Combination of Single- and Paired-Pulse Somatosensory Evoked Potentials in Ischemic Monitoring: Preliminary Investigation in Carotid Endarterectomy.
Cureus 2020;
12:e12206. [PMID:
33489615 PMCID:
PMC7815265 DOI:
10.7759/cureus.12206]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
Introduction
Severe ischemia induces cerebral excitability imbalance before completion of infarct. To investigate the clinical availability of this imbalance with ischemic monitoring, paired-pulse somatosensory evoked potentials (SEPs) were performed in conjunction with conventional SEPs during carotid endarterectomy.
Methods
For carotid endarterectomy patients with hemodynamic deficits of the middle cerebral artery area (n = 34), the excitability imbalances (Q) were measured by paired-pulse SEPs, wherein the second response (A2) was divided by the first (A1; Q = A2/A1). Regional cerebral saturation (rSO2) was also measured. Occlusion was performed twice using shunting.
Results
Each carotid occlusion induced a significant decrease in mean A1 and rSO2, and an increase in mean Q values (p < 0.001), which returned to the baseline level after occlusion. While neuronal imbalances were mostly transient, persistently increased Q values were observed in four cases (11.8%), all indicating postoperative abnormalities in diffusion-weighted magnetic resonance imaging (100%). Meanwhile, A1 detected the postoperative abnormality in only one case (25%). Preoperative Q values at the time of surgery were significantly higher in symptomatic patients having the upper limb deficits than those without (p < 0.01), indicating persistent or permanent imbalances.
Conclusion
Paired-pulse SEPs reliably identified transient, persistent or permanent neuronal imbalances, depending on the ischemic severity. These preliminary results indicated that paired-pulse SEPs, in combination with conventional SEPs (A1), may offer better ischemic monitoring.