Pandey AK, Gautam RK, Behera CK. Microstructure, mechanical strength, chemical resistance, and antibacterial behavior of Ti-5Cu-
x%Nb biomedical alloy.
Biomed Mater 2022;
17. [PMID:
35679847 DOI:
10.1088/1748-605x/ac7763]
[Citation(s) in RCA: 2] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/08/2021] [Accepted: 06/09/2022] [Indexed: 01/01/2023]
Abstract
Titanium-based biomedical alloys are susceptible as they are used as a substitute for human bone. In this study, titanium alloy, Ti-5Cu-x%Nb (x= 0, 5, 10, 15) (%wt) was developed by powder metallurgy route. The effect of alloying niobium with Ti-5Cu alloy and its effect on the microstructure, mechanical strength, corrosion resistance, and antibacterial properties have been evaluated. The results show that the sintered alloy has bothα-Ti and Ti2Cu phases. With increasing niobium content in the alloy,β-Ti was also detected. Additionally, it was found that the micro-hardness and compressive strength of the studied alloy was better than commercially pure titanium (cpTi), while the Young's modulus was lower than cpTi. These properties are highly favorable for using this alloy to replicate the human cortical bone. The alloy was also tested for anticorrosive property in Ringer's solution. The antibacterial activity was also examined forStaphylococcus aureusandEscherichia colibacteria. The alloy showed promising anticorrosive and antibacterial ability.