Abstract
OBJECTIVE
Implantable retinal prostheses aim to provide artificial vision to those suffering from retinal degenerative diseases by electrically stimulating the remaining retinal neurons using a multi-electrode array. The spatial resolution of these devices can be improved by creation of so-called virtual channels (VCs) that are commonly achieved through synchronized stimulation of multiple electrodes. It is largely unclear though if VCs can be created using asynchronous stimulation, which was the primary aim of this study.
APPROACH
A computational model of multi-layered retina and epi-retinal dual-electrode stimulation was developed to simulate the neural activity of populations of retinal ganglion cells (RGCs) using the VC strategy under both synchronous and asynchronous stimulation conditions.
MAIN RESULTS
Our simulation suggests that VCs can be created using asynchronous stimulation. VC performance under both synchronous and asynchronous stimulation conditions can be improved by optimizing stimulation parameters such as current intensity, current ratio (α) between two electrodes, electrode spacing and the stimulation waveform. In particular, two VC performance measures; (1) linear displacement of the centroid of RGC activation, and (2) the RGC activation size consistency as a function of different current ratios α, have comparable performance under asynchronous and synchronous stimulation with appropriately selected stimulation parameters.
SIGNIFICANCE
Our findings support the possibility of creating VCs in the retina under both synchronous and asynchronous stimulation conditions. The results provide theoretical evidence for future retinal prosthesis designs with higher spatial resolution and power efficiency whilst reducing the number of current sources required to achieve these outcomes.