Zafar S, Ahmed TM, Baig R, Jeeva I, Siddiqui MAR. Accuracy of swept source biometry device in detecting macular diseases compared with swept source OCT.
BMC Res Notes 2023;
16:376. [PMID:
38115143 PMCID:
PMC10731895 DOI:
10.1186/s13104-023-06641-3]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/30/2023] [Accepted: 11/28/2023] [Indexed: 12/21/2023] Open
Abstract
OBJECTIVE
To assess the diagnostic accuracy of the IOLMaster 700 foveal scans to detect foveal pathology compared with a standard swept-source optical coherence tomography (SS-OCT) device.
RESULTS
One hundred seventy eye scans of 95 patients were included in the final analyses. Ninety-nine (58.2%) scans were classified as abnormal by SS-OCT. Mean sensitivity of the biometry device was 67.5% (range: 51-84%) and mean specificity was 69.5% (range: 44-95%). Intra-class correlation coefficients were 0.912 and 0.835, for reader 1 and 2, respectively. Area under the curve for receiver operating curve was 0.726. Foveal scans of the IOLMaster 700 can provide clinically useful information. Clinicians should pay attention to the macular scans when reviewing biometry prior to cataract surgery and standard macular OCT should ideally be supplemented in suspicious cases.