Kaakeh Y, Phan H, DeSmet BD, Pasko DA, Glenn DK, Stevenson JG. Enhanced photoemission spectroscopy for verification of high-risk i.v. medications.
Am J Health Syst Pharm 2008;
65:49-54. [PMID:
18159039 DOI:
10.2146/ajhp060626]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/23/2022] Open
Abstract
PURPOSE
The sensitivity and specificity of enhanced photoemission spectroscopy (EPS) for performing an automated final check of compounded i.v. admixtures at a pediatric hospital pharmacy were studied.
METHODS
A tabletop EPS device was used to test samples of seven high-risk drug-diluent combinations compounded in the pharmacy; the drugs were vancomycin, lorazepam, morphine, insulin, hydromorphone, gentamicin, and epinephrine. Ten sets of samples were prepared for each drug. Typically, a sample set consisted of dilutions ranging from 10-fold above to 10-fold below the targeted concentration. Testing was performed twice weekly between November 2005 and March 2006.
RESULTS
The EPS device detected errors departing from the targeted concentration by 20% or more with a sensitivity of at least 95%. Specificity in distinguishing among test medications at targeted concentrations was 100%. The percentage of passing samples with intermediate concentrations varied among the drugs.
CONCLUSION
A tabletop EPS device demonstrated acceptable sensitivity and specificity for validating the identity and concentrations of selected high-risk i.v. medications compounded for pediatric patients. The device may help prevent clinically important medication errors caused by inaccurate compounding.