Yalcın SO, Kaplan AT. Comparison of central corneal thickness measurements with different non-contact devices in healthy children.
Photodiagnosis Photodyn Ther 2022;
40:103045. [PMID:
35908681 DOI:
10.1016/j.pdpdt.2022.103045]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/31/2022] [Revised: 07/17/2022] [Accepted: 07/28/2022] [Indexed: 12/14/2022]
Abstract
PURPOSE
To investigate the agreement of the Combined Scheimpflug-Placido Disk System (Sirius), specular microscopy (Tomey EM-4000), non-contact tono-pachymetry (Canon TX-20P), and optical low-coherence reflectometry (Lenstar LS 900) devices in the measurement of central corneal thickness (CCT) in healthy children.
METHOD
Eighty-three eyes from 83 healthy children were included in the study. The mean age of the children was 11.1 ± 2.8 (range, 7-17) years. CCT was measured using the Sirius topography, Tomey specular microscopy, Canon TX-20P, and Lenstar LS 900 devices. To examine the agreement among these devices, 95% limits of agreement (LoA) and intraclass correlation coefficients (ICC) were determined.
RESULTS
The mean CCT values obtained using the Sirius topography, Tomey SM, Canon TX-20 P, and Lenstar 900 devices were 555.9 ± 37 µm, 553.5 ± 39 µm, 569.3 ± 33 µm, and 559.7 ± 38 µm, respectively. With the ICC method, the agreement rates were determined as 0.875 between Canon TX-20P and Sirius topography, 0.903 between Canon TX-20P and Lenstar LS 900, 0.842 between Canon TX-20P and Tomey EM-4000, 0.979 between Sirius Topography and Lenstar LS 900, 0.978 between Sirius topography and Tomey EM-4000, and 0.969 between Lenstar LS 900 and Tomey EM-4000.
CONCLUSION
The devices with the closest results to each other were the Sirius and Tomey EM-4000 devices, with a statistically strong agreement. There was also a strong agreement between Sirius and Lenstar LS 900, although there was a 3.7-unit difference between them. Nevertheless, despite the strong agreements among these devices, because their CCT values varied statistically significantly, these devices should not be used interchangeably in clinical practice.