Lee CN, Wu CK, Huang IC. Validation of the AViTA BPM636 upper arm blood pressure monitor in adults and pregnant women according to the
ANSI/AAMI/ISO 81060-2:2013.
Blood Press Monit 2023;
28:215-220. [PMID:
37074406 PMCID:
PMC10309106 DOI:
10.1097/mbp.0000000000000648]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/30/2022] [Accepted: 04/04/2023] [Indexed: 04/20/2023]
Abstract
OBJECTIVE
To evaluate the accuracy of the AViTA oscillometric upper arm home blood pressure (BP) monitor in adult and pregnant populations according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/ International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2013).
METHODS
BP measurements on the upper arm were performed on 85 adult subjects and 46 pregnant subjects. The AViTA BPM636 and a standard mercury reference sphygmomanometer were applied and followed the same arm sequential BP measurement method. The universal cuff of the test device was used for arm circumference of 22-42 cm.
RESULTS
For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 1.1 ± 5.49/2.9 ± 5.17 mmHg (systolic/diastolic) for adults; and -2.2 ± 5.93/1.5 ± 4.92 mmHg (systolic/diastolic) for pregnant women. For criterion 2, the SD of the averaged BP differences between the test device and reference BP per adult subject was 4.45/4.20 mmHg (systolic/diastolic) and per pregnant women was 4.66/3.96.
CONCLUSION
The AViTA BPM636 had passed the criteria of the ANSI/AAMI/ISO 81060-2:2013 protocol and can be recommended for home BP measurements in adults and pregnant populations.