Jiang L, Ramesh P, Neph R, Sheng K. Technical note: Multi-MATE, a high-throughput platform for automated image-guided small-animal irradiation.
Med Phys 2023;
50:7383-7389. [PMID:
37341036 PMCID:
PMC10733545 DOI:
10.1002/mp.16563]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/07/2022] [Accepted: 06/07/2023] [Indexed: 06/22/2023] Open
Abstract
BACKGROUND
Small animal irradiation is essential to study the radiation response of new interventions before or parallel to human therapy. Image-guided radiotherapy (IGRT) and intensity-modulated radiotherapy (IMRT) are recently adopted in small animal irradiation to more closely mimic human treatments. However, sophisticated techniques require exceedingly high time, resources, and expertize that are often impractical.
PURPOSE
We propose a high throughput and high precision platform named Multiple Mouse Automated Treatment Environment (Multi-MATE) to streamline image-guided small animal irradiation.
METHODS
Multi-MATE consists of six parallel and hexagonally arranged channels, each equipped with a transfer railing, a 3D-printed immobilization pod, and an electromagnetic control unit, computer-controlled via an Arduino interface. The mouse immobilization pods are transferred along the railings between the home position outside the radiation field and the imaging/irradiation position at the irradiator isocenter. All six immobilization pods are transferred to the isocenter in the proposed workflow for parallel CBCT scans and treatment planning. The immobilization pods are then sequentially transported to the imaging/therapy position for dose delivery. The positioning reproducibility of Multi-MATE are evaluated using CBCT and radiochromic films.
RESULTS
While parallelizing and automating the image-guided small animal radiation delivery, Multi-MATE achieved the average pod position reproducibility of 0.17 ± 0.04 mm in the superior-inferior direction, 0.20 ± 0.04 mm in the left-right direction, and 0.12 ± 0.02mm in the anterior-posterior direction in repeated CBCT tests. Additionally, in image-guided dose delivery tasks, Multi-MATE demonstrated the positioning reproducibility of 0.17 ± 0.06 mm in the superior-inferior direction, 0.19 ± 0.06 mm in the left-right direction.
CONCLUSIONS
We designed, fabricated, and tested a novel automated irradiation platform, Multi-MATE to accelerate and automate image-guided small animal irradiation. The automated platform minimizes human operation and achieves high setup reproducibility and image-guided dose delivery accuracy. Multi-MATE thus removes a major barrier to implementing high-precision preclinical radiation research.
Collapse