Krouwer JS. Comparing Clinical Accuracy and Analytical Accuracy Between Continuous Glucose Monitors. J Diabetes Sci Technol 2024; 18:608-609. [PMID: 38314690 PMCID: PMC11089867 DOI: 10.1177/19322968241229102] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 02/07/2024]
Abstract
This study compares performance between two continuous glucose monitors (CGMs). The study design contains a mix of laboratory results (CGM vs YSI) and home results (CGM vs glucose meter). Analysis is provided for both clinical accuracy and analytical accuracy of CGM glucose measurements. Both types of accuracy are important. Error grid analysis informs about clinical accuracy. Analytical error is important as most users would prefer a CGM with a smaller spread of CGM versus reference differences. The authors provide the percentage of time that no result was obtained. Study design, data analysis, and editorial support were provided by a manufacturer of one of the products studied. This study provides a template for comparisons.
Collapse
|
2
|
Luppa PB, Zeller M, Pieper M, Kaiser P, Weiss N, Vierbaum L, Freckmann G. Quality assessment of glucose measurement with regard to epidemiology and clinical management of diabetes mellitus in Germany. Front Mol Biosci 2024; 11:1371426. [PMID: 38572446 PMCID: PMC10987728 DOI: 10.3389/fmolb.2024.1371426] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/16/2024] [Accepted: 03/05/2024] [Indexed: 04/05/2024] Open
Abstract
Background During the last decade, Germany has seen an increased prevalence and a redistribution from undetected to diagnosed diabetes mellitus. Due to this substantial epidemiological development, the number of people with documented type 2 diabetes was 8.7 million in 2022. An estimated two million undiagnosed subjects are to be added. Beyond that, the life expectancy of diabetic subjects is increasing due to more responsive health systems in terms of care. Possible reasons include improved screening of at-risk individuals, the introduction of HbA1c for diagnosis in 2010, and the higher use of risk scores. Additionally, quality aspects of the laboratory methodology should be taken into consideration. Methods Epidemiology and clinical management of diabetes in Germany are presented in the light of publications retrieved by a selective search of the PubMed database. Additionally, the data from German external quality assessment (EQA) surveys for the measurands glucose in plasma and HbA1c in whole blood, reviewed from 2010 until 2022, were evaluated. Above this, data concerning the analytical performance of near-patient glucometer devices, according to the ISO norm 15197:2013, were analyzed. Results Two laboratory aspects are in good accordance with the observation of an increase in the diabetes mellitus prevalence when retrospectively reviewing the period 2010 to 2022: First, the analytical performance according to the ISO norm 15197:2013 of the glucometer devices widely used by patients with diabetes for the glucose self-testing, has improved during this period. Secondly, concerning the EQA program of INSTAND, the number of participating laboratories raised significantly in Germany. The spreads of variations of the specified results for plasma glucose remained unchanged between 2010 and 2022, whereas for HbA1c a significant decrease of the result scattering could be observed. Conclusion These retrospectively established findings testify to an excellent analytical quality of laboratory diagnostics for glucose and HbA1c throughout Germany which may be involved in a better diagnosis and therapy of previously undetected diabetes mellitus.
Collapse
Affiliation(s)
- Peter B. Luppa
- Institut für Klinische Chemie und Pathobiochemie, Klinikum rechts der Isar der Technische Universität München, Munich, Germany
| | - Michael Zeller
- Institut für Klinische Chemie und Pathobiochemie, Klinikum rechts der Isar der Technische Universität München, Munich, Germany
| | - Marija Pieper
- Institut für Klinische Chemie und Pathobiochemie, Klinikum rechts der Isar der Technische Universität München, Munich, Germany
| | - Patricia Kaiser
- INSTAND e.V., Gesellschaft zur Förderung der Qualitätssicherung in Medizinischen Laboratorien e.V., Düsseldorf, Germany
| | - Nathalie Weiss
- INSTAND e.V., Gesellschaft zur Förderung der Qualitätssicherung in Medizinischen Laboratorien e.V., Düsseldorf, Germany
| | - Laura Vierbaum
- INSTAND e.V., Gesellschaft zur Förderung der Qualitätssicherung in Medizinischen Laboratorien e.V., Düsseldorf, Germany
| | - Guido Freckmann
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| |
Collapse
|
3
|
ElSayed NA, Aleppo G, Bannuru RR, Bruemmer D, Collins BS, Ekhlaspour L, Hilliard ME, Johnson EL, Khunti K, Lingvay I, Matfin G, McCoy RG, Perry ML, Pilla SJ, Polsky S, Prahalad P, Pratley RE, Segal AR, Seley JJ, Stanton RC, Gabbay RA. 7. Diabetes Technology: Standards of Care in Diabetes-2024. Diabetes Care 2024; 47:S126-S144. [PMID: 38078575 PMCID: PMC10725813 DOI: 10.2337/dc24-s007] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/18/2023]
Abstract
The American Diabetes Association (ADA) "Standards of Care in Diabetes" includes the ADA's current clinical practice recommendations and is intended to provide the components of diabetes care, general treatment goals and guidelines, and tools to evaluate quality of care. Members of the ADA Professional Practice Committee, an interprofessional expert committee, are responsible for updating the Standards of Care annually, or more frequently as warranted. For a detailed description of ADA standards, statements, and reports, as well as the evidence-grading system for ADA's clinical practice recommendations and a full list of Professional Practice Committee members, please refer to Introduction and Methodology. Readers who wish to comment on the Standards of Care are invited to do so at professional.diabetes.org/SOC.
Collapse
|
4
|
Afeef S, Tolfrey K, Zakrzewski-Fruer JK, Barrett LA. Performance of the FreeStyle Libre Flash Glucose Monitoring System during an Oral Glucose Tolerance Test and Exercise in Healthy Adolescents. SENSORS (BASEL, SWITZERLAND) 2023; 23:s23094249. [PMID: 37177452 PMCID: PMC10180860 DOI: 10.3390/s23094249] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/20/2023] [Revised: 04/19/2023] [Accepted: 04/21/2023] [Indexed: 05/15/2023]
Abstract
This study's aim was to assess FreeStyle Libre Flash glucose monitoring (FGM) performance during an oral glucose tolerance test (OGTT) and treadmill exercise in healthy adolescents. This should advance the feasibility and utility of user-friendly technologies for metabolic assessments in adolescents. Seventeen healthy adolescents (nine girls aged 12.8 ± 0.9 years) performed an OGTT and submaximal and maximal treadmill exercise tests in a laboratory setting. The scanned interstitial fluid glucose concentration ([ISFG]) obtained by FGM was compared against finger-prick capillary plasma glucose concentration ([CPG]) at 0 (pre-OGTT), -15, -30, -60, -120 min post-OGTT, pre-, mid-, post- submaximal exercise, and pre- and post- maximal exercise. Overall mean absolute relative difference (MARD) was 13.1 ± 8.5%, and 68% (n = 113) of the paired glucose data met the ISO 15197:2013 criteria. For clinical accuracy, 84% and 16% of FGM readings were within zones A and B in the Consensus Error Grid (CEG), respectively, which met the ISO 15197:2013 criteria of having at least 99% of results within these zones. Scanned [ISFG] were statistically lower than [CPG] at 15 (-1.16 mmol∙L-1, p < 0.001) and 30 min (-0.74 mmol∙L-1, p = 0.041) post-OGTT. Yet, post-OGTT glycaemic responses assessed by total and incremental areas under the curve (AUCs) were not significantly different, with trivial to small effect sizes (p ≥ 0.084, d = 0.14-0.45). Further, [ISFGs] were not different from [CPGs] during submaximal and maximal exercise tests (interaction p ≥ 0.614). FGM can be a feasible alternative to reflect postprandial glycaemia (AUCs) in healthy adolescents who may not endure repeated finger pricks.
Collapse
Affiliation(s)
- Sahar Afeef
- School of Sport, Exercise and Health Sciences, Loughborough University, Loughborough LE11 3TU, UK
- Clinical Nutrition Department, Faculty of Applied Medical Sciences, King Abdulaziz University, Jeddah 21589, Saudi Arabia
| | - Keith Tolfrey
- School of Sport, Exercise and Health Sciences, Loughborough University, Loughborough LE11 3TU, UK
| | - Julia K Zakrzewski-Fruer
- Institute for Sport and Physical Activity Research, University of Bedfordshire, Bedford MK41 9EA, UK
| | - Laura A Barrett
- School of Sport, Exercise and Health Sciences, Loughborough University, Loughborough LE11 3TU, UK
| |
Collapse
|
5
|
ElSayed NA, Aleppo G, Aroda VR, Bannuru RR, Brown FM, Bruemmer D, Collins BS, Hilliard ME, Isaacs D, Johnson EL, Kahan S, Khunti K, Leon J, Lyons SK, Perry ML, Prahalad P, Pratley RE, Seley JJ, Stanton RC, Gabbay RA. 7. Diabetes Technology: Standards of Care in Diabetes-2023. Diabetes Care 2023; 46:S111-S127. [PMID: 36507635 PMCID: PMC9810474 DOI: 10.2337/dc23-s007] [Citation(s) in RCA: 110] [Impact Index Per Article: 110.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/14/2022]
Abstract
The American Diabetes Association (ADA) "Standards of Care in Diabetes" includes the ADA's current clinical practice recommendations and is intended to provide the components of diabetes care, general treatment goals and guidelines, and tools to evaluate quality of care. Members of the ADA Professional Practice Committee, a multidisciplinary expert committee, are responsible for updating the Standards of Care annually, or more frequently as warranted. For a detailed description of ADA standards, statements, and reports, as well as the evidence-grading system for ADA's clinical practice recommendations and a full list of Professional Practice Committee members, please refer to Introduction and Methodology. Readers who wish to comment on the Standards of Care are invited to do so at professional.diabetes.org/SOC.
Collapse
|
6
|
Tauschmann M, Forlenza G, Hood K, Cardona-Hernandez R, Giani E, Hendrieckx C, DeSalvo DJ, Laffel LM, Saboo B, Wheeler BJ, Laptev DN, Yarhere I, DiMeglio LA. ISPAD Clinical Practice Consensus Guidelines 2022: Diabetes technologies: Glucose monitoring. Pediatr Diabetes 2022; 23:1390-1405. [PMID: 36537528 PMCID: PMC10107687 DOI: 10.1111/pedi.13451] [Citation(s) in RCA: 28] [Impact Index Per Article: 14.0] [Reference Citation Analysis] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Accepted: 11/05/2022] [Indexed: 12/24/2022] Open
Affiliation(s)
- Martin Tauschmann
- Department of Pediatrics and Adolescent Medicine, Medical University of Vienna, Vienna, Austria
| | - Gregory Forlenza
- Pediatric Diabetes Division, Barbara Davis Center for Diabetes, University of Colorado Anschutz Medical Campus, Aurora, Colorado, USA
| | - Korey Hood
- Department of Psychiatry & Behavioral Sciences, Stanford University School of Medicine, Palo Alto, California, USA
| | | | - Elisa Giani
- Department of Biomedical Sciences, Humanitas University, Milan, Italy
| | - Christel Hendrieckx
- The Australian Centre for Behavioural Research in Diabetes, Diabetes Australia Victoria, Melbourne, Victoria, Australia.,School of Psychology, Deakin University, Geelong, Victoria, Australia
| | - Daniel J DeSalvo
- Texas Children's Hospital, Baylor College of Medicine, Houston, Texas, USA
| | - Lori M Laffel
- Pediatric, Adolescent and Young Adult Section, Joslin Diabetes Center, Boston, Massachusetts, USA.,Division of Endocrinology, Boston Children's Hospital, Boston, Massachusetts, USA
| | - Banshi Saboo
- Department of Diabetology, Diabetes Care and Hormone Clinic, Ambawadi, Ahmedabad, Gujarat, India
| | - Benjamin J Wheeler
- Department of Women's and Children's Health, University of Otago, Dunedin, New Zealand.,Paediatrics Department, Southern District Health Board, Dunedin, New Zealand
| | | | - Iroro Yarhere
- Endocrinology Unit, Paediatrics Department, University of Port Harcourt Teaching Hospital, Port Harcourt, Nigeria
| | - Linda A DiMeglio
- Division of Pediatric Endocrinology and Diabetology, Riley Hospital for Children, Indiana University School of Medicine, Indianapolis, Indiana, USA
| |
Collapse
|
7
|
Pleus S, Freckmann G, Schauer S, Heinemann L, Ziegler R, Ji L, Mohan V, Calliari LE, Hinzmann R. Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus. Diabetes Ther 2022; 13:829-846. [PMID: 35416589 PMCID: PMC9076772 DOI: 10.1007/s13300-022-01254-8] [Citation(s) in RCA: 7] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/18/2022] [Accepted: 03/14/2022] [Indexed: 11/25/2022] Open
Abstract
For decades, self-monitoring of blood glucose (SMBG) has been considered a cornerstone of adequate diabetes management. Structured SMBG can follow different monitoring patterns, and it results in improved glycemic control, reduced hypoglycemia, and a better quality of life of people with diabetes. The technology, usability, and accuracy of SMBG systems have advanced markedly since their introduction a few decades ago. Current SMBG systems are small and easy to use, require small (capillary) blood sample volumes, and provide measurement results within seconds. In addition, devices are increasingly equipped with features such as connectivity to other devices and/or digital diaries and diabetes management tools. Although measurement quality can come close to or equal that of the glucose monitoring systems used by healthcare professionals, several available SMBG systems still do not meet internationally accepted accuracy standards, such as the International Organization for Standardization 15197 standard. Reports from China, India, and Brazil based on local experience suggest that in addition of the accuracy issues of SMBG systems, other obstacles also need to be overcome to optimize SMBG usage. Nonetheless, adequate usage of SMBG data is of high relevance for the management of people with type 2 diabetes mellitus.
Collapse
Affiliation(s)
- Stefan Pleus
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| | - Guido Freckmann
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| | - Sebastian Schauer
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| | | | - Ralph Ziegler
- Diabetes Clinic for Children and Adolescents, Muenster, Germany
| | - Linong Ji
- Peking University People's Hospital, Peking, China
| | - Viswanathan Mohan
- Dr. Mohan's Diabetes Specialities Centre, Chennai, India
- Madras Diabetes Research Foundation, Chennai, India
| | - Luis Eduardo Calliari
- Pediatric Endocrine Unit, Pediatric Department, Santa Casa School of Medical Department, Santa Casa School of Medical Sciences, Sao Paulo, Brazil
| | - Rolf Hinzmann
- Roche Diabetes Care GmbH, Sandhofer Straße 116, 68305, Mannheim, Germany.
| |
Collapse
|
8
|
Eichenlaub M, Pleus S, Shaginian R, Richardson J, Pardo S, Stuhr A, Freckmann G. Impact of Blood Glucose Monitoring System Accuracy on Clinical Decision Making for Diabetes Management. J Diabetes Sci Technol 2022; 17:683-689. [PMID: 35227125 DOI: 10.1177/19322968221080916] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
BACKGROUND The accuracy of blood glucose monitoring systems (BGMS) is crucial for the safe and effective management of diabetes mellitus. Despite standardization of accuracy assessment procedures and requirements, various studies have shown that the accuracy of BGMS on the market can vary considerably. This article therefore provides health care professionals and users with an intuitive illustration of the impact of BGMS accuracy on clinical decision making. MATERIAL AND METHODS Several hypothetical patient scenarios based on blood glucose (BG) levels in the low, normal, and high BG range are devised. Using data from a recent BGMS accuracy study, a method for calculating the expected range of BG readings from four examined BGMS at the selected BG levels is introduced. Based on these ranges, it is illustrated how clinical decisions and subsequent outcomes of the hypothetical patients are affected by the expected inaccuracies of the BGMS. RESULTS The range of expected BGMS readings for the same true BG level can vary considerably between different BGMS. The discussion of hypothetical patient scenarios revealed that the use of some BGMS could be associated with an increased risk of adverse events such as failure to detect hypoglycemia, driving with an unsafe BG level, delay of treatment intervention in diabetes during pregnancy, or the failure to prevent diabetic ketoacidosis. CONCLUSIONS This article can support both health care professionals and patients to understand the impact of BGMS accuracy in a relatable, clinical context. Furthermore, it is suggested that current accuracy requirements might be insufficient for the prevention of adverse clinical outcomes in certain circumstances.
Collapse
Affiliation(s)
- Manuel Eichenlaub
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| | - Stefan Pleus
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| | - Rimma Shaginian
- Ascensia Diabetes Care Holdings AG, Global Medical Affairs, Basel, Switzerland
| | - James Richardson
- Ascensia Diabetes Care Holdings AG, Global Medical Affairs, Basel, Switzerland
| | - Scott Pardo
- Ascensia Diabetes Care, Global Clinical Affairs, Valhalla, USA
| | - Andreas Stuhr
- Ascensia Diabetes Care, Global Medical Affairs, Parsippany, USA
| | - Guido Freckmann
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany
| |
Collapse
|
9
|
Abstract
The American Diabetes Association (ADA) "Standards of Medical Care in Diabetes" includes the ADA's current clinical practice recommendations and is intended to provide the components of diabetes care, general treatment goals and guidelines, and tools to evaluate quality of care. Members of the ADA Professional Practice Committee, a multidisciplinary expert committee (https://doi.org/10.2337/dc22-SPPC), are responsible for updating the Standards of Care annually, or more frequently as warranted. For a detailed description of ADA standards, statements, and reports, as well as the evidence-grading system for ADA's clinical practice recommendations, please refer to the Standards of Care Introduction (https://doi.org/10.2337/dc22-SINT). Readers who wish to comment on the Standards of Care are invited to do so at professional.diabetes.org/SOC.
Collapse
|
10
|
Tanaka K, Higuchi R, Mizusawa K, Nakamura T, Nakajima K. Fasting biochemical hypoglycemia and related-factors in non-diabetic population: Kanagawa Investigation of Total Check-up Data from National Database-8. World J Diabetes 2021; 12:1131-1140. [PMID: 34326960 PMCID: PMC8311474 DOI: 10.4239/wjd.v12.i7.1131] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/08/2021] [Revised: 04/28/2021] [Accepted: 06/22/2021] [Indexed: 02/06/2023] Open
Abstract
BACKGROUND In healthy people, the lowest daily blood glucose concentration is usually observed in the early morning, after overnight fasting. However, the clinical relevance and the prevalence of fasting biochemical hypoglycemia (FBH) are poorly understood in people who do not have diabetes, although the clinical implications of such hypoglycemia have been extensively studied in patients with diabetes. FBH can be influenced by many factors, including age, sex, body mass, smoking, alcohol drinking, exercise levels, medications, and eating behaviors, such as breakfast skipping and late-night eating.
AIM To determine the prevalence of FBH and investigated its association with potential risk factors in a population without diabetes.
METHODS Clinical parameters and lifestyle-related factors were assessed in a cross-sectional study of 695613 people aged 40-74 years who had undergone a health check-up (390282 men and 305331 women). FBH was defined as fasting plasma glucose < 70 mg/dL (3.9 mmol/L) after overnight fasting, regardless of any symptoms. The absence of diabetes was defined as HbA1c < 6.5%, fasting plasma glucose < 126 mg/dL (7.0 mmol/L), and no pharmacotherapy for diabetes. Multivariate logistic regression analysis, with adjustment for confounding factors, was used to identify associations.
RESULTS FBH was present in 1842 participants (0.26%). There were significantly more women in the FBH group (59.1%) than in the non-FBH group (43.9%). Values of most of the clinical parameters, but not age, were significantly lower in the FBH group than in the non-FBH group. Logistic regression analysis showed that a body mass index of ≤ 20.9 kg/m2 (reference: 21-22.9 kg/m2) and current smoking were significantly associated with FBH, and this was not altered by adjustment for age, sex, and pharmacotherapy for hypertension or dyslipidemia. Female sex was associated with FBH. When the data were analyzed according to sex, men in their 60s or 70s appeared more likely to experience FBH compared with their 40s, whereas men in their 50s and women aged ≥ 50 years appeared less likely to experience FBH. The relationships of FBH with other factors including alcohol drinking and pharmacotherapies for hypertension and dyslipidemia also differed between men and women.
CONCLUSION FBH occurs even in non-diabetic people, albeit at a very low frequency. FBH is robustly associated with low body mass and smoking, and its relationship with lifestyle factors varies according to sex.
Collapse
Affiliation(s)
- Kotone Tanaka
- Graduate School of Health Innovation, Kanagawa University of Human Services, Kawasaki 210-0821, Japan
| | - Ryoko Higuchi
- School of Nutrition and Dietetics, Faculty of Health and Social Services, Kanagawa University of Human Services, Yokosuka 238-8522, Japan
| | - Kaori Mizusawa
- School of Nutrition and Dietetics, Faculty of Health and Social Services, Kanagawa University of Human Services, Yokosuka 238-8522, Japan
| | - Teiji Nakamura
- School of Nutrition and Dietetics, Faculty of Health and Social Services, Kanagawa University of Human Services, Yokosuka 238-8522, Japan
| | - Kei Nakajima
- School of Nutrition and Dietetics, Faculty of Health and Social Services, Kanagawa University of Human Services, Yokosuka 238-8522, Japan
- Department of Endocrinology and Diabetes, Saitama Medical Center, Saitama Medical University, Kawagoe 350-8550, Japan
| |
Collapse
|
11
|
Waist Circumference Is Not Associated with Impaired Fasting Blood Glucose in a Sample of Mexican Children and Teenagers: Results from a State Screening Program. CHILDREN-BASEL 2021; 8:children8030172. [PMID: 33668305 PMCID: PMC7996355 DOI: 10.3390/children8030172] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 02/04/2021] [Revised: 02/18/2021] [Accepted: 02/20/2021] [Indexed: 11/16/2022]
Abstract
A high waist circumference (WC) is used as a risk factor for impaired fasting blood glucose (IFG) in adults. This association is less studied in children and teenagers. The purpose of this study was to analyze the association between having a high WC and IFG by sex and age in a sample of Mexican children and teenagers. We analyzed the data of 12979 participants aged 5 to 17 years to calculate percentile references for uncorrected WC, corrected for height (WC/HT) and by height squared (WC/HT2) with quantile regression. A subsample of 2309 participants with fasting blood glucose samples (FBG), WC, WC/HT and WC/HT2 values was analyzed with logistic regression. A high WC, WC/HT, and WC/HT2 were considered at the sex- and age-specific 90th percentile from the subsample. The IFG was considered as FBG ≥100 mg/dL. Having a high WC, WC/HT, nor WC/HT2 was not significantly associated with IFG for either sex and age group (all p > 0.05). A high WC, either on its uncorrected or corrected for height values, was not an important assessment for predicting IFG in a sample of Mexican children and teenagers. This study provides percentile reference values specific for sex and age.
Collapse
|
12
|
Abstract
The American Diabetes Association (ADA) "Standards of Medical Care in Diabetes" includes the ADA's current clinical practice recommendations and is intended to provide the components of diabetes care, general treatment goals and guidelines, and tools to evaluate quality of care. Members of the ADA Professional Practice Committee, a multidisciplinary expert committee (https://doi.org/10.2337/dc21-SPPC), are responsible for updating the Standards of Care annually, or more frequently as warranted. For a detailed description of ADA standards, statements, and reports, as well as the evidence-grading system for ADA's clinical practice recommendations, please refer to the Standards of Care Introduction (https://doi.org/10.2337/dc21-SINT). Readers who wish to comment on the Standards of Care are invited to do so at professional.diabetes.org/SOC.
Collapse
|
13
|
Abstract
There has been a rapid advancement in the pace of development of new diabetes technologies and therapies for the management of type 1 diabetes over the past decade. The Diabetes Control and Complications Trial conclusively established that tight glycemic control with intensive insulin therapy decreases the rates of diabetes complications in proportion to glycemic control, and diabetes technologies have accordingly been developed to help patients reach these goals. In this review, the authors discuss new diabetes therapeutics and technologies, including new insulin analogues, insulin pumps, continuous glucose monitoring systems, and automated insulin delivery systems."
Collapse
Affiliation(s)
- Jordan S Sherwood
- Diabetes Research Center, Massachusetts General Hospital, 50 Staniford Street, Suite 301, Boston, MA 02114, USA
| | - Steven J Russell
- Diabetes Research Center, Massachusetts General Hospital, 50 Staniford Street, Suite 301, Boston, MA 02114, USA
| | - Melissa S Putman
- Diabetes Research Center, Massachusetts General Hospital, 50 Staniford Street, Suite 301, Boston, MA 02114, USA.
| |
Collapse
|
14
|
Pleus S, Baumstark A, Jendrike N, Mende J, Link M, Zschornack E, Haug C, Freckmann G. System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015. BMJ Open Diabetes Res Care 2020; 8:8/1/e001067. [PMID: 31958308 PMCID: PMC7039612 DOI: 10.1136/bmjdrc-2019-001067] [Citation(s) in RCA: 17] [Impact Index Per Article: 4.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/20/2019] [Revised: 12/20/2019] [Accepted: 12/25/2019] [Indexed: 01/12/2023] Open
Abstract
OBJECTIVE Accuracy of 18 current-generation blood glucose monitoring systems (BGMS) available in Europe was evaluated applying criteria adapted from EN ISO 15197:2015 with one reagent system lot. BGMS were selected based on market research data. RESEARCH DESIGN AND METHODS The BGMS ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Flex, Pic GlucoTest, Rightest GM700S, TRUEyou, and WaveSense JAZZ Wireless were tested using capillary blood from 100 different subjects and assessing the percentage of results within ±15 mg/dL (0.83 mmol/L) or 15% of comparison method results for BG concentrations below or above 100 mg/dL (5.55 mmol/L), respectively. In addition, the minimal deviation from comparison method results within which ≥95% of results of the respective BGMS were found was calculated. RESULTS In total, 14 BGMS had ≥95% of results within ±15 mg/dL (0.83 mmol/L) or ±15% and 3 BGMS had ≥95% of results within ±10 mg/dL (0.55 mmol/L) or ±10% of the results obtained with the comparison method. The smallest deviation from comparison method results within which ≥95% of results were found was ±7.7 mg/dL (0.43 mmol/L) or ±7.7%; the highest deviation was ±19.7 mg/dL (1.09 mmol/L) or ±19.7%. CONCLUSIONS This accuracy evaluation shows that not all CE-labeled BGMS fulfill accuracy requirements of ISO 15197 reliably and that there is considerable variation even among BGMS fulfilling these criteria. This safety-related information should be taken into account by patients and healthcare professionals when making therapy decisions. TRIAL REGISTRATION NUMBER NCT03737188.
Collapse
Affiliation(s)
- Stefan Pleus
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Annette Baumstark
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Nina Jendrike
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Jochen Mende
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Manuela Link
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Eva Zschornack
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Cornelia Haug
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| | - Guido Freckmann
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Universität Ulm, Ulm, Germany
| |
Collapse
|
15
|
Abstract
The American Diabetes Association (ADA) "Standards of Medical Care in Diabetes" includes the ADA's current clinical practice recommendations and is intended to provide the components of diabetes care, general treatment goals and guidelines, and tools to evaluate quality of care. Members of the ADA Professional Practice Committee, a multidisciplinary expert committee (https://doi.org/10.2337/dc20-SPPC), are responsible for updating the Standards of Care annually, or more frequently as warranted. For a detailed description of ADA standards, statements, and reports, as well as the evidence-grading system for ADA's clinical practice recommendations, please refer to the Standards of Care Introduction (https://doi.org/10.2337/dc20-SINT). Readers who wish to comment on the Standards of Care are invited to do so at professional.diabetes.org/SOC.
Collapse
|
16
|
Fabris C, Ozaslan B, Breton MD. Continuous Glucose Monitors and Activity Trackers to Inform Insulin Dosing in Type 1 Diabetes: The University of Virginia Contribution. SENSORS 2019; 19:s19245386. [PMID: 31817678 PMCID: PMC6961036 DOI: 10.3390/s19245386] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 10/31/2019] [Revised: 11/19/2019] [Accepted: 11/21/2019] [Indexed: 01/11/2023]
Abstract
Objective: Suboptimal insulin dosing in type 1 diabetes (T1D) is frequently associated with time-varying insulin requirements driven by various psycho-behavioral and physiological factors influencing insulin sensitivity (IS). Among these, physical activity has been widely recognized as a trigger of altered IS both during and following the exercise effort, but limited indication is available for the management of structured and (even more) unstructured activity in T1D. In this work, we present two methods to inform insulin dosing with biosignals from wearable sensors to improve glycemic control in individuals with T1D. Research Design and Methods: Continuous glucose monitors (CGM) and activity trackers are leveraged by the methods. The first method uses CGM records to estimate IS in real time and adjust the insulin dose according to a person’s insulin needs; the second method uses step count data to inform the bolus calculation with the residual glucose-lowering effects of recently performed (structured or unstructured) physical activity. The methods were tested in silico within the University of Virginia/Padova T1D Simulator. A standard bolus calculator and the proposed “smart” systems were deployed in the control of one meal in presence of increased/decreased IS (Study 1) and following a 1-hour exercise bout (Study 2). Postprandial glycemic control was assessed in terms of time spent in different glycemic ranges and low/high blood glucose indices (LBGI/HBGI), and compared between the dosing strategies. Results: In Study 1, the CGM-informed system allowed to reduce exposure to hypoglycemia in presence of increased IS (percent time < 70 mg/dL: 6.1% versus 9.9%; LBGI: 1.9 versus 3.2) and exposure to hyperglycemia in presence of decreased IS (percent time > 180 mg/dL: 14.6% versus 18.3%; HBGI: 3.0 versus 3.9), tending toward optimal control. In Study 2, the step count-informed system allowed to reduce hypoglycemia (percent time < 70 mg/dL: 3.9% versus 13.4%; LBGI: 1.7 versus 3.2) at the cost of a minor increase in exposure to hyperglycemia (percent time > 180 mg/dL: 11.9% versus 7.5%; HBGI: 2.4 versus 1.5). Conclusions: We presented and validated in silico two methods for the smart dosing of prandial insulin in T1D. If seen within an ensemble, the two algorithms provide alternatives to individuals with T1D for improving insulin dosing accommodating a large variety of treatment options. Future work will be devoted to test the safety and efficacy of the methods in free-living conditions.
Collapse
|
17
|
Weissenbacher S, Yang CY, Kuan TC, Demircik F, Hanna M, Pfützner A. System accuracy assessments with a blood glucose meter with combined glucose and ß-hydroxybutyrate measurement capabilities. Expert Rev Mol Diagn 2019; 19:1043-1048. [PMID: 31482753 DOI: 10.1080/14737159.2019.1662300] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/26/2022]
Abstract
Background: We evaluated the Wellion Galileo GLU/KET blood and ketone (ß-Hydroxybutyrate, ß-OHB) meter to demonstrate that it meets ISO15107:2015 regulatory approval criteria. Research Design and Methods: A total of 100 subjects (52 female, age: 30 to 84 years, diabetes: 10 type 1/90 type 2) with blood glucose levels distributed over the entire measurement range as required by the ISO15197 protocol were tested (double determinations with 3 strip lots and two devices). A similar test protocol was followed to test ß-OHB strip performance (reference devices: YSI 2300plus for glucose and STANBIO ß-HOB LiquiColor TestKit for ß-OHB). Precision was tested for glucose with 3 blood glucose concentrations (ß-OHB: 2 control solutions). Results: All glucose test-strip lots met the strict ISO acceptance criteria. Mean absolute relative difference (MARD) was 4.9% and all data pairs were in zone A of the consensus error grid. The ß-OHB test-strips also met the pre-defined acceptance criteria. Within-run and between-run precision was calculated to be 2.3% and 0.7% for the glucose strips (3.7%/0.8% for the ketone strips). Conclusions: When tested according to the ISO15197:2015 guideline, the device showed very accurate measurement performance for glucose and ß-OHB testing and fully met regulatory accuracy approval criteria.
Collapse
Affiliation(s)
| | | | | | | | - Mina Hanna
- Pfützner Science & Health Institute , Mainz , Germany
| | - Andreas Pfützner
- Pfützner Science & Health Institute , Mainz , Germany.,Department of Biotechnology, Technical University Bonn-Rhein-Sieg , Rheinbach , Germany.,Department of Internal Medicine and Laboratory Medicine, University for Digitalized Technologies in Medicine & Dentistry , Luxembourg
| |
Collapse
|
18
|
Klonoff DC. Postmarket Surveillance of Blood Glucose Monitor Systems Is Needed for Safety of Subjects and Accurate Determination of Effectiveness in Clinical Trials of Diabetes Drugs and Devices. J Diabetes Sci Technol 2019; 13:419-423. [PMID: 30974987 PMCID: PMC6501534 DOI: 10.1177/1932296819843398] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
Affiliation(s)
- David C. Klonoff
- Diabetes Research Institute, Mills-Peninsula Medical Center, San Mateo, CA, USA
- David C. Klonoff, MD, FACP, FRCP (Edin), Fellow AIMBE, Diabetes Research Institute, Mills-Peninsula Medical Center, 100 South San Mateo Drive, Rm 5147, San Mateo, CA 94401, USA.
| |
Collapse
|