1
|
Wachter A, Bahlke F, Popa M, Foerschner L, Krafft H, Maurer S, Telishevska M, Englert F, Lengauer S, Kottmaier M, Lennerz C, Reents T, Hessling G, Deisenhofer I, Bourier F. Correlation of local impedance, global impedance and contact force with resulting lesion size in RF ablation. Europace 2022. [DOI: 10.1093/europace/euac053.108] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
During clinical application of RF-energy, several parameters are constantly observed, e.g. RF time, power, global impedance and temperature. The new parameter "local impedance" is gaining importance and might be a possible real-time marker for predicting long-lasting RF-lesions and increasing safety. The aim of this study was to investigate the correlation between local and global impedance as well as RF power and contact force.
Methods
RF-lesions were created using an ex vivo model with porcine cardiac preparations. These were put in a saline-filled container with a dispersive electrode. Global impedance was held at 120 Ohm by adjusting saline and water. Additionally, a heated thermostat and a circulation pump were installed to imitate blood flow. RF-lesions were produced with different settings of energy and contact force (20, 30, 40 and 50 watts; contact force of 0-5g, 10-15g and 20-25g). While creating the lesions, global and local impedance, temperature, energy and RF time were documented, as well as the current lesion width and depth. Local and global impedance drops were calculated as the difference between baseline impedance and current impedance.
Results
In total, 1223 measurements were made during application of RF-energy. 6 steam pops occurred. Contact force, local and global impedance changes showed highly significant correlations with lesion depth and diameter. Amongst analyzed values, local impedance drop showed highest correlation with lesion diameter and depth (r = 0,391 and 0,613; p<0,001). Visualized in Figure 1 and 2, global and local impedance correlate with lesion diameter and depth.
Discussion
Local impedance was found to be a suitable real-time marker for lesion size, showing significantly higher correlations with RF lesion size than contact force and global impedance. Further investigation is necessary to find a cut off for a safe but long-lasting RF-lesion.
Collapse
Affiliation(s)
- A Wachter
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Bahlke
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Popa
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - L Foerschner
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - H Krafft
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - S Maurer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Telishevska
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Englert
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - S Lengauer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Kottmaier
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - C Lennerz
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - T Reents
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - G Hessling
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - I Deisenhofer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Bourier
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| |
Collapse
|
2
|
Wachter A, Bahlke F, Popa M, Foerschner L, Krafft H, Maurer S, Englert F, Telishevska M, Lengauer S, Kottmaier M, Lennerz C, Reents T, Hessling G, Deisenhofer I, Bourier F. Electrode tissue coupling is a new main determinant of RF lesion creation, experience from a cardiac animal model. Europace 2022. [DOI: 10.1093/europace/euac053.109] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/14/2022] Open
Abstract
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
Radiofrequency (RF) current is a classic ablation technology used in the EP lab. By now it is assumed that RF lesion size is mainly determined by RF power, duration and contact force. This ex vivo animal model study shows that a new parameter - "electrode tissue coupling" (ETC) - plays an additional major role in RF lesion creation. The ETC level describes the amount of electric contact between the RF electrode and cardiac tissue: In minor ETC levels, only the distal electrode connects to tissue. In full ETC levels, the whole RF electrode is embedded in cardiac tissue.
Methods
RF-lesions were created using an ex vivo porcine cardiac model. The experimental setup consisted of a saline-filled container, a dispersive electrode, a heated thermostat and a circulation pump to imitate in vivo conditions. Global impedance was kept at 120 Ohm as well as the temperature at 37°C. RF-lesions were created using identic values of RF duration and contact force. A RF power of 20W, 30W, 40W, and 50W was used. The ETC levels were systematically varied between minor and full coupling. All parameters (power, temperature, global and local impedance, contact force, ETC, lesion size) were measured constantly during application of RF-current, enabling real-time correlation of RF parameters and lesion size.
Results
In total, 1923 measurements during application of RF-current were analyzed. In ETC III (full tissue coupling), lesions became significantly wider and deeper. In 20W ablations, lesion diameters were significantly (1.68 fold) larger when applying ETC III instead of ETC I. This relation was found in variations of RF parameters (Table 1).
Interestingly, baseline local impedance and local impedance drop showed a high correlation with selected ETC levels. The average baseline local impedance in ETC I was 207.2 Ohm, compared to 267.3 Ohm in ETC III (p < 0.01).
Discussion
In addition to by now known parameters (power, duration, contact force), electrode tissue coupling is a main determinant of lesion size. Higher ETC levels result in higher amounts of RF current going into adjacent tissue instead of current dissipation into the blood pool. In clinical practice, the ETC level can be predicted by baseline local and global impedance. Observation of these parameters should become clinical practice during RF ablation.
Collapse
Affiliation(s)
- A Wachter
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Bahlke
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Popa
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - L Foerschner
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - H Krafft
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - S Maurer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Englert
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Telishevska
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - S Lengauer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - M Kottmaier
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - C Lennerz
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - T Reents
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - G Hessling
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - I Deisenhofer
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| | - F Bourier
- German Heart Center Munich, Technical University of Munich, Munich, Germany
| |
Collapse
|
3
|
Knoll K, Oconnor M, Chouchane A, Haller B, Schaarschmidt C, Bock M, Foerschner L, Froehlich R, Kottmaier M, Bourier F, Reents T, Hessling G, Deisenhofer I, Kolb C, Lennerz C. Procalcitonin as a biomarker of cardiac implantable electronic device pocket infection: a prospective validation study. Europace 2022. [DOI: 10.1093/europace/euac053.542] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/13/2022] Open
Abstract
Abstract
Funding Acknowledgements
Type of funding sources: None.
Introduction
The implantation of cardiac implantable electronic devices (CIEDs) such as pacemakers and implantable cardioverter-defibrillators is increasing along with the complexitly of these devices. CIED infection is an uncommon, but severe complication associated with the presence of a device and is associated with a high mortality and morbidity. Lead-related infections and frank endocarditis are associated with a systemic inflammatory response and, in general, are readily identified. Isolated pocket infections do not produce such a systemic response and are thus more complex to diagnose. There is a reliance on clinical accumen and examination of local signs of infection. There is thus a need for a reliable biomaker to help identify cases of pocket infection.
Aim
Our group have previously shown procalcitonin (PCT) to be a potentially useful biomaker in the clinial situation of possible pocket infection. We aim to prospectively validate the proposed cut-off value of 0.05ng/ml for the procalcitonin (PCT) biomaker in an independent cohort, which we have previously identified as showing promise in this clinical situation.
Methods
In this prospective case-control validation study the PCT levels of 81 patients with confirmed pocket infections were compared to 81 controls, matched for age and renal function, presenting for elective generator replacement or lead revision unrelated to infection. Exclusion criteria included: concomitant infectious or inflammatory diseases, end-stage renal failure, active malignancy or receiving immunosuppressive therapy.
Results
A PCT over 0.05 ng/ml was found in 68% (n= 55) of pocket infections and 24% (n= 19) of controls. Using the predefined cut-off value of 0.05 ng/ml PCT had a sensitivity of 68% and a specificity of 77% for diagnosing pocket infections. ROC analysis revealed area under the curve of 0.752 (standard error 0.039, p <0.001 ) for PCT. In patients presenting with minimal infective signs the sensivity remained high (67% vs 70% with extensive inflammation) and similarly remained high in thus who had received anti-biotic therapy prior to PCT sampling (65% vs 69%).
Conclusion
PCT is a potentially useful biomarker to aid the diagnosis of a pocket infection when used with the prospecitvely validated cut-off value of 0.05ng/ml. The sensitivity of the PCT positive result remained high even in patients pre-treated with antibiotics or with minimal clinical signs of inflammation.
Collapse
Affiliation(s)
- K Knoll
- German Heart Centre Munich, Munich, Germany
| | - M Oconnor
- Royal Brompton and Harefield Hospital, London, United Kingdom of Great Britain & Northern Ireland
| | | | - B Haller
- Clinic rechts der Isar of the University of Technology, Institut für KI und Informatik in der Medizin, Munich, Germany
| | | | - M Bock
- German Heart Centre Munich, Munich, Germany
| | | | | | | | - F Bourier
- German Heart Centre Munich, Munich, Germany
| | - T Reents
- German Heart Centre Munich, Munich, Germany
| | - G Hessling
- German Heart Centre Munich, Munich, Germany
| | | | - C Kolb
- German Heart Centre Munich, Munich, Germany
| | - C Lennerz
- German Heart Centre Munich, Munich, Germany
| |
Collapse
|