Ichikawa S, Sugimori H, Ichijiri K, Yoshimura T, Nagaki A.
Acquisition time reduction in pediatric
99m Tc-DMSA planar imaging using deep learning.
J Appl Clin Med Phys 2023:e13978. [PMID:
37021382 DOI:
10.1002/acm2.13978]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/25/2022] [Revised: 02/27/2023] [Accepted: 03/10/2023] [Indexed: 04/07/2023] Open
Abstract
PURPOSE
Given the potential risk of motion artifacts, acquisition time reduction is desirable in pediatric 99m Tc-dimercaptosuccinic acid (DMSA) scintigraphy. The aim of this study was to evaluate the performance of predicted full-acquisition-time images from short-acquisition-time pediatric 99m Tc-DMSA planar images with only 1/5th acquisition time using deep learning in terms of image quality and quantitative renal uptake measurement accuracy.
METHODS
One hundred and fifty-five cases that underwent pediatric 99m Tc-DMSA planar imaging as dynamic data for 10 min were retrospectively collected for the development of three deep learning models (DnCNN, Win5RB, and ResUnet), and the generation of full-time images from short-time images. We used the normalized mean squared error (NMSE), peak signal-to-noise ratio (PSNR), and structural similarity index metrics (SSIM) to evaluate the accuracy of the predicted full-time images. In addition, the renal uptake of 99m Tc-DMSA was calculated, and the difference in renal uptake from the reference full-time images was assessed using scatter plots with Pearson correlation and Bland-Altman plots.
RESULTS
The predicted full-time images from the deep learning models showed a significant improvement in image quality compared to the short-time images with respect to the reference full-time images. In particular, the predicted full-time images obtained by ResUnet showed the lowest NMSE (0.4 [0.4-0.5] %) and the highest PSNR (55.4 [54.7-56.1] dB) and SSIM (0.997 [0.995-0.997]). For renal uptake, an extremely high correlation was achieved in all short-time and three predicted full-time images (R2 > 0.999 for all). The Bland-Altman plots showed the lowest bias (-0.10) of renal uptake in ResUnet, while short-time images showed the lowest variance (95% confidence interval: -0.14, 0.45) of renal uptake.
CONCLUSIONS
Our proposed method is capable of producing images that are comparable to the original full-acquisition-time images, allowing for a reduction of acquisition time/injected dose in pediatric 99m Tc-DMSA planar imaging.
Collapse