1
|
Gottfried TM, Galeazzi P, Föger A, Dejaco D, Tröger A, Fischer N, Innerhofer V, Di Trapani F, Weiss N, Seebacher J, Dierker A, Schmutzhard J. Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation. Eur Arch Otorhinolaryngol 2024; 281:4121-4131. [PMID: 38564010 PMCID: PMC11266372 DOI: 10.1007/s00405-024-08584-2] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/15/2024] [Accepted: 02/26/2024] [Indexed: 04/04/2024]
Abstract
PURPOSE Cochlear implantation is a prevalent remedy for severe-to-profound hearing loss. Optimising outcomes and hearing preservation, and minimising insertion trauma, require precise electrode placement. Objective monitoring during the insertion process can provide valuable insights and enhance surgical precision. This study assesses the feasibility and performance of an impedance-based method for monitoring electrode insertion, compared to the surgeon's feedback. METHODS The study utilised the Insertion Monitoring Tool (IMT) research software, allowing for real-time measurement of impedance and evoked compound action potential (eCAP) during electrode insertion in 20 patient implantations. This enabled an impedance-based method to continuously assess the status of each electrode during the insertion process. The feasibility and performance was evaluated and compared to the surgeon's feedback approach. eCAP measurements focused merely on feasibility without searching specific responses. RESULTS The IMT demonstrated feasibility in measuring real-time impedances and eCAP during the insertion of the electrode array. The impedance-based method exhibited potential for accurately monitoring the insertion depth with a high success rate. However, further development is needed to improve the number of usable contacts. CONCLUSIONS Objective monitoring with the impedance-based method shows promise as a valuable tool to enhance the precision of cochlear implant electrode insertion respecting insertion distance estimation. The IMT research software proved feasible in recording real-time impedances and eCAP during electrode insertion. While this impedance-based method exhibits high success rates, further improvements are required to optimise the number of usable contacts. This study highlights the potential of objective monitoring techniques to enhance cochlear implantation outcomes.
Collapse
Affiliation(s)
- Timo M Gottfried
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Pablo Galeazzi
- MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria
| | - Aline Föger
- MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria
| | - Daniel Dejaco
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Andrea Tröger
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Natalie Fischer
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Veronika Innerhofer
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Francesco Di Trapani
- MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria
| | - Nora Weiss
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Bochum, North Rhine-Westphalia, Bleichstraße 15, 44787, Bochum, Germany
- Department of Otorhinolaryngology, TUM School of Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany
- Department of Translational Neurosciences, Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium
| | - Josef Seebacher
- Departement of Hearing, Speech and Voice Disorders, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria
| | - Angelika Dierker
- MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria
| | - Joachim Schmutzhard
- Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria.
| |
Collapse
|
2
|
Pisa JFD, Almutairi WH, Mackie K, Stangherlin D, Sulkers J, Hochman JB. Impedance Change as an Indicator of Cochlear Implant Failure. Otol Neurotol 2023; 44:e211-e215. [PMID: 36764700 DOI: 10.1097/mao.0000000000003819] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/12/2023]
Abstract
OBJECTIVE In February 2020, Advanced Bionics initiated a field action notice to remove the MS Ultra (V1) implant from circulation. In this study, we quantify a single site's incidence with device failure and examine the relationship between impedance change and declining speech perception. STUDY DESIGN Retrospective/cohort study. SETTING Tertiary healthcare center. PATIENTS Forty-nine adult patients (52 devices) were implanted between October 2017 and December 2019, with the following exclusion criteria: <18 years of age, medical/surgical failures, and lack of English-language proficiency. INTERVENTIONS Diagnostic. MAIN OUTCOME MEASURES Postoperative speech perception (AzBio sentence test) scores at 12 months postactivation were compared with repeat testing at 6- to 12-month intervals. Degree of change in impedances from 1 month postactivation was analyzed at similar intervals. Device failures were suspected when impedance levels on three or more electrodes dropped to <3.5 kΩ and AzBio scores declined >15%. Device failures were confirmed through analyses completed by the manufacturer. RESULTS To date, 21% (11/52) of devices have met all three criteria for a confirmed failure with an average decline in speech perception of 27.7 ± 22%. The average length of time for detection of device failure was 21.9 ± 5 months postimplantation. An additional 7 devices (13%) are currently being monitored for failure but have not yet met full criteria. There was no predictive relationship between degree of impedance change and speech perception scores. Electrodes along the entire array (channels 1-16) can be affected. CONCLUSIONS Impedance values can be used as a reliable indicator of device malfunction, thus facilitating patient counseling and early intervention. No correlation was identified between impedance changes and speech perception scores.
Collapse
Affiliation(s)
| | | | - Kristy Mackie
- Central Speech and Hearing Clinic, Winnipeg, Manitoba, Canada
| | | | - Jacob Sulkers
- Central Speech and Hearing Clinic, Winnipeg, Manitoba, Canada
| | | |
Collapse
|