1
|
Kurukuti NM, Nadeau M, Paschalis EI, Houston KE. An Adjustable Magnetic Levator Prosthesis for Customizable Eyelid Re-Animation in Severe Blepharoptosis: Design and Proof-of-Concept. Transl Vis Sci Technol 2023; 12:11. [PMID: 37566397 PMCID: PMC10424798 DOI: 10.1167/tvst.12.8.11] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/01/2023] [Accepted: 06/20/2023] [Indexed: 08/12/2023] Open
Abstract
Purpose Blepharoptosis is a common oculoplastic condition causing incomplete opening of the upper eyelid. Surgical approaches, the mainstay for correction, often fail to improve blink function. The purpose of this study was to develop a nonsurgical treatment option for severe ptosis that allows blink re-animation. Methods Magnetic force required to perform blink re-animation was characterized by evaluation of eye-opening and closing using inter-palpebral fissure (IPF) outcomes with various combinations of eyelid array and box magnets. Optimal size of the spectacle magnet that achieved forces required for optimal blink dynamics was selected using simulation. The adjustable magnetic levator prosthesis (aMLP) included an eyelid array magnet and an adjustable rotating spectacle magnet that allowed change in the magnetic direction, thus changing the net magnetic interactive force between the magnets. The clinical feasibility of aMLP in improving eye opening without limiting eye closing was evaluated in patients with ptosis through a proof-of-concept study using IPF and comfort outcomes. Results Optimal eye opening and closing was achieved by a magnet-array combination providing 45 grams of surface force (gF) in the tested ptosis population. The aMLP was able to modulate eye opening and closing with change in rotation of the spectacle magnet in two patients with ptotis. The best fitting of an aMLP improved IPF opening without limiting eye closing and with good comfort reported. Conclusions Preliminary results suggest that the an aMLP can correct ptosis without adversely affecting blink function. Further evaluation in a larger patient population is warranted. Translational Relevance A nonsurgical, proof of concept, adjustable magnetic treatment option for blink re-animation in patients with severe ptosis is presented.
Collapse
Affiliation(s)
- Nish Mohith Kurukuti
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
- Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, IL, USA
| | - Melanie Nadeau
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
| | - Eleftherios I. Paschalis
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
- The Boston Keratoprosthesis Laboratory, Department of Ophthalmology, Massachusetts Eye & Ear, Boston, MA, USA
| | - Kevin E. Houston
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
- University of Massachusetts Medical School, Department of Ophthalmology and Neurology, MA, USA
- Central Western Massachusetts Veterans Affairs, Departments of Optometry and Neurology, MA, USA
| |
Collapse
|
2
|
Houston KE, Tomasi M, Amaral C, Finch N, Yoon MK, Lee H, Paschalis EI. The Magnetic Levator Prosthesis for Temporary Management of Severe Blepharoptosis: Initial Safety and Efficacy. Transl Vis Sci Technol 2018; 7:7. [PMID: 29367892 PMCID: PMC5777174 DOI: 10.1167/tvst.7.1.7] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/21/2017] [Accepted: 11/03/2017] [Indexed: 11/24/2022] Open
Abstract
Purpose We further optimized and evaluated the safety of the magnetic levator prosthesis (MLP) for temporary management of severe blepharoptosis, and compared efficacy and comfort against the ptosis crutch. Methods The interpalpebral fissure (IPF) of participants (n = 12) with ptosis was measured during attempted eyelid opening, volitional closing, and spontaneous closing with no device, ptosis crutch, or the MLP. A 10-point scale documented comfort. Additionally, a 20 minute and then 1 week trial of the MLP was offered. Safety measures were skin erythema rating, change in visual acuity, and change in corneal staining. Results The MLP and crutch opened the eye (IPF 11.2 and 9.3 mm), but the MLP allowed better volitional closure (IPF 1.0 vs. 4.9 mm, P = 0.009), but was no better in allowing spontaneous blink (IPF 7.5 vs. 7.7 mm, P = 0.722). Both devices were equally comfortable (both median 8/10 comfort, P = 0.46). With extended use, opening with the MLP showed IPF 9.24 mm at 20 minutes and 9.46 mm at 1 week, and volitional closure was IPF 0.95 and 0.52 mm, respectively. Closure on spontaneous blink improved with extended wear to IPF 5.14 and 5.18 mm, respectively (P = 0.002). Two participants exhibited moderate skin erythema and one had increased corneal staining without change in acuity. Conclusions The MLP is safe and feasible for temporary correction of severe ptosis. Translational Relevance First group data in patients showing successful reanimation of the eyelid with magnetic force.
Collapse
Affiliation(s)
- Kevin E Houston
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA.,Spaulding Rehabilitation Hospital, Boston, MA, USA
| | - Matteo Tomasi
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
| | - Christina Amaral
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA
| | - Nicole Finch
- Spaulding Rehabilitation Hospital, Boston, MA, USA
| | - Michael K Yoon
- Oculoplastics Service, Department of Ophthalmology, Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, MA, USA
| | - Hang Lee
- Biostatistics Center, Massachusetts General Hospital, Harvard Medical School, Boston, MA, USA
| | - Eleftherios I Paschalis
- Schepens Eye Research Institute, Massachusetts Eye and Ear, Department of Ophthalmology, Harvard Medical School, Boston, MA, USA.,Boston Keratoprosthesis Laboratory, Department of Ophthalmology, Schepens Eye Research Institute - Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, MA, USA.,Disruptive Technology Laboratory (DTL), Massachusetts Eye and Ear, Harvard Medical School, Boston, MA, USA
| |
Collapse
|