1
|
Romero M, Macchione MA, Mattea F, Strumia M. The role of polymers in analytical medical applications. A review. Microchem J 2020. [DOI: 10.1016/j.microc.2020.105366] [Citation(s) in RCA: 10] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
|
2
|
Resende TD, Lizar JC, Dos Santos FM, Borges LF, Pavoni JF. Study of the formulation optimization and reusability of a MAGAT gel dosimeter. Phys Med 2019; 63:105-111. [PMID: 31221401 DOI: 10.1016/j.ejmp.2019.05.018] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/26/2018] [Revised: 05/23/2019] [Accepted: 05/25/2019] [Indexed: 01/29/2023] Open
Abstract
PURPOSE This study aims to optimize the formulation of a methacrylic acid gelatine and tetrakis (hydroxymethyl) phosphonium chloride (MAGAT) gel dosimeter to achieve acceptable dosimetric characteristics and the lowest final costs. This study also evaluates the reusability of the dosimeter. METHODS The MAGAT gel dosimeter formulation was optimized. Tetrakis (hydroxymethyl) phosphonium chloride (THPC) concentrations (2, 5, 8, 10, 20, and 65 mM), methacrylic acid (MA) concentrations (2.0, 2.5, 3.0, 3.5, and 4.0% w/w) and gelatin concentrations (4.36, 6.45, 8.36, and 10.45% w/w) were evaluated to provide an adequate dosimetric response. The final dosimeter formulation linearity and dose rate dependence were evaluated. The reutilization methodology of the optimized gel formulation, but containing 2 mM of THPC, which was previously irradiated with a dose of 2 Gy, is also presented. RESULTS The optimized mass concentration of the dosimeter consists of 88.60% deionized water, 8.36% gelatin, 3.00% of MA and 0.04% THPC (5 mM). It presents a linear response for doses up to 10 Gy with a 1.16 Gy-1 s-1 sensitivity. A maximum sensitivity variation of less than 4.0% was found when varying the dose rate of the radiation beams from 300 to 500 cGy/min. It was possible to reuse the dosimeter, however the sensitivity decreased by 15% from the first to the second irradiation. CONCLUSIONS A low-cost MAGAT gel dosimeter with optimized formulation that responds to radiation in a dose range of 0 to 10 Gy with small dose-rate dependence is presented. The MAGAT gel can be reused after a 2 Gy irradiation.
Collapse
Affiliation(s)
- Thiago Dias Resende
- Physics Department, Faculty of Philosophy, Sciences and Letters at Ribeirão Preto, University of São Paulo, Av. Bandeirantes 3900, 14040-901, Monte Alegre, Ribeirão Preto, SP, Brazil
| | - Jessica Caroline Lizar
- Physics Department, Faculty of Philosophy, Sciences and Letters at Ribeirão Preto, University of São Paulo, Av. Bandeirantes 3900, 14040-901, Monte Alegre, Ribeirão Preto, SP, Brazil
| | - Fred Müller Dos Santos
- Physics Department, Faculty of Philosophy, Sciences and Letters at Ribeirão Preto, University of São Paulo, Av. Bandeirantes 3900, 14040-901, Monte Alegre, Ribeirão Preto, SP, Brazil
| | - Leandro Federiche Borges
- Radiotherapy Service, Clinics Hospital of University of São Paulo Medical School at Ribeirão Preto, Av. Bandeirantes 3900, 14040-900, Monte Alegre, Ribeirão Preto, SP, Brazil
| | - Juliana Fernandes Pavoni
- Physics Department, Faculty of Philosophy, Sciences and Letters at Ribeirão Preto, University of São Paulo, Av. Bandeirantes 3900, 14040-901, Monte Alegre, Ribeirão Preto, SP, Brazil; Radiotherapy Service, Clinics Hospital of University of São Paulo Medical School at Ribeirão Preto, Av. Bandeirantes 3900, 14040-900, Monte Alegre, Ribeirão Preto, SP, Brazil.
| |
Collapse
|
3
|
Shih CT, Hsu JT, Han RP, Hsieh BT, Chang SJ, Wu J. A novel method of estimating dose responses for polymer gels using texture analysis of scanning electron microscopy images. PLoS One 2013; 8:e67281. [PMID: 23843998 PMCID: PMC3699568 DOI: 10.1371/journal.pone.0067281] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/06/2013] [Accepted: 05/16/2013] [Indexed: 12/20/2022] Open
Abstract
Polymer gels are regarded as a potential dosimeter for independent validation of absorbed doses in clinical radiotherapy. Several imaging modalities have been used to convert radiation-induced polymerization to absorbed doses from a macro-scale viewpoint. This study developed a novel dose conversion mechanism by texture analysis of scanning electron microscopy (SEM) images. The modified N-isopropyl-acrylamide (NIPAM) gels were prepared under normoxic conditions, and were administered radiation doses from 5 to 20 Gy. After freeze drying, the gel samples were sliced for SEM scanning with 50×, 500×, and 3500× magnifications. Four texture indices were calculated based on the gray level co-occurrence matrix (GLCM). The results showed that entropy and homogeneity were more suitable than contrast and energy as dose indices for higher linearity and sensitivity of the dose response curves. After parameter optimization, an R2 value of 0.993 can be achieved for homogeneity using 500× magnified SEM images with 27 pixel offsets and no outlier exclusion. For dose verification, the percentage errors between the prescribed dose and the measured dose for 5, 10, 15, and 20 Gy were −7.60%, 5.80%, 2.53%, and −0.95%, respectively. We conclude that texture analysis can be applied to the SEM images of gel dosimeters to accurately convert micro-scale structural features to absorbed doses. The proposed method may extend the feasibility of applying gel dosimeters in the fields of diagnostic radiology and radiation protection.
Collapse
Affiliation(s)
- Cheng-Ting Shih
- Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, Hsinchu, Taiwan, R.O.C
| | - Jui-Ting Hsu
- School of Dentistry, China Medical University, Taichung, Taiwan, R.O.C
| | - Rou-Ping Han
- Department of Management Information Systems, Central Taiwan University of Science and Technology, Taichung, Taiwan, R.O.C
| | - Bor-Tsung Hsieh
- Department of Medical Imaging and Radiological Science, Central Taiwan University of Science and Technology, Taichung, Taiwan, R.O.C
| | - Shu-Jun Chang
- Health Physics Division, Institute of Nuclear Energy Research, Taoyuan, Taiwan, R.O.C
| | - Jay Wu
- Department of Biomedical Imaging and Radiological Science, China Medical University, Taichung, Taiwan, R.O.C
- * E-mail:
| |
Collapse
|
4
|
An evaluation of dosimetric characteristics of MAGIC gel modified by adding formaldehyde (MAGIC-f). RADIAT MEAS 2012. [DOI: 10.1016/j.radmeas.2012.10.004] [Citation(s) in RCA: 33] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
|
5
|
Sedaghat M, Bujold R, Lepage M. Preliminary studies on the role and reactions of tetrakis(hydroxymethyl)phosphonium chloride in polyacrylamide gel dosimeters. Phys Med Biol 2012; 57:5981-94. [DOI: 10.1088/0031-9155/57/19/5981] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
6
|
Pavoni JF, Pike TL, Snow J, DeWerd L, Baffa O. Tomotherapy dose distribution verification using MAGIC-fpolymer gel dosimetry. Med Phys 2012; 39:2877-84. [DOI: 10.1118/1.4704496] [Citation(s) in RCA: 26] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/27/2023] Open
|
7
|
Cheng HW, Ho CJ, Lee CC, Tu SJ, Shih BY, Chao TC. Development of a novel optical CT employing a laser to create a collimated line-source with a flat-top intensity distribution. RADIAT MEAS 2011. [DOI: 10.1016/j.radmeas.2011.05.032] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/18/2022]
|
8
|
Baldock C. Historical overview of the development of gel dosimetry: a personal perspective. ACTA ACUST UNITED AC 2007. [DOI: 10.1088/1742-6596/56/1/002] [Citation(s) in RCA: 45] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/27/2022]
|
9
|
|
10
|
|
11
|
Fuxman AM, McAuley KB, Schreiner LJ. Modelling of polyacrylamide gel dosimeters with spatially non-uniform radiation dose distributions. Chem Eng Sci 2005. [DOI: 10.1016/j.ces.2004.09.079] [Citation(s) in RCA: 30] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
|
12
|
Richardson SL, Tomé WA, Orton NP, McNutt TR, Paliwal BR. IMRT delivery verification using a spiral phantom. Med Phys 2004; 30:2553-8. [PMID: 14528978 DOI: 10.1118/1.1603965] [Citation(s) in RCA: 30] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/09/2023] Open
Abstract
In this paper we report on the testing and verification of a system for IMRT delivery quality assurance that uses a cylindrical solid water phantom with a spiral trajectory for radiographic film placement. This spiral film technique provides more complete dosimetric verification of the entire IMRT treatment than perpendicular film methods, since it samples a three-dimensional dose subspace rather than using measurements at only one or two depths. As an example, the complete analysis of the predicted and measured spiral films is described for an intracranial IMRT treatment case. The results of this analysis are compared to those of a single field perpendicular film technique that is typically used for IMRT QA. The comparison demonstrates that both methods result in a dosimetric error within a clinical tolerance of 5%, however the spiral phantom QA technique provides a more complete dosimetric verification while being less time consuming. To independently verify the dosimetry obtained with the spiral film, the same IMRT treatment was delivered to a similar phantom in which LiF thermoluminescent dosimeters were arranged along the spiral trajectory. The maximum difference between the predicted and measured TLD data for the 1.8 Gy fraction was 0.06 Gy for a TLD located in a high dose gradient region. This further validates the ability of the spiral phantom QA process to accurately verify delivery of an IMRT plan.
Collapse
Affiliation(s)
- Susan L Richardson
- Department of Radiology, Baylor College of Medicine, Houston, Texas 77030, USA
| | | | | | | | | |
Collapse
|
13
|
Islam KTS, Dempsey JF, Ranade MK, Maryanski MJ, Low DA. Initial evaluation of commercial optical CT-based 3D gel dosimeter. Med Phys 2003; 30:2159-68. [PMID: 12945982 DOI: 10.1118/1.1593636] [Citation(s) in RCA: 60] [Impact Index Per Article: 2.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
We evaluated the OCTOPUS-ONE research laser CT scanner developed and manufactured by MGS Research, Inc. (Madison, CT). The scanner is designed for imaging 3D optical density distributions in BANG gels. The scanner operates in a translate-rotate configuration with a single scanning laser beam. The rotating cylindrical gel phantom is immersed in a refractive index matching solution and positioned at the center of a square tank made of plastic and glass. A stationary polarized He-Ne laser beam (633 nm) is reflected from a mirror moving parallel to the tank wall and scans the gel. Another mirror moves synchronously along the opposite side of the tank and collects the transmitted light and sends it to a single stationary silicon photodetector. A filtered backprojection algorithm is used to reconstruct projection data in a plane. The laser-mirrors-detector assembly is mounted on a horizontal platform that moves vertically for slice selection. We have tested the mechanical and optical setup, projection centering on the axis of rotation, linearity, and spatial resolution. We found the optical detector to respond linearly to transmitted light from control samples. The spatial resolution of the scanner was determined by employing a split field resolution technique. We obtained the horizontal and vertical full widths at half maxima of the laser beam intensity profiles as 0.6 and 0.8 mm, respectively. Dose calibration tests of the gel were performed using a nine-field (2 x 2 cm2 each) dose pattern irradiated at different dose levels. Finally, we compared gel-derived 2D planar dose distribution against radiochromic film measured dose distribution for both the nine-field and a uniform 5 x 5 cm2 field of 6 MV x rays. Very similar dose distributions were observed in gel and radiochromic film except in regions of steep dose gradient and highest dose. A dose normalization of 15.6% was required between the two dosimeters due to differences in overall radiation response. After normalization, analysis using the gamma evaluation showed that the radiochromic film and gel-measured dose distributions differed by a maximum gamma of 1.3 using 5% and 1.5 mm dose difference and distance-to-agreement criteria. The optical CT scanner has great potential as a 3D dosimeter, but a few refinements and further testing are necessary before its routine clinical use.
Collapse
Affiliation(s)
- K T S Islam
- Department of Radiation Oncology, Mallinckrodt Institute, Washington University School of Medicine, St. Louis, Missouri 63110, USA
| | | | | | | | | |
Collapse
|
14
|
MacDougall ND, Pitchford WG, Smith MA. A systematic review of the precision and accuracy of dose measurements in photon radiotherapy using polymer and Fricke MRI gel dosimetry. Phys Med Biol 2002; 47:R107-21. [PMID: 12433119 DOI: 10.1088/0031-9155/47/20/201] [Citation(s) in RCA: 44] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/23/2022]
Abstract
The purpose of this work is to undertake a critical appraisal of the evidence in the published literature concerning the basic parameters of accuracy and precision associated with the use of Fricke and polymer gels (in conjunction with MR imaging) as radiation dosimeters in photon radiotherapy, condensing and analysing the body of published information (to the end of April 2002). A systematic review was undertaken addressing specific issues of precision and accuracy asking defined questions of the published literature. Accuracy and precision in relation to gel dosimetry were defined. Information was obtained from published, peer-reviewed journals. A defined search strategy utilizing MeSH headings and keywords, with extensive use of cross-referencing, identified 115 references dealing with gel dosimetry. Exclusion criteria were used to select only data from publications which would give unequivocal evidence. For accuracy, results had to be compared with an ionization chamber as gold standard and all gel samples had to be manufactured in the same batch. For precision, in addition to gels being from the same batch, samples must all have been irradiated at the same time and scanned simultaneously (or within a short time frame). Many results were found demonstrating 'dose mapping' examples using gels. However, there were very few publications containing firm evidence of precision and accuracy. There was no evidence which fulfilled our criteria about accuracy or precision using Fricke gels. For polymer gels only one paper was found for accuracy (4% (Low et al 1999 Med. Phys. 26 1542-51)) and precision (1.7% (Baldock et al 1998 Phys. Med. Biol. 43695-702)); however, both were carried out at only one dose level. If the exclusion criteria were relaxed to include accuracy results comparing gel to a non gold standard dosimeter (e.g. TLD), results give a median accuracy of 10% (range 8-23.5%) for polymer gel (Cosgrove et al 2000 Phys. Med. Biol. 45 1195-210, De Deene et al 1998 Radiother: Oncol. 48 283-91, Farajollahi et al 2000 Br. J. Radiol. 72 1085-92, McJury et al 1999b Phys. Med. Biol. 44 2431-44, Murphy et al 2000b Phys. Med. Biol. 45 835-45, Oldham et al 2001 Med. Phys. 28 1436-45) and 5% for Fricke gel (Chan and Ayyangar 1995b Med. Phys. 22 1171-5). Evidence also points to accuracy worsening at lower dose levels for both gels. The precision data should be viewed with caution as repeated MR measurements were not performed with the same samples. The only precision data for Fricke gels was 1.5% (Johansson Back et al 1998 Phys. Med. Biol. 43 261-76), but for zero dose. In conclusion, despite the amount of published data, sparse research has been undertaken which provides clear evidence of the accuracy and precision for both gels. That which has been published has used higher doses than would be routine in radiotherapy. The basic radiation dosimeter qualities of accuracy and precision have yet to be fully quantified for polymer and Fricke gels at clinically relevant dose levels.
Collapse
Affiliation(s)
- N D MacDougall
- Department of Medical Physics, Cookridge Hospital, Leeds, UK
| | | | | |
Collapse
|
15
|
Lepage M, McMahon K, Galloway GJ, De Deene Y, Bäck SAJ, Baldock C. Magnetization transfer imaging for polymer gel dosimetry. Phys Med Biol 2002; 47:1881-90. [PMID: 12108773 DOI: 10.1088/0031-9155/47/11/304] [Citation(s) in RCA: 53] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
Off-resonance RF pre-saturation was used to obtain contrast in MRI images of polymer gel dosimeters irradiated to doses up to 50 Gy. Two different polymer gel dosimeters composed of 2-hydroxyethyl-acrylate or methacrylic acid monomers mixed with N, N'-methylene-bisacrylamide (BIS), dispersed in an aqueous gelatin matrix were evaluated. Radiation-induced polymerization of the co-monomers generates a fast-relaxing insoluble polymer. Saturation of the polymer using off-resonance Gaussian RF pulses prior to a spin-echo readout with a short echo time leads to contrast that is dependent on the absorbed dose. This contrast is attributed to magnetization transfer (MT) between free water and the polymer, and direct saturation of water was found to be negligible under the prevailing experimental conditions. The usefulness of MT imaging was assessed by computing the dose resolution obtained with this technique. We found a low value of dose resolution over a wide range of doses could be obtained with a single experiment. This is an advantage over multiple spin echo (MSE) experiments using a single echo spacing where an optimal dose resolution is achieved over only very limited ranges of doses. The results suggest MT imaging protocols may be developed into a useful tool for polymer gel dosimetry.
Collapse
Affiliation(s)
- M Lepage
- Centre for Medical, Health and Environmental Physics, Queensland University of Technology, Brisbane, Australia.
| | | | | | | | | | | |
Collapse
|
16
|
Lepag M, Jayasakera PM, Bäck SA, Baldock C. Dose resolution optimization of polymer gel dosimeters using different monomers. Phys Med Biol 2001; 46:2665-80. [PMID: 11686281 DOI: 10.1088/0031-9155/46/10/310] [Citation(s) in RCA: 83] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
Polymer gel dosimeters of different formulations were manufactured from different monomers of acrylamide, acrylic acid, methacrylic acid, 1-vinyl-2-pyrrolidinone, 2-hydroxyethyl methacrylate and 2-hydroxyethyl acrylate. Gelatin and agarose were used as the gelling agents and N,N'-methylene-bis-acrylamide was used as a co-monomer in each polymer gel dosimeter. The T2 dependence of each dosimeter was analysed using a model of fast exchange of magnetization. The influence of the half-dose and the apparent T2 of the polymer-proton pool on the dose resolution (Dpdelta) were examined. Comparisons are made with the commonly employed R2-dose sensitivity. Differences exist suggesting that experiments reported in the literature using what were thought to be more optimal dosimeters may not actually be so. Based on Dpdelta of each formulation, conclusions are drawn on the optimal formulation required for a specific range of absorbed doses. In addition, information about the extent of polymerization of the monomers used along with some characteristics of the polymer network formed are reported. The influence of the concentration of monomers and gelling agent was subsequently evaluated using a model of fast exchange of magnetization. Based on these calculations, further improvement in Dpdelta can be expected.
Collapse
Affiliation(s)
- M Lepag
- Centre for Medical, Health and Environmental Physics, Queensland University of Technology, Brisbane, Australia
| | | | | | | |
Collapse
|
17
|
Papagiannis P, Pappas E, Kipouros P, Angelopoulos A, Sakelliou L, Baras P, Karaiskos P, Seimenis I, Sandilos P, Baltas D. Dosimetry close to an 192Ir HDR source using N-vinylpyrrolidone based polymer gels and magnetic resonance imaging. Med Phys 2001; 28:1416-26. [PMID: 11488573 DOI: 10.1118/1.1382603] [Citation(s) in RCA: 38] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/27/2023] Open
Abstract
In this work, the utilization of polymer gel-MRI dosimetry for measurements at distances relevant to clinical brachytherapy and intravascular applications [i.e., in the mm range, where steep three-dimensional (3-D) dose gradients exist] is investigated using N-vinylpyrrolidone-based gels. Transverse axis radial dose distributions, dose distributions parallel to the source axis, and 2-D dose distributions around the commonly used microSelectron 192Ir HDR source are measured for single source dwell position irradiations. Experimental results are found in good agreement with verified Monte Carlo calculations, even for distances less than 3 mm from the source. The effect of various MRI parameters, such as slice thickness, slice mispositioning, and in-plane resolution, on the accuracy of the method is also investigated. Possible limitations of the method are discussed, and its' overall potential in brachytherapy dosimetry is evaluated. Experimental 2-D dose distributions for an intravascular application following the Paris irradiation protocol are compared to corresponding commercial treatment planning system calculations. Results suggest that polymer gel-MRI dosimetry is capable of experimentally verifying dose distributions in relevant clinical intravascular applications.
Collapse
Affiliation(s)
- P Papagiannis
- Department of Physics, University of Athens, Greece.
| | | | | | | | | | | | | | | | | | | |
Collapse
|