McNeil DS, Trenholm JAI. Establishing Safety Parameters for Orthopaedic Cast Saw Blade Usage.
J Pediatr Orthop 2021;
41:e884-e888. [PMID:
34516468 DOI:
10.1097/bpo.0000000000001928]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 02/07/2023]
Abstract
BACKGROUND
The incidence of injuries from cast saws during cast removal ranges from 0.12% to 4.3%. With 1 second or less of exposure time, a temperature of 65°C can cause partial thickness burns. Despite numerous studies that recommend avoiding the use of a dull blade, there is no objective measure of what defines dullness.
METHODS
Plaster and fiberglass casts were collected and measured after removal from patients in the clinic. A series of slabs were constructed based on these measurements. To simulate our emergency department setting, a Stryker 940 cast saw without an attached vacuum was used to split plaster slabs. A thermocouple was used to directly measure the 940-23 ion-nitride saw blade temperature after each use. To simulate our orthopaedic clinic setting, a Stryker 940 cast saw with an attached vacuum was used to split fiberglass and plaster slabs. Three blades were tested in each setting, bivalving 50 slabs each.
RESULTS
For the plaster slabs split without a vacuum, average blade temperature of the 3 blades reached 65°C on the 42nd cast. However, the individual blades exceeded 65°C on the 33rd, 31st, and 38th casts, respectively. For the fiberglass and plaster slabs split with a vacuum, average blade temperature reached a maximum of 57.5°C in the first 50 trials. Extrapolating from this data, the blade is predicted to exceed 65°C on the 104th cast.
CONCLUSIONS
When a Stryker 940 cast saw without vacuum is used to cut plaster casts, the ion-nitride blade should be changed frequently, at minimum after 60 casts have been split, or 30 casts have been bivalved. When a Stryker 940 cast saw with vacuum is used to remove fiberglass and plaster casts, the ion-nitride blade should be changed after removing 103 casts. A cast saw with an attached vacuum should be used whenever possible to minimize the risk of burning patients.
CLINICAL RELEVANCE
Determine how often a cast saw blade should be changed to minimize risk of burning patients.
Collapse