Gavazzi S, van den Berg CAT, Sbrizzi A, Kok HP, Stalpers LJA, Lagendijk JJW, Crezee H, van Lier ALHMW. Accuracy and precision of electrical
permittivity mapping at 3T: the impact of three
B 1 + mapping techniques.
Magn Reson Med 2019;
81:3628-3642. [PMID:
30737816 PMCID:
PMC6593818 DOI:
10.1002/mrm.27675]
[Citation(s) in RCA: 14] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/30/2018] [Revised: 01/07/2019] [Accepted: 01/08/2019] [Indexed: 12/29/2022]
Abstract
Purpose
To investigate the sequence‐specific impact of B1+ amplitude mapping on the accuracy and precision of permittivity reconstruction at 3T in the pelvic region.
Methods
B1+ maps obtained with actual flip angle imaging (AFI), Bloch–Siegert (BS), and dual refocusing echo acquisition mode (DREAM) sequences, set to a clinically feasible scan time of 5 minutes, were compared in terms of accuracy and precision with electromagnetic and Bloch simulations and MR measurements. Permittivity maps were reconstructed based on these B1+ maps with Helmholtz‐based electrical properties tomography. Accuracy and precision in permittivity were assessed. A 2‐compartment phantom with properties and size similar to the human pelvis was used for both simulations and measurements. Measurements were also performed on a female volunteer’s pelvis.
Results
Accuracy was evaluated with noiseless simulations on the phantom. The maximum B1+ bias relative to the true B1+ distribution was 1% for AFI and BS and 6% to 15% for DREAM. This caused an average permittivity bias relative to the true permittivity of 7% to 20% for AFI and BS and 12% to 35% for DREAM. Precision was assessed in MR experiments. The lowest standard deviation in permittivity, found in the phantom for BS, measured 22.4 relative units and corresponded to a standard deviation in B1+ of 0.2% of the B1+ average value. As regards B1+ precision, in vivo and phantom measurements were comparable.
Conclusions
Our simulation framework quantitatively predicts the different impact of B1+ mapping techniques on permittivity reconstruction and shows high sensitivity of permittivity reconstructions to sequence‐specific bias and noise perturbation in the B1+ map. These findings are supported by the experimental results.