Pathak PK, Vashisht SK, Baby S, Jithin PK, Jain Y, Mahawar R, Sharan VGGK. Commissioning and quality assurance of Halcyon
TM 2.0 linear accelerator.
Rep Pract Oncol Radiother 2021;
26:433-444. [PMID:
34277097 PMCID:
PMC8281907 DOI:
10.5603/rpor.a2021.0065]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/20/2020] [Accepted: 02/25/2021] [Indexed: 12/02/2022] Open
Abstract
BACKGROUND
Varian Medical Systems has introduced a new medical linear accelerator called HalcyonTM 2.0, which is based on the ring delivery system (RDS). It is a true IGRT machine having 6MV FFF photon energy. In addition to the planar and MV-CBCT imaging techniques it also has an option of ultra-fast kV-iCBCT which enhances the image reconstruction and improves the visualization of soft tissue. The field portals are shaped by a unique dual layer MLC with special stacked and staggered design which enables high modulation with low radiation leakage. Recently, we have commissioned our first Halcyon 2.0 machine. The aim of this work was to systematically investigate various parameters of a newly installed HalcyonTM 2.0 linear accelerator.
MATERIALS AND METHODS
Detailed measurements were conducted as per various guidelines. Also, the measurements were performed to fulfil the national regulatory requirements. Commissioning data of Halcyon 6 MV-FFF beam was performed in a water tank. For absolute measurements, a 0.6-cc waterproof Farmer chamber and electrometer were used. All relative measurements (PDDs, in-line, cross-line and angular profiles) were performed with 0.0125 cc point chamber.
RESULTS
All the tests were within the acceptable limit. Measured data were compared with factory data as well as the existing medical linear accelerator of the same category. The obtained results were quite satisfactory.
CONCLUSIONS
This study summarizes the commissioning experience with Halcyon linear accelerator. Evaluation of mechanical, radiation safety and dosimetric parameters were performed. The obtained parameters were well below the specified tolerance limits.