiac surgery. INTERVENTIONS Patients were randomized into a percutaneous superior vena cava cannula group and a bilateral internal jugular vein sheaths group. The superior vena cava drainage for cardiopulmonary bypass was performed with one percutaneous superior vena cava cannula (14-18 Fr) or the bilateral internal jugular vein sheaths (8 Fr). MEASUREMENTS AND MAIN RESULTS Both interventions reached theoretic flow rate in all patients. In patients weighing<50 kg (n=38) and 50-70 kg (n=64), the average central venous pressure values during cardiopulmonary bypass of both groups showed no significant differences. The patients weighing>70 kg (n=15) in the bilateral internal jugular vein sheaths group had a normal average central venous pressure value, but it was significantly higher than that of percutaneous superior vena cava cannula group ([10.5±3.1] mmHg vs. [4.5±4.4] mmHg, p=0.013). The patient satisfaction scale scores for the cervical incisions were significantly higher in the bilateral internal jugular vein sheaths group than in the percutaneous superior vena cava cannula group ([2.6±0.9] vs. [2.1±0.8], p=0.002). CONCLUSIONS The bilateral internal jugular vein sheaths were a feasible and effective option to replace one percutaneous superior vena cava cannula during thoracoscopic cardiac surgery, with better patient satisfaction.
Collapse
Affiliation(s)
- Qian Lei
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Qing-Shi Zeng
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Xiao-Shen Zhang
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Bin Xie
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Zhi-Chao Luo
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Hui-Ming Guo
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China.
| | - Ji-Mei Chen
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| | - Jian Zhuang
- Department of Cardiac Surgery and Anesthesiology, Guangdong Cardiovascular Institute, Guangdong General Hospital, and Guangdong Academy of Medical Sciences, Guangzhou, China
| |
Collapse
|
14
|
Characterizing the impact of minor cannula design modification. Int J Artif Organs 2012; 35:132-8. [PMID: 22388942 DOI: 10.5301/ijao.5000042] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 09/19/2011] [Indexed: 11/20/2022]
Abstract
OBJECTIVE Bench evaluation of the hydrodynamic behavior of venous cannulas is a valuable technique for the analysis of their performance during cardiopulmonary bypass (CPB). The aim of this study was to investigate the effect of the internal diameter of the extracorporeal connecting tube of venous cannulas on flow rate (Q), pressure drop (delta P), and cannula resistance (delta P/Q²) values, using a computer assisted test bench. METHODS An in vitro circuit was set up with silicone tubing between the test cannula encased in a movable reservoir, and a static reservoir. The delta P, defined as the difference between the drainage pressure and the preload pressure, was measured using high-fidelity Millar pressure transducers. Q was measured using an ultrasonic flowmeter. Data display and data recording were controlled using virtual instruments in a stepwise fashion. RESULTS The 27 F smartcanula® with a 9 mm connecting tube diameter showed 17% less resistance compared to that with an 8 mm connecting tube diameter. Q values were 7.22±0.1 and 7.81±0.04 L/min for cannulas with 8 mm and 9 mm connecting tube diameters, respectively. The delta P/Q² ratio values were 72% lower for the Medtronic cannula with a 9 mm connecting tube diameter compared to that with an 8 mm connecting tube diameter. Q values for the Medtronic cannula were 3.94±0.23 and 6.58±0.04 L/min with 8 mm and 9 mm connecting tube diameters, respectively. The 27 F smartcanula® showed 13% more flow rate compared to the 28 F Medtronic cannula using the unpaired Student t-test (p<0.0001). CONCLUSIONS Our results demonstrated that Q was increased but delta P and delta P/Q² values were significantly decreased when the connecting tube diameter was increased for venous cannulas. The connecting tube diameter significantly affected the resistance to liquid flow through the cannula. Smartcanulas® outperform Medtronic cannulas.
Collapse
|
15
|
Ohtake H, Yamaguchi S, Yashiki N, Kato H, Seguchi R, Ishikawa N, Watanabe G. Inferior vena cava occluder for remote access perfusion in robotic cardiac surgery. MINIM INVASIV THER 2011; 19:214-8. [PMID: 20166838 DOI: 10.3109/13645701003644426] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022]
Abstract
Robotic cardiac surgery requires remote access perfusion. We have developed an inferior vena cava (IVC) occluder that can safely and conveniently drain blood from the IVC. This device has been clinically applied in one patient, as described herein. The framework for this device is made from a single superelastic nitinol wire, 0.30 mm in diameter. Diameter of the spreading site of the device is decided from computed tomographic images. A polyester fabric membrane (thickness 0.10 mm) is set at the tip of this framework. The occluder is deployed through an 18-F sheath. This device was used in a 64-year-old woman with lipoma in the right atrial wall near the IVC-right atrium interface. In this patient, it might not have been possible to completely reset the tumour by conventional IVC occlusion using a snare. The occluder was smoothly and safely deployed and retracted. During placement of the occluder, blood did not flow from the IVC into the right atrium. During extracorporeal circulation, vacuum drainage was performed with no air contamination. The tumour was resected by a three-arm da Vinci Surgical System. The IVC occluder needs to completely block the IVC and avoid obstructing the inflow region of the hepatic vein. This device obviates the need to place a snare on the IVC, and thus should directly improve the safety of robotic cardiac surgery and shorten the operating time.
Collapse
Affiliation(s)
- Hiroshi Ohtake
- Department of General & Cardiothoracic Surgery, Kanazawa University, Kanazawa, Japan.
| | | | | | | | | | | | | |
Collapse
|
16
|
Transfusion-free neonatal cardiopulmonary bypass using a TinyPump. Ann Thorac Surg 2010; 90:1615-21. [PMID: 20971275 DOI: 10.1016/j.athoracsur.2010.07.034] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/23/2009] [Revised: 07/09/2010] [Accepted: 07/14/2010] [Indexed: 11/22/2022]
Abstract
BACKGROUND We devised a miniaturized circuit incorporating a TinyPump in the venous line to amplify the venous return. We compared this system to the conventional blood-primed circuit and investigated whether this circuit could maintain hematocrit levels without blood transfusion and reduce coagulation and inflammatory cascades. METHODS Thirteen 1-week-old piglets (3.7 ± 0.2 kg) were divided into group M (miniaturized circuits with TinyPump-assisted venous drainage without blood, n = 7) and group C (conventional circuits with blood priming, n = 6). Cardiopulmonary bypass (CPB) was performed at 150 to 180 mL·kg(-1)·min(-1) for 2 hours, including 60 minutes of cardioplegic cardiac arrest. Modified ultrafiltration (MUF) was subsequently performed. Data were acquired before CPB and after the end of MUF. RESULTS The priming volume including the hemofilter circuit of the main circuit required 152 mL in group M and 300 mL in group C. The mean hematocrit values in group M and group C were not significantly different during CPB (21.5% ± 2.0% versus 23.2% ± 1.3%) or after MUF (30.7% ± 2.1% versus 32.9% ± 4.0%). After MUF, group M had lower thrombin-antithrombin complex levels (16.7 ± 5.0 ng/mL versus 28.4 ± 8.4 ng/mL, p < 0.01) and interleukin-8 levels (2,867 ± 758 pg/mL versus 13,730 ± 5,220 pg/mL, p < 0.01) than group C. The pulmonary vascular resistance index was lower in group M after MUF (4,105 ± 862 dynes·cm(-5)·kg(-1) versus 6,304 ± 1,477 dynes·cm(-5)·kg(-1), p < 0.01). The lung water content was also better in group M (83.7% ± 0.5% versus 84.9% ± 0.5%, p < 0.01). CONCLUSIONS The minicircuit with TinyPump-assisted venous drainage successfully maintained acceptable hematocrit levels and the cardiopulmonary function in neonatal piglets. Employing this technique may attenuate blood requirements and inflammatory responses, thereby improving the clinical outcomes of neonatal open-heart surgery.
Collapse
|
17
|
Apostolakis EE, Koletsis EN, Baikoussis NG, Siminelakis SN, Papadopoulos GS. Strategies to prevent intraoperative lung injury during cardiopulmonary bypass. J Cardiothorac Surg 2010; 5:1. [PMID: 20064238 PMCID: PMC2823729 DOI: 10.1186/1749-8090-5-1] [Citation(s) in RCA: 101] [Impact Index Per Article: 6.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/24/2009] [Accepted: 01/11/2010] [Indexed: 12/02/2022] Open
Abstract
During open heart surgery the influence of a series of factors such as cardiopulmonary bypass (CPB), hypothermia, operation and anaesthesia, as well as medication and transfusion can cause a diffuse trauma in the lungs. This injury leads mostly to a postoperative interstitial pulmonary oedema and abnormal gas exchange. Substantial improvements in all of the above mentioned factors may lead to a better lung function postoperatively. By avoiding CPB, reducing its time, or by minimizing the extracorporeal surface area with the use of miniaturized circuits of CPB, beneficial effects on lung function are reported. In addition, replacement of circuit surface with biocompatible surfaces like heparin-coated, and material-independent sources of blood activation, a better postoperative lung function is observed. Meticulous myocardial protection by using hypothermia and cardioplegia methods during ischemia and reperfusion remain one of the cornerstones of postoperative lung function. The partial restoration of pulmonary artery perfusion during CPB possibly contributes to prevent pulmonary ischemia and lung dysfunction. Using medication such as corticosteroids and aprotinin, which protect the lungs during CPB, and leukocyte depletion filters for operations expected to exceed 90 minutes in CPB-time appear to be protective against the toxic impact of CPB in the lungs. The newer methods of ultrafiltration used to scavenge pro-inflammatory factors seem to be protective for the lung function. In a similar way, reducing the use of cardiotomy suction device, as well as the contact-time between free blood and pericardium, it is expected that the postoperative lung function will be improved.
Collapse
|
18
|
Jegger D, Sundaram S, Shah K, Mallabiabarrena I, Mucciolo G, von Segesser LK. Using computational fluid dynamics to evaluate a novel venous cannula (Smart canula) for use in cardiopulmonary bypass operating procedures. Perfusion 2008; 22:257-65. [PMID: 18181514 DOI: 10.1177/0267659107083657] [Citation(s) in RCA: 13] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
Peripheral access cardiopulmonary bypass (CPB) is initiated with percutaneous cannulae (CTRL) and venous drainage is often impeded due to smaller vessel and cannula size. A new cannula (Smartcanula, SC) was developed which can change shape in situ and, therefore, may improve venous drainage. Its performance was evaluated using a 2-D computational fluid dynamics (CFD) model. The Navier-Stokes equations could be simplified due to the fact that we use a steady state and a 2-dimensional system while the equation of continuity (p constant) was also simplified. We compared the results of the SC to the CTRL using CFDRC (Version 6.6, CFDRC research corporation, Huntsville, USA) at two preloads (300 and 700 Pa). The SC's mass flow rate outperformed the CTRL by 12.1% and 12.2% at a pressures of 300 and 700 Pa, respectively. At 700 Pa, a pressure gradient of 50% was measured for the CTRL and 11% for the SC. The mean velocity at the 700 Pa for the CTRL was 1.0 m.s(-1) at exit while the SC showed an exit velocity of 1.3 m.s(-1). Shear rates inside the cannulae were similar between the two cannulae. In conclusion, the prototype shows greater mass flow rates compared to the classic cannula; thus, it is more efficient. This is also advocated by a better pressure gradient and higher average velocities. By reducing cannula-tip surface area or increasing hole surface area, greater flow rates are achieved.
Collapse
Affiliation(s)
- D Jegger
- Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland.
| | | | | | | | | | | |
Collapse
|
19
|
von Segesser LK, Jegger D, Mucciolo G, Tozzi P, Mucciolo A, Delay D, Mallabiabarrena I, Horisberger J. The Smartcanula: a new tool for remote access perfusion in limited access cardiac surgery. Heart Surg Forum 2007; 8:E241-5. [PMID: 16112936 DOI: 10.1532/hsf98.20051127] [Citation(s) in RCA: 22] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
Devices for venous cannulation have seen significant progress over time: the original, rigid steel cannulas have evolved toward flexible plastic cannulas with wire support that prevents kinking, very thin walled wire wound cannulas allowing for percutaneous application, and all sorts of combinations. In contrast to all these rectilinear venous cannula designs, which present the same cross-sectional area over their entire intravascular path, the smartcanula concept of "collapsed insertion and expansion in situ" is the logical next step for venous access. Automatically adjusting cross-sectional area up to a pre-determined diameter or the vessel lumen provides optimal flow and ease of use for both, insertion and removal. Smartcanula performance was assessed in a small series of patients (76 +/- 17 kg) undergoing redo procedures. The calculated target pump flow (2.4 L/min/m2) was 4.42 +/- 61 L/ min. Mean pump flow achieved during cardiopulmonary bypass was 4.84 +/- 87 L/min or 110% of the target. Reduced atrial chatter, kink resistance in situ, and improved blood drainage despite smaller access orifice size, are the most striking advantages of this new device. The benefits of smart cannulation are obvious in remote cannulation for limited access cardiac surgery, but there are many other cannula applications where space is an issue, and that is where smart cannulation is most effective.
Collapse
Affiliation(s)
- Ludwig K von Segesser
- Department of Cardio-vascular Surgery, Centre Hospitalier Universitaire Vaudois, CHUV, Lausanne, Switzerland.
| | | | | | | | | | | | | | | |
Collapse
|
20
|
Gasparovic H, Byrne JG. The use of a femoral arterial cannula for effective dilation for percutaneous femoral venous cannulation. J Card Surg 2005; 20:68-9. [PMID: 15673414 DOI: 10.1111/j.0886-0440.2005.200403.x] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
Abstract
We present a technique for percutaneous femoral venous cannulation for initiation of cardiopulmonary bypass in the setting of previous femoral vessel exposure in which previous scarring prevents safe dilation of the subcutaneous tissues using standard techniques. The technique presented may be particularly helpful when redo sternotomy is deemed hazardous and cardiopulmonary bypass is judged mandatory prior to redo sternotomy. The patient presented in this paper had two previous cardiac operations with prior surgical exposure of femoral vessels in whom institution of cardiopulmonary bypass prior to sternotomy was of paramount importance due to a 7-cm ascending aortic aneurysm.
Collapse
Affiliation(s)
- Hrvoje Gasparovic
- Division of Cardiac Surgery, Brigham and Women's Hospital, 75 Francis Street, Boston, MA 02115, USA
| | | |
Collapse
|
21
|
Jegger D, Horisberger J, Boone Y, Seigneuil I, Jachertz M, Holzmann I, von Segesser LK. Vascular Access for Cardiopulmonary Bypass Procedures. Artif Organs 2004; 28:649-54. [PMID: 15209858 DOI: 10.1111/j.1525-1594.2004.07372.x] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/28/2022]
Abstract
Since the initiation of cardiac surgery using cardiopulmonary bypass, little progress has been made concerning the design of catheters for vascular access. However, in the last few years, research in this specialized field has established that catheter performance not only depends on size but also on the catheter's design. The catheter's drainage hole surface area correlates with its performance, i.e., flow; the ratio of the catheter's diameter to the patient's vein diameter also correlates with flow. These findings should influence the design of future models. An example is presented with the development of the Smartcanula which maximizes hole surface area and minimizes the wall thickness in order to optimize flow rate and vascular access to the patient.
Collapse
Affiliation(s)
- David Jegger
- Department of Cardiovascular Surgery, Center Hospitalier Universitaire Vaudois (CHUV), Rue du Bugnon, Lausanne, Switzerland.
| | | | | | | | | | | | | |
Collapse
|
22
|
Jegger D, Corno AF, Mucciolo A, Mucciolo G, Boone Y, Horisberger J, Seigneul I, Jachertz M, von Segesser LK. A prototype paediatric venous cannula with shape change in situ. Perfusion 2003; 18:61-5. [PMID: 12705652 DOI: 10.1191/0267659103pf640oa] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
During cardiopulmonary bypass (CPB), venous drainage may be impeded due to small vessel and cannula size or chattering, thus, blood return to the heart-lung machine is reduced. We designed a self-expandable prototype cannula, which is able to maintain the vein open and overcome this problem and analysed its performance capability. This prototype and several other cannulae were tested using an access vessel diameter of 7 mm. An in vitro circuit was set up with a 10 mm penrose latex tube simulating the patient's vein placed between the patient preload reservoir and the cannula, encasing the cannula's inlet(s). Maximum flow rate was determined for passive venous drainage (PVD) at preloads (P) of 2 and 4 mmHg. We compared these results to three classic single-stage venous cannulae: basket tip, thoracic drain and percutaneous tip. By comparing the other cannulae to the prototype, under PVD conditions and a central venous pressure (CVP) of 2 mmHg, the prototype cannula's flow rate (1.32 +/- 0.04 L/min) outperformed the basket type (the best performing comparator) (1.02 +/- 0.08 L/min) by 23% (p < 0.005). When the preload was increased to 4 mmHg under PVD conditions, the same trend was noted with the prototype cannula (1.65 +/- 0.05 L/min), outperforming the basket cannula's value (1.26 +/- 0.05 L/min) by 24% (p < 0.001). This new cannula design provides superior flow characteristics, under all test conditions, compared to the classic single-stage venous cannulae used for paediatric CPB surgery.
Collapse
Affiliation(s)
- David Jegger
- Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland.
| | | | | | | | | | | | | | | | | |
Collapse
|
23
|
Jegger D, Tevaearai HT, Mueller XM, Pierrel N, Horisberger J, von Segesser LK. Flow dynamic comparison of peripheral venous cannulas used with centrifugal pump assistance in vitro. Artif Organs 2002; 26:390-2. [PMID: 11952513 DOI: 10.1046/j.1525-1594.2002.06820.x] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
Because of the risk of vein collapse, the benefits of using a centrifugal pump to assist venous drainage for cardiopulmonary bypass are limited when the tips of peripheral cannulas are maintained within the vena cava. Using a mock circuit including 20 mm diameter latex tubing to mimic a vena cava, we compared the performance of 6 commercially available peripheral venous cannulas and attempted to determine potential factors influencing maximal flow drainage before vein collapse. A close correlation was observed with the total hole area of the cannula. Best performance (5.10 +/- 0.08 L/min) was obtained with an 8 mm internal diameter (ID) cannula and 343 mm2 total hole area. A larger cannula (ID 9.2 mm) with only 209 mm2 total hole area drained 5.03 +/- 0.05 L/min whereas a smaller cannula (ID 6.7 mm) with a total hole area of 586 mm2 also allowed a similar flow of 5.03 +/- 0.12 L/min. Therefore, the total hole area appears to be a critical factor in designing peripheral cannulas used in restricted chambers such as vena cavae.
Collapse
Affiliation(s)
- David Jegger
- Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland
| | | | | | | | | | | |
Collapse
|
24
|
Jegger D, Mueller X, Mucciolo G, Mucciolo A, Boone Y, Seigneul I, Horisberger J, von Segesser LK. A new expandable cannula to increase venous return during peripheral access cardiopulmonary bypass surgery. Int J Artif Organs 2002; 25:136-40. [PMID: 11908488 DOI: 10.1177/039139880202500208] [Citation(s) in RCA: 14] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
Peripheral cannulation for cardiopulmonary bypass (CPB) is of prime interest in minimally invasive open heart surgery. As CPB is initiated with percutaneous cannulae, venous drainage is impeded due to smaller vessel and cannula size. A new cannula was developed which can change shape in situ and therefore may improve venous drainage. An in vitro circuit was set-up with a penrose latex tubing placed between the preload reservoir and the cannula, encasing the cannula's inlet and simulating the vena cava. The preload (P) was stabilised at 2 and at 5 mmHg respectively. The maximum flow rate was determined for 4 conditions: passive venous drainage (PVD) and assisted venous drainage (AVD) using a centrifugal pump at the 2 preload settings. We compared the results of the prototype cannula to classical femoral venous cannulae: basket 28Fr, a thoracic 28Fr and a percutaneous 27Fr. Under PVD conditions and a CVP of 2 mmHg, the prototype cannula's flow rate outperformed the next best cannula by 14% (p=0.0002) and 13% under AVD conditions (p=0.0001). Under PVD conditions and a CVP of 5 mmHg, the prototype cannula outperformed the percutaneous cannula by 19% (p=0.0001) and 14% under AVD conditions (p=0.0002). The new cannula outperforms the classical percutaneous venous cannulae during all of the four conditions tested in vitro.
Collapse
Affiliation(s)
- D Jegger
- Department of Cardio-Vascular Surgery, Centre Hospitalier Universitaire Vaudois (CHUV), Lausanne, Switzerland.
| | | | | | | | | | | | | | | |
Collapse
|
25
|
Mueller XM, Tevaearai HT, Horisberger J, Augstburger M, Burki M, von Segesser LK. Vacuum assisted venous drainage does not increase trauma to blood cells. ASAIO J 2001; 47:651-4. [PMID: 11730205 DOI: 10.1097/00002480-200111000-00016] [Citation(s) in RCA: 17] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022] Open
Abstract
Although gravity drainage has been the standard technique for cardiopulmonary bypass (CPB), the development of min imally invasive techniques for cardiac surgery has renewed interest in using vacuum assisted venous drainage (VAVD) Dideco (Mirandola, Italy) has modified the D903 Avant oxygenator to apply a vacuum to its venous reservoir. The impact of VAVD on blood damage with this device is analyzed. Six calves (mean body weight, 71.3 +/- 4.1 kg) were con nected to CPB by jugular venous and carotid arterial cannu lation, with a flow rate of 4-4.51 L/min for 6 h. They were assigned to gravity drainage (standard D903 Avant oxygen ator, n = 3) or VAVD (modified D903 Avant oxygenator, n = 3). The animals were allowed to survive for 7 days. A standard battery of blood samples was taken before bypass, throughout bypass, and 24 h, 48 h, and 7 days after bypass. Analysis of variance was used for repeated measurements. Thrombocyte and white blood cell counts, corrected by hematocrit and normalized by prebypass values, were not significantly different between groups throughout all study periods. The same holds true for hemolytic parameters (lactate dehydrogenase [LDH] and plasma hemoglobin). Both peaked at 24 hr in the standard and VAVD groups: LDH, 2,845 +/- 974 IU/L vs. 2,537 +/- 476 IU/L (p = 0.65), respectively; and plasma hemoglobin, 115 +/- 31 mg/L vs. 89 +/- 455 mg/L (p = 0.45), respectively. In this experimental setup with prolonged perfusion time, VAVD does not increase trauma to blood cells in comparison with standard gravity drainage.
Collapse
Affiliation(s)
- X M Mueller
- Clinic for Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland
| | | | | | | | | | | |
Collapse
|
26
|
Tevaearai HT, Mueller XM, Jegger D, Ruchat P, von Segesser LK. Venous drainage with a single peripheral bicaval cannula for less invasive atrial septal defect repair. Ann Thorac Surg 2001; 72:1772-3. [PMID: 11722098 DOI: 10.1016/s0003-4975(01)02985-x] [Citation(s) in RCA: 13] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
New peripheral venous cannulae have recently been proposed for minimally invasive open cardiac surgery. We present a femoral venous cannula designed to simultaneously drain both superior and inferior vena cavae. Used in adult patients for atrial septal defect repair, the cannula allowed passive blood drainage of 70.6% +/- 11.7% of theoretical cardiac output. Drainage was augmented to 93.4% +/- 8.6% of theoretical cardiac output by means of a centrifugal pump.
Collapse
Affiliation(s)
- H T Tevaearai
- Department of Cardiovascular Surgery, University Hospital (CHUV), Lausanne, Switzerland.
| | | | | | | | | |
Collapse
|
27
|
Ni YM, Leskosek B, Shi LP, Chen YL, Qian LF, Li RY, Tu ZL, von Segesser LK. Optimization of venous return tubing diameter for cardiopulmonary bypass. Eur J Cardiothorac Surg 2001; 20:614-20. [PMID: 11509288 DOI: 10.1016/s1010-7940(01)00840-5] [Citation(s) in RCA: 20] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/18/2022] Open
Abstract
OBJECTIVE To determine the optimal venous tubing diameter for adult cardiopulmonary bypass (CPB) to improve gravity drainage and to reduce priming volume. METHODS (A) Maximum bovine blood flow rates by gravity drainage were assessed in vitro for four different tubing diameters (1/2, 3/8, 5/16,1/4 inch) with three different lengths and various pre- and afterloads. Based on the results of (A) and multiple regression analyses, we developed equations to predict tubing sizes as a function of target flows. (C) The equations obtained in (B) were validated by ex vivo bovine experiments. (D) The clinically required maximal flows were determined retrospectively by reviewing 119 perfusion records at Zurich University. (E) Based on our model (B), the clinical patient and hardware requirements, the optimal venous tubing diameter was calculated. (F) The optimized venous tubing was evaluated in a prospective clinical trial involving 312 patients in Hangzhou. RESULTS For a mean body surface area of 1.83+/-0.2 m(2), the maximal perfusion flow rate (D) achieved with 1/2-inch (=1.27 cm(2)) venous tubing was 4.62+/-0.57 l/min (range: 2.50-6.24 l/min). Our validated model (B,C) predicted 1.0 cm(2) as optimal cross-sectional area for the venous line. New tubing packs developed accordingly were used routinely thereafter. The maximal flow rate was 4.93+/-0.58 l/min (range: 3.9-7.0) in patients with a mean body surface area of 1.62+/-0.21 m(2). CONCLUSION The new venous tubing with 1.0-cm(2) cross-sectional area improves the drainage in the vast majority of adult patients undergoing CPB and reduces the priming volume (-27 ml/m). Reduced hemodilution can prevent homologous transfusions if a predefined transfusion trigger level is not reached.
Collapse
Affiliation(s)
- Y M Ni
- Department of Cardiothoracic Surgery, The First Affiliated Hospital, Medical College, Zhejiang University, Hangzhou, China
| | | | | | | | | | | | | | | |
Collapse
|