Van Hecke I, Vanden Bempt I, Malfait R, Van Den Bossche J. Evaluation of the Sysmex pocH-1001 haematology analyser in an outdoor oncology service.
Acta Clin Belg 2010;
65:248-51. [PMID:
20954464 DOI:
10.1179/acb.2010.054]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/31/2022]
Abstract
Since rapid blood count analysis as near patient testing is expanding, we evaluated the use of a Sysmex pocH-100i compact haematology analyser in an outdoor oncology setting according to the recently published International Council for Standardization in Haematology (ICSH) guidelines. In total, 838 blood samples from oncology patients were analysed by pocH-100i and re-analysed by a high-throughput haematology analyser for comparison (Abbott CD-4000 or Sysmex XE-2100) to evaluate in use imprecision, comparability and vote-outs. Imprecision was less than 5%, except for platelet enumeration in the low range (within-run imprecision 7%). Good comparability was found even for platelet enumeration in the low range (r2 = 0.82). Vote-outs were found in 10.6% of examined samples. In conclusion, the Sysmex pocH-100i demonstrates good imprecision conform with former publications, produces reliable results in normal and in lower ranges comparable to the results of high throughput haematology analysers. In a well controlled management plan the Sysmex pocH-100i is suitable for near patient testing in oncology.