Lloyd MA, Baerveldt G, Fellenbaum PS, Sidoti PA, Minckler DS, Martone JF, LaBree L, Heuer DK. Intermediate-term results of a randomized clinical trial of the 350- versus the 500-mm2 Baerveldt implant.
Ophthalmology 1994;
101:1456-63; discussion 1463-4. [PMID:
8058290 DOI:
10.1016/s0161-6420(94)31152-3]
[Citation(s) in RCA: 114] [Impact Index Per Article: 3.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/28/2023] Open
Abstract
BACKGROUND
The Baerveldt glaucoma implant is a large equatorial aqueous shunting device that is installed through a single-quadrant conjunctival incision. The intermediate-term results of a randomized study comparing the 350- and 500-mm2 Baerveldt implants are reported.
METHODS
Seventy-three patients with medically uncontrollable, nonneovascular glaucomas associated with aphakia, pseudophakia, or failed filters were enrolled in a randomized, prospective study comparing 350- and 500-mm2 Baerveldt implants. Surgical success was defined as 6 mmHg < or = final intraocular pressure < or = 21 mmHg without glaucoma reoperation or devastating complication.
RESULTS
Of patients with 350- and 500-mm2 implants, 93% and 88%, respectively, achieved surgical success (18-month life-table analysis, P = 0.93). The 500-mm2 implants afforded intraocular pressure control with significantly fewer medications (0.7 versus 1.3; P = 0.006). The postoperative visual acuities remained within one line of the preoperative visual acuities or improved in 62% and 66% of patients in the 350- and 500-mm2 groups, respectively (P = 0.93). Complication rates were statistically similar. The most frequent ones in the 350- and 500-mm2 groups, respectively, were serous choroidal effusion (16% and 32%), strabismus (16% and 19%), anterior uveitis (14% and 11%), and corneal or corneal graft edema (11% each).
CONCLUSION
The intermediate-term results of the 350- and 500-mm2 Baerveldt implants were statistically comparable with respect to surgical and visual outcomes, as well as complications, although the larger implant was associated with a higher rate of some complications. However, the 500-mm2 Baerveldt implant afforded intraocular pressure control with fewer medications than the 350-mm2 implant.
Collapse