Hirata M, Hanaoka K, Matsumoto K, Tamura M, Okumura M, Monzen H. The impact of the offset distance between the planning target volume and isocenter on irradiation time in TomoTherapy: A phantom study.
Phys Med 2019;
63:19-24. [PMID:
31221404 DOI:
10.1016/j.ejmp.2019.05.009]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [What about the content of this article? (0)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/25/2018] [Revised: 04/19/2019] [Accepted: 05/18/2019] [Indexed: 11/23/2022] Open
Abstract
PURPOSE
The influence of the offset distance from treatment target to gantry isocenter (GIC) on the dosimetric parameter and irradiation time was investigated using TomoTherapy METHODS: The reference position was defined as the centers of both the I'mRT phantom and planning target volume (PTV) with a spherical of 4 cm diameter aligned with the GIC. The dose calculations were performed in two offset methods with 2 and 12 Gy/fr, Method 1. The PTV was moved from 0.0 to 12.5 cm along the RL direction and -5.0 to 5.0 cm along the AP direction (PTV offset), Method 2. The phantom was moved from 0.0 to -7.5 cm along the RL direction and -5.0 to 5.0 cm along the AP direction (Phantom offset). The maximum, minimum and mean doses, homogeneity index, conformity index, irradiation time, and monitor unit were compared.
RESULTS
The irradiation times increased with increasing PTV offset. The increases in the irradiation time were 54.4% and 40.8% at PTV offsets of 12.5 cm along the RL direction for 2 and 12 Gy/fr, while the increases were 20.1% and 15.0% at a PTV offset of 5.0 cm along the AP direction. An increased irradiation time was not observed for the phantom offset. The offset didn't affect the other parameters.
CONCLUSIONS
The PTV location offset of ≥5 cm from the GIC along the RL and AP axes increased the irradiation time; therefore, the PTV should be aligned with the GIC as much as possible to reduce the irradiation time on TomoTherapy.