Gutmann S, Heiderhoff M, Möbius R, Siegel T, Flegel T. Application accuracy of a frameless optical neuronavigation system as a guide for craniotomies in dogs.
Acta Vet Scand 2023;
65:54. [PMID:
38098105 PMCID:
PMC10722823 DOI:
10.1186/s13028-023-00720-y]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/07/2023] [Accepted: 12/08/2023] [Indexed: 12/17/2023] Open
Abstract
BACKGROUND
Optical neuronavigation systems using infrared light to create a virtual reality image of the brain allow the surgeon to track instruments in real time. Due to the high vulnerability of the brain, neurosurgical interventions must be performed with a high precision. The aim of the experimental cadaveric study was to determine the application accuracy of a frameless optical neuronavigation system as guide for craniotomies by determining the target point deviation of predefined target points at the skull surface in the area of access to the cerebrum, cerebellum and the pituitary fossa. On each of the five canine cadaver heads ten target points were marked in a preoperative computed tomography (CT) scan. These target points were found on the cadaver skulls using the optical neuronavigation system. Then a small drill hole (1.5 mm) was drilled at these points. Subsequently, another CT scan was made. Both CT data sets were fused into the neuronavigation software, and the actual target point coordinates were identified. The target point deviation was determined as the difference between the planned and drilled target point coordinates. The calculated deviation was compared between two observers.
RESULTS
The analysis of the target point accuracies of all dogs in both observers taken together showed a median target point deviation of 1.57 mm (range: 0.42 to 5.14 mm). No significant differences were found between the observers or the different areas of target regions.
CONCLUSION
The application accuracy of the described system is similar to the accuracy of other optical neuronavigation systems previously described in veterinary medicine, in which mean values of 1.79 to 4.3 mm and median target point deviations of 0.79 to 3.53 mm were determined.
Collapse