Vis A, Arfaee M, Khambati H, Slaughter MS, Gummert JF, Overvelde JTB, Kluin J. The ongoing quest for the first total artificial heart as destination therapy. Nat Rev Cardiol 2022; 19:813-828. [PMID: 35668176 DOI: 10.1038/s41569-022-00723-8] [Citation(s) in RCA: 9] [Impact Index Per Article: 4.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Accepted: 05/09/2022] [Indexed: 12/18/2022]
Abstract
Many patients with end-stage heart disease die because of the scarcity of donor hearts. A total artificial heart (TAH), an implantable machine that replaces the heart, has so far been successfully used in over 1,700 patients as a temporary life-saving technology for bridging to heart transplantation. However, after more than six decades of research on TAHs, a TAH that is suitable for destination therapy is not yet available. High complication rates, bulky devices, poor durability, poor biocompatibility and low patient quality of life are some of the major drawbacks of current TAH devices that must be addressed before TAHs can be used as a destination therapy. Quickly emerging innovations in battery technology, wireless energy transmission, biocompatible materials and soft robotics are providing a promising opportunity for TAH development and might help to solve the drawbacks of current TAHs. In this Review, we describe the milestones in the history of TAH research and reflect on lessons learned during TAH development. We summarize the differences in the working mechanisms of these devices, discuss the next generation of TAHs and highlight emerging technologies that will promote TAH development in the coming decade. Finally, we present current challenges and future perspectives for the field.
Collapse
Affiliation(s)
- Annemijn Vis
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Maziar Arfaee
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Husain Khambati
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Mark S Slaughter
- Department of Cardiovascular and Thoracic Surgery, University of Louisville, Louisville, KY, USA
| | - Jan F Gummert
- Department of Thoracic and Cardiovascular Surgery, Heart and Diabetes Center NRW, Ruhr-University Bochum, Bad Oeynhausen, Germany
| | - Johannes T B Overvelde
- Autonomous Matter Department, AMOLF, Amsterdam, The Netherlands.,Institute for Complex Molecular Systems and Department of Mechanical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands
| | - Jolanda Kluin
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands. .,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands.
| |
Collapse
|
2
|
Kwant PB, Finocchiaro T, Förster F, Reul H, Rau G, Morshuis M, El Banayosi A, Körfer R, Schmitz-Rode T, Steinseifer U. The MiniACcor: Constructive Redesign of an Implantable Total Artificial Heart, Initial Laboratory Testing and Further Steps. Int J Artif Organs 2018; 30:345-51. [PMID: 17520573 DOI: 10.1177/039139880703000411] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
The Aachen Total Artificial Heart (ACcor) has been under development at the Helmholtz Institute in Aachen over the last decade. It may serve as a bridge to transplant or as a long-term replacement of the natural heart. Based upon previous in vivo experiments with the ACcor total artificial heart, it was decided to optimize and redesign the pump unit. Smaller dimensions, passive filling and separability into three components were the three main design goals. The new design is called the MiniACcor, which is about 20% smaller than its predecessor, and weighs only 470 grams. Also its external driver/control unit was miniaturized and a new microcontroller was selected. To validate the design, it was extensively tested in laboratory mock loops. The MiniACcor was able to pump between 4.5 and 7 l/min at different pump rates against normal physiological pressures. Several requirements for the future compliance chamber and transcutaneous energy transmission (TET) system were also measured in the same mock loop. Further optimization and validation are being performed in cooperation with the Heart and Diabetes Centre North Rhine-Westphalia.
Collapse
Affiliation(s)
- P B Kwant
- Helmholtz Institute, Aachen, Germany.
| | | | | | | | | | | | | | | | | | | |
Collapse
|
3
|
Fukamachi K, Horvath DJ, Massiello AL, Fumoto H, Horai T, Rao S, Golding LAR. An innovative, sensorless, pulsatile, continuous-flow total artificial heart: device design and initial in vitro study. J Heart Lung Transplant 2009; 29:13-20. [PMID: 19782599 DOI: 10.1016/j.healun.2009.05.034] [Citation(s) in RCA: 69] [Impact Index Per Article: 4.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/29/2009] [Revised: 05/27/2009] [Accepted: 05/27/2009] [Indexed: 11/17/2022] Open
Abstract
BACKGROUND We are developing a very small, innovative, continuous-flow total artificial heart (CFTAH) that passively self-balances left and right pump flows and atrial pressures without sensors. This report details the CFTAH design concept and our initial in vitro data. METHODS System performance of the CFTAH was evaluated using a mock circulatory loop to determine the range of systemic and pulmonary vascular resistance (SVR and PVR) levels over which the design goal of a maximum absolute atrial pressure difference of 10 mm Hg is achieved for a steady-state flow condition. Pump speed was then modulated at 2,600 +/- 900 rpm to induce flow and arterial pressure pulsation to evaluate the effects of speed pulsations on the system performance. An automatic control mode was also evaluated. RESULTS Using only passive self-regulation, pump flows were balanced and absolute atrial pressure differences were maintained at <10 mm Hg over a range of SVR (750 to 2,750 dyne.sec.cm(-5)) and PVR (135 to 600 dyne.sec.cm(-5)) values far exceeding normal levels. The magnitude of induced speed pulsatility affected relative left/right performance, allowing for an additional active control to improve balanced flow and pressure. The automatic control mode adjusted pump speed to achieve targeted pump flows based on sensorless calculations of SVR and CFTAH flow. CONCLUSIONS The initial in vitro testing of the CFTAH with a single, valveless, continuous-flow pump demonstrated its passive self-regulation of flows and atrial pressures and a new automatic control mode.
Collapse
Affiliation(s)
- Kiyotaka Fukamachi
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA
| | | | | | | | | | | | | |
Collapse
|
4
|
Kamohara K, Weber S, Klatte RS, Ootaki Y, Akiyama M, Kopcak MW, Luangphakdy V, Flick CR, Chen JF, Navia JL, Smith WA, Fukamachi K. Replacement of the left-side valves of an implanted total artificial heart. ASAIO J 2006; 52:368-72. [PMID: 16883114 DOI: 10.1097/01.mat.0000227731.46835.1e] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022] Open
Abstract
The MagScrew total artificial heart (TAH) is under development. Despite its anticipated durability and reliability, the possibility of a bioprosthetic valve malfunction exists. As a result, the potential for valve replacement surgery, instead of device replacement, would be desirable after a TAH implant. In two of our 90-day animal experiments, we successfully replaced the left-side valves through a left thoracotomy opposite to the right-sided incision site for the initial TAH implant. The results of these cases suggest that the left-side valves could also be replaced through a left thoracotomy approach in humans. To confirm the ability to access the left-side valves in humans, four human cadaver studies were performed with the use of a mock pump designed for human application. This report describes the operative techniques for left-side valve replacement in animals and discusses the advantages of a left thoracotomy in clinical situations, based on results from the human cadaver studies.
Collapse
Affiliation(s)
- Keiji Kamohara
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA
| | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
5
|
Weber S, Kamohara K, Klatte RS, Luangphakdy V, Flick C, Chen JF, Casas F, Ootaki Y, Kopcak M, Akiyama M, Hirschman GB, Chapman PA, Donahue A, Wetterau W, Prisco C, Mast R, Sherman C, Fukamachi K, Smith WA. MagScrew TAH: an update. ASAIO J 2006; 51:xxxvi-xlvi. [PMID: 16340348 DOI: 10.1097/01.mat.0000187395.29817.36] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022] Open
Abstract
The MagScrew Total Artificial Heart (TAH) system is the result of a close collaboration among the Cleveland Clinic Foundation, Foster Miller Technologies, Wilson Greatbatch Ltd, and Whalen Biomedical Inc. The system components are the thoracic blood pumping unit with attached compliance chamber and refill port, implantable electronic control unit, implantable battery pack, transcutaneous energy transmission system, external battery pack, and a telemetry system for communication with the electronic control unit. System in vitro tests are underway for system characterization and durability demonstration, whereas in vivo tests were conducted to evaluate system performance and biocompatibility under physiologic conditions. The passively filling pump uses a left master alternate left and right ejection control mode and has a Starling law-like response to venous pressure. The in vitro tests documented excellent hydraulic pump performance with high device output of over 9 l/min at left atrial pressures below 12 mm Hg. Atrial balance was well maintained under all test conditions. The in vivo tests demonstrated good biocompatibility without use of anticoagulant therapy. Experimental durations have ranged between 0 and 92 days. Postexplant evaluation of tissue samples did not reveal any sign of thromboembolic events or tissue damage due to device operation.
Collapse
Affiliation(s)
- Stephan Weber
- The Cleveland Clinic Foundation, Cleveland, OH 44195, USA.
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
6
|
Schenk S, Weber S, Luangphakdy V, Klatte RS, Flick CR, Chen JF, Kopcak MW, Ootaki Y, Kamohara K, Hirschman GB, Vitale NG, Chapman PA, Smith WA, Fukamachi K. Magscrew Total Artificial Heart In Vivo Performance Above 200 Beats Per Minute. Ann Thorac Surg 2005; 79:1378-83; discussion 1383. [PMID: 15797082 DOI: 10.1016/j.athoracsur.2004.03.064] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Accepted: 03/15/2004] [Indexed: 10/25/2022]
Abstract
PURPOSE Downsizing pulsatile devices requires an increase of beat rate if flow capacity is to be maintained. We applied this concept to the preclinical MagScrew total artificial heart (TAH). DESCRIPTION The device fills passively with a stroke volume of 45 ml and beat rates up to 250 beats per minute (bpm). EVALUATION Stable hemodynamics were observed during a 30-day bovine implant with a flow of 8.7 +/- 1.2 L/min at beat rates of 204 +/- 18 bpm. Device filling was exceptional up to 250 bpm generating flow of greater than 12 L/min. Beat rate adapted to preload in a way similar to a Frank-Starling response. Left and right atrial pressures were balanced. The aortic pulse pressure was 49-70 mm Hg, which translates to a pulsatility index of 0.49-0.77. Organ functions were preserved and blood damage did not occur. CONCLUSIONS Increasing the beat rate while downsizing the MagScrew TAH was successful with strong flow generation by passive filling. Pulsatility was maintained at high beat rates. This innovative approach may be used to develop small pulsatile pumps.
Collapse
Affiliation(s)
- Soren Schenk
- Department of Biomedical Engineering, The Cleveland Clinic Foundation, Cleveland, Ohio 44195, USA
| | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
7
|
Schenk S, Weber S, Luangphakdy V, Flick CR, Chen JF, Inoue M, Kopcak MW, Ootaki Y, Doi K, Dessoffy R, Hirschman GB, Vitale NG, Chapman PA, Smith WA, Fukamachi K. In vivo performance and biocompatibility of the MagScrew ventricular assist device. ASAIO J 2004; 49:594-8. [PMID: 14524571 DOI: 10.1097/01.mat.0000084107.46300.21] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022] Open
Abstract
Currently available ventricular assist devices (VADs) have limitations in long-term durability and blood compatibility. We evaluated a prototype of a pulsatile MagScrew VAD for in vivo hemodynamic performance and biocompatibility. The device is composed of an actuator, blood pump housing, diaphragm, pusher plate, and bioprosthetic valves. Its protein-coated ("biolized") blood-contacting surface inhibits clot formation. Forces between moving parts of the actuator are transmitted magnetically, eliminating a primary source of friction and wear. The pump fills passively and is highly preload sensitive. The device was implanted into three calves for 90, 10, and 57 days, respectively. No anticoagulants were given postoperatively. The device functioned without technical problems during the entire course of each experiment, with mean device flow ranging between 5.4 and 9.0 L/min. Autopsy of the first two calves revealed no sign of embolization and clean blood-contacting surfaces of the devices. The third experiment was complicated by a prosthetic valve endocarditis with infectious embolization, and a few small depositions were found in the pump. In conclusion, the MagScrew VAD has demonstrated a high level of performance and biocompatibility in three calves studied for 10-90 days. Vigorous development is in progress to bring this device to preclinical readiness and thus provide surgeons with the VAD of choice for permanent implantation.
Collapse