EUS-guided portal pressure gradient measurement with a simple novel device: a human pilot study.
VIDEOGIE : AN OFFICIAL VIDEO JOURNAL OF THE AMERICAN SOCIETY FOR GASTROINTESTINAL ENDOSCOPY 2018;
3:361-363. [PMID:
30402586 PMCID:
PMC6205538 DOI:
10.1016/j.vgie.2018.07.013]
[Citation(s) in RCA: 12] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 04/21/2023]
Abstract
BACKGROUND AND AIMS
Portal hypertension is a serious adverse event of liver cirrhosis. Recently, we developed a simple novel technique for EUS-guided portal pressure gradient (PPG) measurement (PPGM). Our animal studies showed excellent correlation between EUS-PPGM and interventional radiology-acquired PPGM. In this video we demonstrate the results of the first human pilot study of EUS-PPGM in patients with liver disease.
METHODS
EUS-PPGM was performed by experienced endosonographers using a linear echoendoscope, a 25-gauge FNA needle, and a novel compact manometer. The portal vein and hepatic vein (or inferior vena cava) were targeted by use of a transgastric or transduodenal approach. Feasibility was defined as successful PPGM in each patient. Safety was based on adverse events captured in a postprocedural interview.
RESULTS
Twenty-eight patients underwent EUS-PPGM with 100% technical success and no adverse events. PPG ranged from 1.5 to 19 mm Hg and had excellent correlation with clinical parameters of portal hypertension, including the presence of varices (P = .0002), PH gastropathy (P = .007), and thrombocytopenia (P = .036).
CONCLUSION
This novel technique of EUS-PPGM using a 25-gauge needle and compact manometer is feasible and appears safe. Given the availability of EUS and the simplicity of the manometry setup, EUS-guided PPG may represent a promising breakthrough for procuring indispensable information in the management of patients with liver disease.
Collapse