1
|
Ramesh TV, Narongrit FW, Rispoli JV. Adaptable, wearable, and stretchable coils: A review. Magn Reson Med 2025; 93:2186-2208. [PMID: 39902582 DOI: 10.1002/mrm.30428] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/25/2024] [Revised: 12/18/2024] [Accepted: 12/26/2024] [Indexed: 02/05/2025]
Abstract
Over the last four decades, there have been various evolutions in the design and development of coils, from volume coils to the recent introduction of wireless receive arrays. A recent aim has been to develop coils that can closely conform to the anatomy of interest to increase the acquired signal. This goal has given rise to designs ranging from adaptable transmit coils to on-body stretchable receive arrays made using fabric or elastomer substrates. This review covers the design, fabrication details, experimental setup, and MRI results of adaptable, wearable, and stretchable MRI coils. The active and passive automatic tuning and matching strategies are examined with respect to mitigating signal-to-noise ratio reduction when the coil form is altered. A brief discussion of wireless MRI coils, which provide a solution to overcome the cabling issues associated with MRI coil development, is also included. The adaptable, wearable, and stretchable coils and various coil tuning techniques represent innovative radiofrequency coil solutions that pave the way for next-generation MRI hardware development.
Collapse
Affiliation(s)
- Thejas Vishnu Ramesh
- Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana, USA
| | - Folk W Narongrit
- Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana, USA
- Elmore Family School of Electrical and Computer Engineering, Purdue University, West Lafayette, Indiana, USA
| | - Joseph V Rispoli
- Department of Radiology and Medical Imaging, University of Virginia, Charlottesville, Virginia, USA
| |
Collapse
|
2
|
Cui J, Hollingsworth NA, Wright SM. A Review of Current Control and Decoupling Methods for MRI Transmit Arrays. IEEE Rev Biomed Eng 2025; 18:388-400. [PMID: 38194402 DOI: 10.1109/rbme.2024.3351713] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/11/2024]
Abstract
The shortened radio frequency wavelength in high field MRI makes it challenging to create a uniform excitation pattern over a large field of view, or to achieve satisfactory transmission efficiency at a local area. Transmit arrays are one tool that can be used to create a desired excitation pattern. To be effective, it is important to be able to control the current amplitude and phase at the array elements. The control of the current may get complicated by the coil coupling in many applications. Various methods have been proposed to achieve current control, either in the presence of coupling, or by effectively decouple the array elements. These methods are applied in different subsystems in the RF transmission chain: coil; coil-amplifier interface; amplifier, etc. In this review paper, we provide an overview of the various approaches and aspects of transmit current control and decoupling.
Collapse
|
3
|
Puchnin V, Jandaliyeva A, Hurshkainen A, Solomakha G, Nikulin A, Petrova P, Lavrenteva A, Andreychenko A, Shchelokova A. Quadrature transceive wireless coil: Design concept and application for bilateral breast MRI at 1.5 T. Magn Reson Med 2023; 89:1251-1264. [PMID: 36336799 DOI: 10.1002/mrm.29507] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/21/2022] [Revised: 09/20/2022] [Accepted: 10/09/2022] [Indexed: 11/09/2022]
Abstract
PURPOSE Development of a novel quadrature inductively driven transceive wireless coil for breast MRI at 1.5 T. METHODS A quadrature wireless coil (HHMM-coil) design has been developed as a combination of two linearly polarized coils: a pair of 'metasolenoid' coils (MM-coil) and a pair of Helmholtz-type coils (HH-coil). The MM-coil consisted of an array of split-loop resonators. The HH-coil design included two electrically connected flat spirals. All the wireless coils were coupled to a whole-body birdcage coil. The HHMM-coil was studied and compared to the linear coils in terms of transmit and SAR efficiencies via numerical simulations. A prototype of HHMM-coil was built and tested on a 1.5 T scanner in a phantom and healthy volunteer. We also proposed an extended design of the HHMM-coil and compared its performance to a dedicated breast array. RESULTS Numerical simulations of the HHMM-coil with a female voxel model have shown more than a 2.5-fold increase in transmit efficiency and a 1.7-fold enhancement of SAR efficiency compared to the linearly polarized coils. Phantom and in vivo imaging showed good agreement with the numerical simulations. Moreover, the HHMM-coil provided good image quality, visualizing all areas of interest similar to a multichannel breast array with a 32% reduction in signal-to-noise ratio. CONCLUSION The proposed quadrature HHMM-coil allows the B 1 + $$ {\mathrm{B}}_1^{+} $$ -field to be significantly better focused in the region-of-interest compared to the linearly polarized coils. Thus, the HHMM-coil provides high-quality breast imaging on a 1.5 T scanner using a whole-body birdcage coil for transmit and receive.
Collapse
Affiliation(s)
- Viktor Puchnin
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| | | | - Anna Hurshkainen
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| | - Georgiy Solomakha
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| | - Anton Nikulin
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| | - Polina Petrova
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| | - Anna Lavrenteva
- Medical Institute named after Berezin Sergey (MIBS), St. Petersburg, Russia
| | - Anna Andreychenko
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia.,Research and Practical Clinical Center for Diagnostics and Telemedicine Technologies, Moscow Health Care Department, Moscow, Russia
| | - Alena Shchelokova
- School of Physics and Engineering, ITMO University, St. Petersburg, Russia
| |
Collapse
|
4
|
Alkandari D, Bosshard JC, Huang CH, Wright SM. Multiple slot modules for high field magnetic resonance imaging array coils. Magn Reson Med 2023; 89:2485-2498. [PMID: 36763854 DOI: 10.1002/mrm.29610] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/07/2022] [Revised: 01/15/2023] [Accepted: 01/16/2023] [Indexed: 02/12/2023]
Abstract
PURPOSE Mitigating coupling effects between coil elements represents a continuing challenge. Here, we present a 16-bowtie slot volume coil arranged in eight independent dual-slot modules without the use of any decoupling circuits. METHODS Two electrically short "bowtie" slot antennas were used to form a "module." A bowtie configuration was chosen because electromagnetic modeling results show that bowtie slots exhibit improved B 1 + P in $$ \frac{B_1^{+}}{\sqrt{P_{in}}} $$ efficiency when compared to thin rectangular slots. An eight-module volume coil was evaluated through electromagnetic modeling, bench tests, and MRI experiments at 4.7 T. RESULTS Bench tests indicate that worst-case coupling between modules did not exceed -14.5 dB. MR images demonstrate well-localized patterns about single excited modules confirming the low coupling between modules. Homogeneous MR images were acquired from a synthesized quadrature birdcage transmit mode. MRI experiments show that the RF power requirements for the proposed coil are 9.2 times more than a birdcage coil. Whereas from simulations performed to assess the proposed coil losses, the total power dissipated in the phantom was 1.1 times more for the birdcage. Simulation results at 7 T reveal an equivalent B1 + homogeneity when compared with an eight-dipole coil. CONCLUSION Although exhibiting higher RF power requirements, as a transmit coil when the power availability is not a restriction, the inherently low coupling between electrically short slots should enable the use of many slot elements around the imaging volume. The slot module described in this paper should be useful in the design of multi-channel transmit coils.
Collapse
Affiliation(s)
- Dheyaa Alkandari
- Department of Electrical Engineering, Kuwait University, Kuwait City, Kuwait
| | - John C Bosshard
- Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| | - Chung-Huan Huang
- Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| | - Steven M Wright
- Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA.,Department of Biomedical Engineering, Texas A&M University, College Station, Texas, USA
| |
Collapse
|
5
|
Wilcox M, Ogier S, Cheshkov S, Dimitrov I, Malloy C, Wright S, McDougall M. A 16-Channel 13C Array Coil for Magnetic Resonance Spectroscopy of the Breast at 7T. IEEE Trans Biomed Eng 2021; 68:2036-2046. [PMID: 33651680 DOI: 10.1109/tbme.2021.3063061] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/13/2023]
Abstract
OBJECTIVE Considering the reported elevation of ω-6/ω-3 fatty acid ratios in breast neoplasms, one particularly important application of 13C MRS could be in more fully understanding the breast lipidome's relationship to breast cancer incidence. However, the low natural abundance and gyromagnetic ratio of the 13C isotope lead to detection sensitivity challenges. Previous 13C MRS studies have relied on the use of small surface coils with limited field-of-view and shallow penetration depths to achieve adequate signal-to-noise ratio (SNR), and the use of receive array coils is still mostly unexplored. METHODS This work presents a unilateral breast 16-channel 13C array coil and interfacing hardware designed to retain the surface sensitivity of a single small loop coil while improving penetration depth and extending the field-of-view over the entire breast at 7T. The coil was characterized through bench measurements and phantom 13C spectroscopy experiments. RESULTS Bench measurements showed receive coil matching better than -17 dB and average preamplifier decoupling of 16.2 dB with no evident peak splitting. Phantom MRS studies show better than a three-fold increase in average SNR over the entirety of the breast region compared to volume coil reception alone as well as an ability for individual array elements to be used for coarse metabolite localization without the use of single-voxel or spectroscopic imaging methods. CONCLUSION Our current study has shown the benefits of the array. Future in vivo lipidomics studies can be pursued. SIGNIFICANCE Development of the 16-channel breast array coil opens possibilities of in vivo lipidomics studies to elucidate the link between breast cancer incidence and lipid metabolics.
Collapse
|
6
|
Del Bosque R, Cui J, Ogier S, Cheshkov S, Dimitrov IE, Malloy C, Wright SM, McDougall M. A 32-channel receive array coil for bilateral breast imaging and spectroscopy at 7T. Magn Reson Med 2020; 85:551-559. [PMID: 32820540 DOI: 10.1002/mrm.28425] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/12/2019] [Revised: 06/19/2020] [Accepted: 06/22/2020] [Indexed: 11/09/2022]
Abstract
PURPOSE This work describes the construction and evaluation of a bilateral 32-channel receive array for breast imaging at 7T. METHODS The receive array consisted of 32 receive coils, placed on two 3D-printed hemispherical formers. Each side of the receive array consisted of 16 receive loops, each loop having a corresponding detachable board with match/tune capacitors, active detuning circuitry, and a balun. Coil performance was evaluated on homogeneous canola oil phantoms using a Philips Achieva 7T system. Array coil performance was compared with a bilateral forced current excitation volume coil in transmit/receive mode and with a previously reported 16-channel unilateral coil with a similar design. RESULTS The 32-channel array had an increase in average SNR throughout both phantoms by a factor of five as compared with the volume coil, with SNR increases up to 10 times along the periphery and three times in the center. Noise measurements showed low interelement noise correlation (average: 5.4%; maximum: 16.8%). Geometry factor maps were acquired for various acceleration factors and showed mean geometry factors <1.2, for combined acceleration factors of up to six. CONCLUSIONS The improvements achieved demonstrate the clear potential for use in dynamic contrast-enhanced or diffusion-weighted MR studies, while maintaining diagnostically relevant spatial and temporal resolutions.
Collapse
Affiliation(s)
- Romina Del Bosque
- Department of Biomedical Engineering, Texas A&M University, College Station, Texas, USA
| | - Jiaming Cui
- Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| | - Stephen Ogier
- Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| | - Sergey Cheshkov
- Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA.,Department of Radiology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.,Center for Brain Health, University of Texas at Dallas, Dallas, Texas, USA
| | - Ivan E Dimitrov
- Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA.,Philips Healthcare, Gainesville, Florida, USA
| | - Craig Malloy
- Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA.,Department of Radiology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.,Department of Internal Medicine, University of Texas Southwestern Medical Center, Dallas, Texas, USA
| | - Steven M Wright
- Department of Biomedical Engineering, Texas A&M University, College Station, Texas, USA.,Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| | - Mary McDougall
- Department of Biomedical Engineering, Texas A&M University, College Station, Texas, USA.,Department of Electrical and Computer Engineering, Texas A&M University, College Station, Texas, USA
| |
Collapse
|
7
|
Chhetri A, Li X, Rispoli JV. Current and Emerging Magnetic Resonance-Based Techniques for Breast Cancer. Front Med (Lausanne) 2020; 7:175. [PMID: 32478083 PMCID: PMC7235971 DOI: 10.3389/fmed.2020.00175] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/15/2019] [Accepted: 04/15/2020] [Indexed: 01/10/2023] Open
Abstract
Breast cancer is the most commonly diagnosed cancer among women worldwide, and early detection remains a principal factor for improved patient outcomes and reduced mortality. Clinically, magnetic resonance imaging (MRI) techniques are routinely used in determining benign and malignant tumor phenotypes and for monitoring treatment outcomes. Static MRI techniques enable superior structural contrast between adipose and fibroglandular tissues, while dynamic MRI techniques can elucidate functional characteristics of malignant tumors. The preferred clinical procedure-dynamic contrast-enhanced MRI-illuminates the hypervascularity of breast tumors through a gadolinium-based contrast agent; however, accumulation of the potentially toxic contrast agent remains a major limitation of the technique, propelling MRI research toward finding an alternative, noninvasive method. Three such techniques are magnetic resonance spectroscopy, chemical exchange saturation transfer, and non-contrast diffusion weighted imaging. These methods shed light on underlying chemical composition, provide snapshots of tissue metabolism, and more pronouncedly characterize microstructural heterogeneity. This review article outlines the present state of clinical MRI for breast cancer and examines several research techniques that demonstrate capacity for clinical translation. Ultimately, multi-parametric MRI-incorporating one or more of these emerging methods-presently holds the best potential to afford improved specificity and deliver excellent accuracy to clinics for the prediction, detection, and monitoring of breast cancer.
Collapse
Affiliation(s)
- Apekshya Chhetri
- Magnetic Resonance Biomedical Engineering Laboratory, Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, United States
- Basic Medical Sciences, College of Veterinary Medicine, Purdue University, West Lafayette, IN, United States
| | - Xin Li
- Magnetic Resonance Biomedical Engineering Laboratory, Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, United States
| | - Joseph V. Rispoli
- Magnetic Resonance Biomedical Engineering Laboratory, Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN, United States
- Center for Cancer Research, Purdue University, West Lafayette, IN, United States
- School of Electrical & Computer Engineering, Purdue University, West Lafayette, IN, United States
| |
Collapse
|
8
|
Yeh JNT, Lin JFL. A Flexible and Modular Receiver Coil Array for Magnetic Resonance Imaging. IEEE TRANSACTIONS ON MEDICAL IMAGING 2019; 38:824-833. [PMID: 30295617 DOI: 10.1109/tmi.2018.2873317] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/08/2023]
Abstract
We propose a flexible form-fittingMRI receiver coil array assembledby individualcoilmodules. This design targetsMRI applications requiring a receiver array conforming to the anatomy of various shapes or sizes. Coil modules in our proposed array were arranged with gaps between them. Each coil module had a circumferential shielding structure stacked on top of the coil. Together they achieve robust decoupling when the array was bent differently. Two types of the circumferential shielding structure were investigatedby using full-wave electromagnetic simulations and imaging experiments. Results showed that our flexible coil array had good decoupling between coils whether they were on a flat or curved surface with the S21 magnitude ranged between -18.1 dB and -19.9 dB in simulations, and with the average of off-diagonal entries of the noise correlationmatrix less than 0.047 in experimentalmeasurements. Anatomical images of human brain, calf, and knee were acquired by our seven-channel prototype on a 3T MRI system. The maximal and the average SNR within 50 mm from our array surpassed those from the commercial 32-channel head and 4-channel flexible coil arrays by 2.63/1.35-fold and 3.89/1.50-fold, respectively.
Collapse
|
9
|
Krikken E, Steensma BR, Voogt IJ, Luijten PR, Klomp DW, Raaijmakers AJ, Wijnen JP. Homogeneous B 1+ for bilateral breast imaging at 7 T using a five dipole transmit array merged with a high density receive loop array. NMR IN BIOMEDICINE 2019; 32:e4039. [PMID: 30489661 PMCID: PMC6587506 DOI: 10.1002/nbm.4039] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 06/26/2018] [Revised: 09/24/2018] [Accepted: 10/12/2018] [Indexed: 06/09/2023]
Abstract
To explore the use of five meandering dipole antennas in a multi-transmit setup, combined with a high density receive array for breast imaging at 7 T for improved penetration depth and more homogeneous B1 field. Five meandering dipole antennas and 30 receiver loops were positioned on two cups around the breasts. Finite difference time domain simulations were performed to evaluate RF safety limits of the transmit setup. Scattering parameters of the transmit setup and coupling between the antennas and the detuned loops were measured. In vivo parallel imaging performance was investigated for various acceleration factors. After RF shimming, a B1 map, a T1 -weighted image, and a T2 -weighted image were acquired to assess B1 efficiency, uniformity in contrast weighting, and imaging performance in clinical applications. The maximum achievable local SAR10g value was 7.0 W/kg for 5 × 1 W accepted power. The dipoles were tuned and matched to a maximum reflection of -11.8 dB, and a maximum inter-element coupling of -14.2 dB. The maximum coupling between the antennas and the receive loops was -18.2 dB and the mean noise correlation for the 30 receive loops 7.83 ± 8.69%. In vivo measurements showed an increased field of view, which reached to the axilla, and a high transmit efficiency. This coil enabled the acquisition of T1 -weighted images with a high spatial resolution of 0.7 mm3 isotropic and T2 -weighted spin echo images with uniformly weighted contrast.
Collapse
Affiliation(s)
- Erwin Krikken
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| | - Bart R. Steensma
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| | - Ingmar J. Voogt
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| | - Peter R. Luijten
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| | - Dennis W.J. Klomp
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| | - Alexander J.E. Raaijmakers
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
- Department of Biomedical EngineeringEindhoven University of TechnologyEindhovenThe Netherlands
| | - Jannie P. Wijnen
- Department of RadiologyUniversity Medical Center UtrechtUtrechtThe Netherlands
| |
Collapse
|
10
|
Li X, Rispoli JV. Toward 7T breast MRI clinical study: safety assessment using simulation of heterogeneous breast models in RF exposure. Magn Reson Med 2018; 81:1307-1321. [PMID: 30216530 DOI: 10.1002/mrm.27395] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/21/2018] [Revised: 04/19/2018] [Accepted: 05/17/2018] [Indexed: 12/27/2022]
Abstract
PURPOSE To facilitate assessment of RF power deposition and temperature rise within the breast, we present a method to seamlessly join heterogeneous breast models with standard whole-body models and demonstrate simulations at 7 T. METHODS Finite-difference time-domain electromagnetic and bioheat simulations are performed to analyze the specific absorption rate (SAR) and temperature rise distributions in 36 Breast Imaging Reporting and Data System (BI-RADS) categorized breast models fused to 2 female whole-body models while transmitting from a 7T breast volume coil. The breast models are uncompressed in the prone position and feature heterogeneous tissue contents; fusion with human models uses affine transformation and the level-set method. RESULTS The fusion method produces a continuous transient from the chest region to the posterior portion of breast models while preserving the original volume and shape of breast models. Simulation results of both Ella and Hanako models indicate that the maximum local SAR, partial body SAR, and local tissue temperature rise are positively correlated with both breast density and the highest BI-RADS density classification. Additionally, maximum local tissue temperature rise is positively correlated with maximum 10-g SAR values. CONCLUSION Fibroglandular tissue content plays an important role in the distribution of SAR and temperature rise within breast tissue. The combined body-breast models preserve the integrity of breast models while concurrently exhibiting the loading of whole-body human models. The procedures presented in this simulation study facilitate safety assessments for breast MRI across the population at both clinical and ultrahigh field strengths.
Collapse
Affiliation(s)
- Xin Li
- Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana
| | - Joseph V Rispoli
- Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana.,School of Electrical and Computer Engineering, Purdue University, West Lafayette, Indiana.,Purdue University Center for Cancer Research, Purdue University, West Lafayette, Indiana
| |
Collapse
|
11
|
Cui J, Dimitrov IE, Cheshkov S, Gu M, Malloy CR, Wright SM. An Adjustable-Length Dipole Using Forced-Current Excitation for 7T MR. IEEE Trans Biomed Eng 2018; 65:2259-2266. [PMID: 29989961 DOI: 10.1109/tbme.2017.2788864] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
Ultrahigh field imaging of the body and the spine is challenging due to the large field-of-view (FOV) required. It is especially difficult for RF transmission due to its requirement on both the length and the depth of the ${\rm{B}}_{1}^{{\rm + }}$ field. One solution is to use a long dipole to provide continuous current distribution. The drawback is the natural falloff of the ${\rm{B}}_{1}$ field toward the ends of the dipole, therefore the ${\rm{B}}_{1}^{{\rm + }}$ per unit square root of maximum specific absorption rate ${\rm{(B}}_{1}^{{\rm + }}{\rm{/ \surd SAR}}_{{\rm{max}}})$ performance is particularly poor toward the end of the dipole. In this study, a segmented element design using forced-current excitation and a switching circuit is presented. The design provides long FOV when desired and allows flexible FOV switching and power distribution without additional power amplifiers. Different element types and arrangements were explored and a segmented dipole design was chosen as the best design. The segmented dipole was implemented and tested on the bench and with a phantom on a 7T whole body scanner. The switchable mode dipole enabled a large FOV in the long mode and improved ${\rm{B}}_{1}^{{\rm + }}{\rm{/ \surd SAR}}_{{\rm{max}}}$ efficiency in a smaller FOV in the short mode.
Collapse
|
12
|
Kim J, Krishnamurthy N, Santini T, Zhao Y, Zhao T, Bae KT, Ibrahim TS. Experimental and numerical analysis of B1(+) field and SAR with a new transmit array design for 7T breast MRI. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2016; 269:55-64. [PMID: 27240143 PMCID: PMC4979605 DOI: 10.1016/j.jmr.2016.04.012] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/15/2016] [Revised: 04/20/2016] [Accepted: 04/21/2016] [Indexed: 05/15/2023]
Abstract
Developing a radiofrequency (RF) coil system that produces a uniform B1(+) field (circularly polarized component of the transverse magnetic field responsible for excitation) and low specific absorption rate (SAR) is critical for high performance ultrahigh field human imaging. In this study, we provide the design of a new eight channel radiofrequency (RF) transmit (Tx) array for breast MRI at 7T. A numerical analysis utilizing an in-house finite difference time domain (FDTD) package was carried out in (1) four breast models, (2) homogeneous spherical model and (3) full body model to calculate the B1(+) intensity (μT) and homogeneity represented by coefficient of variation (CoV=standard deviation/mean) in the proposed RF array design. The numerical results were compared with that measured in breast phantom (Bphantom) and homogeneous spherical phantom at 7T MRI and showed very good agreement. Average and peak SARs were also calculated in the four breast models and the temperature rises due to the operation of the RF array were also measured in the Bphantom. The proposed RF array; which can operate in a single or multi transmit modes, demonstrates homogeneous RF field excitation with acceptable local/average SAR levels for breast MRI at 7T.
Collapse
Affiliation(s)
- Junghwan Kim
- Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15261, USA
| | | | - Tales Santini
- Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15261, USA
| | - Yujuan Zhao
- Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15261, USA
| | - Tiejun Zhao
- MR Research Support, Siemens Healthcare, Pittsburgh, PA 15213, USA
| | - Kyongtae Ty Bae
- Department of Radiology, University of Pittsburgh, Pittsburgh, PA 15213, USA; Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15261, USA
| | - Tamer S Ibrahim
- Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15261, USA; Department of Radiology, University of Pittsburgh, Pittsburgh, PA 15213, USA.
| |
Collapse
|