Villa-Tamayo MF, Builes-Montaño CE, Ramirez-Rincón A, Carvajal J, Rivadeneira PS. Accuracy of an Off-Label Transmitter and Data Manager Paired With an Intermittent Scanned Continuous Glucose Monitor in Adults With Type 1 Diabetes.
J Diabetes Sci Technol 2024;
18:701-708. [PMID:
36281579 DOI:
10.1177/19322968221133405]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
BACKGROUND
This work evaluates the accuracy and agreement between the FreeStyle Libre sensor (FSL) and an off-label converted real-time continuous glucose monitor (c-rtCGM) device consisting of the MiaoMiao transmitter and the xDrip+ application which can be coupled to the FSL.
METHODS
Four weeks of glucose data were collected from 21 participants with type 1 diabetes using the c-rtCGM and FSL: two weeks with a single initial calibration (uncalibrated) and two weeks with a daily calibration (calibrated). Accuracy and agreement evaluation included mean absolute relative difference (MARD), the %20/20 rule, Bland-Altman plots, and the Consensus Error Grid analysis.
RESULTS
Values reported by the c-rtCGM system compared with the FSL resulted in an overall MARD of 12.06% and 84.71% of the results falling within Consensus Error Grid Zone A when the device is calibrated. For uncalibrated devices, an overall MARD of 17.49% was obtained. Decreased accuracy was shown in the hypoglycemic range and for rates of change greater than 2 mg/dL/min. The between-device bias also incremented with increasing glucose values.
CONCLUSION
Measurements recorded by the c-rtCGM were found to be accurate when compared with FSL data only when performing daily c-rtCGM device calibrations. High drops in accuracy and agreement between devices occurred when the c-rtCGM was not calibrated.