1
|
Xing T, Liu J, Sun H, Gao Y, Ju Y, Liu X, Song D. Commutability assessment of reference materials for homocysteine. Clin Chem Lab Med 2022; 60:1562-1569. [PMID: 35977428 DOI: 10.1515/cclm-2022-0388] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/21/2022] [Accepted: 07/18/2022] [Indexed: 11/15/2022]
Abstract
OBJECTIVES Commutability of reference materials is essential for ensuring the traceability of patient measurement results and the technical basis for the use of reference materials. Commutability is only relevant for matrixed reference material; it is a prerequisite for the accuracy and authenticity of calibration methods. In this study, we evaluated the commutability of reference materials for homocysteine. METHODS Five conventional measurement methods were applied to simultaneously measure 30 serum samples and seven homocysteine reference materials from the National Institute of Standards and Technology and the National Institute of Metrology. Liquid chromatography tandem-mass spectrometry was used as a reference method. Two methods were used to evaluate the commutability of the seven reference materials according to the Clinical and Laboratory Standards Institute EP30-A and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) commutability assessment document. RESULTS Among 35 combinations of the five conventional methods and seven reference materials, after evaluation in accordance with the EP30-A, the seven reference materials passed the commutability assessment, and 34 combinations were commutable. According to the IFCC, the commutability evaluation of 28 combinations was conclusive (commutable or non-commutable), while results for the remaining seven combinations could not be determined. CONCLUSIONS The homocysteine reference materials showed good commutability. The sensitivity of the measurement procedure, measurement deviation and uncertainty, and differences in the "measurand" selected by different methods may affect the evaluation results. Additionally, different judgment standards for different methods may explain the observed variations in evaluation results.
Collapse
Affiliation(s)
- Tongtong Xing
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology (Key Laboratory of Chemical Metrology and Applications on Nutrition and Health for State Market Regulation), Beijing, P.R. China.,School of Chemical Engineering, Beijing University of Chemical Technology, Beijing, P.R. China
| | - Jianyi Liu
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology (Key Laboratory of Chemical Metrology and Applications on Nutrition and Health for State Market Regulation), Beijing, P.R. China
| | - Haofeng Sun
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology (Key Laboratory of Chemical Metrology and Applications on Nutrition and Health for State Market Regulation), Beijing, P.R. China.,School of Chemical and Engineering, Nanjing University of Science and Technology, Nanjing, Jiangsu, P.R. China
| | - Yanhong Gao
- Laboratory Department of the First Medical Center of PLA General Hospital, Beijing, P.R. China
| | - Yi Ju
- Shanghai Center for Clinical Laboratory, Shanghai, P.R. China
| | - Xiaolin Liu
- School of Chemical Engineering, Beijing University of Chemical Technology, Beijing, P.R. China
| | - Dewei Song
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology (Key Laboratory of Chemical Metrology and Applications on Nutrition and Health for State Market Regulation), Beijing, P.R. China
| |
Collapse
|
2
|
Verstraete J, Kiekens F, Strobbe S, De Steur H, Gellynck X, Van Der Straeten D, Stove CP. Clinical determination of folates: recent analytical strategies and challenges. Anal Bioanal Chem 2019; 411:4383-4399. [DOI: 10.1007/s00216-019-01574-y] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/22/2018] [Revised: 12/04/2018] [Accepted: 01/03/2019] [Indexed: 01/10/2023]
|
3
|
Hyung SW, Lee J, Seo HY, Kim B. Quantification of Folic Acid in Human Serum Using Isotope Dilution Ultra-High-Pressure Liquid Chromatography/Mass Spectrometry. B KOREAN CHEM SOC 2018. [DOI: 10.1002/bkcs.11356] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/10/2022]
Affiliation(s)
- Seok-Won Hyung
- Division of Chemical and Medical Metrology; Korea Research Institute of Standards and Science; Daejeon 34113 South Korea
| | - Joonhee Lee
- Division of Chemical and Medical Metrology; Korea Research Institute of Standards and Science; Daejeon 34113 South Korea
| | - Hye-Young Seo
- Hygienic Safety and Analysis Center; World Institute of Kimchi; Gwangju 61755 South Korea
| | - Byungjoo Kim
- Division of Chemical and Medical Metrology; Korea Research Institute of Standards and Science; Daejeon 34113 South Korea
| |
Collapse
|
4
|
Fazili Z, Sternberg MR, Paladugula N, Pfeiffer CM. Two International Round-Robin Studies Showed Good Comparability of 5-Methyltetrahydrofolate but Poor Comparability of Folic Acid Measured in Serum by Different High-Performance Liquid Chromatography-Tandem Mass Spectrometry Methods. J Nutr 2017; 147:1815-1825. [PMID: 28768831 PMCID: PMC5712443 DOI: 10.3945/jn.117.254144] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/08/2017] [Revised: 05/31/2017] [Accepted: 06/28/2017] [Indexed: 01/16/2023] Open
Abstract
Background: Serum folate methods produce different results. The comparability of HPLC-mass spectrometry (MS)/MS methods is not well documented.Objective: We conducted an international "round-robin" investigation to assess the comparability, precision, and accuracy of serum folate HPLC-MS/MS methods.Methods: The CDC laboratory, 7 laboratories with independently developed methods (group 1), and 6 laboratories with an adapted CDC method (group 2) analyzed folate forms in 6 serum pools and 6 calibrators from the CDC (duplicate analysis over 2 d) and in two 3-level reference materials (duplicate analysis).Results: All laboratories measured 5-methyltetrahydrofolate (5-methylTHF) and folic acid; some measured additional folate forms. The geometric mean (range) concentrations (nanomoles per liter) for 5-methylTHF in the 6 serum pools were 18.3 nmol/L (CDC), 13.8-28.9 nmol/L (group 1), and 16.8-18.6 nmol/L (group 2); for folic acid the concentrations were 3.42 nmol/L (CDC), 1.09-4.74 nmol/L (group 1), and 1.74-2.90 nmol/L (group 2). The median imprecision (CV) for 5-methylTHF was 4.1% (CDC), 4.6-11% (group 1), and 1.7-6.0% (group 2); for folic acid it was 6.9% (CDC), 4.9-20% (group 1), and 3.9-23% (group 2). The mean ± SD (range) recovery of 5-methylTHF spiked into serum was 98% ± 27% (59-138%) for group 1 and 98% ± 10% (82-111%) for group 2; for folic acid it was 93% ± 29% (67-198%) for group 1 and 81% ± 16% (64-102%) for group 2. The mean relative bias for 5-methylTHF compared with the reference material certificate value was 12% (CDC), -24% to 30% (group 1), and -0.6% to 16% (group 2); for folic acid it was 73% (CDC), -47% to 578% (group 1), and -3.3% to 67% (group 2).Conclusions: For 5-methylTHF, group 2 laboratories demonstrated better agreement and precision, less variable spiking recovery, and less bias by using a reference material. Laboratory performance for folic acid was highly variable and needs improvement. Certified reference materials for serum folate forms and total folate are needed to improve method accuracy.
Collapse
Affiliation(s)
- Zia Fazili
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| | - Maya R Sternberg
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| | - Neelima Paladugula
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| | - Christine M Pfeiffer
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| |
Collapse
|
5
|
Liu Y, Song D, Xu B, Li H, Dai X, Chen B. Development of a matrix-based candidate reference material of total homocysteine in human serum. Anal Bioanal Chem 2017; 409:3329-3335. [PMID: 28271225 DOI: 10.1007/s00216-017-0272-3] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/09/2017] [Revised: 02/10/2017] [Accepted: 02/21/2017] [Indexed: 10/20/2022]
Abstract
We developed and evaluated a candidate serum reference material to help improve clinical routine measurement, and to provide traceability of the measurement results. D8-Homocystine, dithiothreitol, and acetonitrile were used as an internal standard, the reducing agent, and the protein precipitating agent, respectively. A triple quadrupole mass spectrometer with an electrospray ionization source was used for monitoring the transitions (m/z 140.0 → 94.0, 136.0 → 90.0) in multiple-reaction-monitoring mode. We used a calibration model relying on bracketing and gravimetric measurements to give SI-traceability and higher accuracy to serum value assignments. The method was evaluated for accuracy using NIST Standard Reference Material SRM1955. The results of the three concentrations (1, 2, and 3) of total homocysteine in human serum samples were determined by an isotope-dilution liquid chromatography-tandem mass spectrometry method; tHcy 1 is 28.8 ± 1.1 μmol/L, tHcy 2 is 17.93 ± 0.57 μmol/L, and tHcy 3 is 14.38 ± 0.46 μmol/L. Graphical abstract The workflow diagram.
Collapse
Affiliation(s)
- Yu Liu
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology, Beijing, 100029, China
| | - Dewei Song
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology, Beijing, 100029, China.
| | - Bei Xu
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology, Beijing, 100029, China
| | - Hongmei Li
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology, Beijing, 100029, China
| | - Xinhua Dai
- Division of Chemical Metrology and Analytical Science, National Institute of Metrology, Beijing, 100029, China
| | - Baorong Chen
- Beijing Aerospace General Hospital, Beijing, 100076, China
| |
Collapse
|
6
|
Kopp M, Morisset R, Koehler P, Rychlik M. Stable Isotope Dilution Assays for Clinical Analyses of Folates and Other One-Carbon Metabolites: Application to Folate-Deficiency Studies. PLoS One 2016; 11:e0156610. [PMID: 27276031 PMCID: PMC4898732 DOI: 10.1371/journal.pone.0156610] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/14/2015] [Accepted: 05/17/2016] [Indexed: 01/21/2023] Open
Abstract
Folate deficiency is generally accepted as a potential direct or indirect risk factor for diseases including spina bifida, coronary heart diseases, malfunctions of the central nervous system, and cancer. The direct inclusion of folates in the methylation cycle, including the remethylation of homocysteine and regeneration of S-adenosylmethionine, underlines the importance of these vitamins and other components of one-carbon metabolism. Therefore, the aim of the present study was to develop a multiple stable isotope dilution assay (SIDA) for the respective analytes in plasma and tissue samples to allow for a closer look at the interaction between a severe folate deficiency and local folate status, as well as further interactions with circulating S-adenosylmethionine, S-adenosylhomocysteine, and homocysteine. The analytical methods were based on SIDAs coupled with liquid chromatography-tandem mass spectrometry (LC-MS/MS) analysis using the deuterated folates [2H4]-5-methyltetrahydrofolic acid, [2H4]-5-formyltetrahydrofolic acid, [2H4]-tetrahydrofolic acid, [2H4]-10-formylfolic acid, and [2H4]-folic acid and the deuterated one-carbon metabolites [2H4]-homocysteine, [2H4]-S-adenosylhomocysteine, and [2H3]-S-adenosylmethionine as internal standards. Three analytical methods have been developed for the analysis of homocysteine, S-adenosylmethionine, S-adenosylhomocysteine, and six folate vitamers. Validation data for the analysis of C1-metabolites in plasma and tissue samples or folate analysis in tissue samples revealed excellent sensitivity, precision, and recovery for all analytes studied. The miniaturized methods using sample volumes as low as 50 μL and weighed portions of 5-25 mg will allow the assessment of the status of folates and additional biomarkers of impaired one-carbon metabolism during folate deficiency.
Collapse
Affiliation(s)
- Markus Kopp
- Chair of Analytical Food Chemistry, Technische Universität München, Alte Akademie 10, D-85354 Freising, Germany
- Institute for Food & Health (Z I E L), Technische Universiät München, Weihenstephaner Berg 1, D-85354 Freising, Germany
| | - Rosalie Morisset
- Institute for Food & Health (Z I E L), Technische Universiät München, Weihenstephaner Berg 1, D-85354 Freising, Germany
- Chair of Nutritional Physiology, Technische Universität München, Gregor-Mendel-Straße 2, D-85354 Freising, Germany
| | - Peter Koehler
- Deutsche Forschungsanstalt für Lebensmittelchemie, Leibniz Institut, Lise-Meitner-Straße 34, D-85354 Freising, Germany
| | - Michael Rychlik
- Chair of Analytical Food Chemistry, Technische Universität München, Alte Akademie 10, D-85354 Freising, Germany
- * E-mail:
| |
Collapse
|
7
|
Bailey LB, Stover PJ, McNulty H, Fenech MF, Gregory JF, Mills JL, Pfeiffer CM, Fazili Z, Zhang M, Ueland PM, Molloy AM, Caudill MA, Shane B, Berry RJ, Bailey RL, Hausman DB, Raghavan R, Raiten DJ. Biomarkers of Nutrition for Development-Folate Review. J Nutr 2015; 145:1636S-1680S. [PMID: 26451605 PMCID: PMC4478945 DOI: 10.3945/jn.114.206599] [Citation(s) in RCA: 296] [Impact Index Per Article: 32.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/03/2014] [Revised: 12/11/2014] [Accepted: 04/14/2015] [Indexed: 12/13/2022] Open
Abstract
The Biomarkers of Nutrition for Development (BOND) project is designed to provide evidence-based advice to anyone with an interest in the role of nutrition in health. Specifically, the BOND program provides state-of-the-art information and service with regard to selection, use, and interpretation of biomarkers of nutrient exposure, status, function, and effect. To accomplish this objective, expert panels are recruited to evaluate the literature and to draft comprehensive reports on the current state of the art with regard to specific nutrient biology and available biomarkers for assessing nutrients in body tissues at the individual and population level. Phase I of the BOND project includes the evaluation of biomarkers for 6 nutrients: iodine, iron, zinc, folate, vitamin A, and vitamin B-12. This review represents the second in the series of reviews and covers all relevant aspects of folate biology and biomarkers. The article is organized to provide the reader with a full appreciation of folate's history as a public health issue, its biology, and an overview of available biomarkers (serum folate, RBC folate, and plasma homocysteine concentrations) and their interpretation across a range of clinical and population-based uses. The article also includes a list of priority research needs for advancing the area of folate biomarkers related to nutritional health status and development.
Collapse
Affiliation(s)
- Lynn B Bailey
- Department of Foods and Nutrition, University of Georgia, Athens, GA;
| | - Patrick J Stover
- Division of Nutritional Sciences, Cornell University, Ithaca, NY
| | - Helene McNulty
- Northern Ireland Centre for Food and Health, Biomedical Sciences Research Institute, University of Ulster, Londonderry, United Kingdom
| | - Michael F Fenech
- Genome Health Nutrigenomics Laboratory, Food, Nutrition, and Bioproducts Flagship, Commonwealth Scientific and Industrial Research Organization, Adelaide, Australia
| | - Jesse F Gregory
- Food Science and Human Nutrition Department, University of Florida, Gainesville, FL
| | - James L Mills
- Eunice Kennedy Shriver National Institute of Child Health and Human Development, NIH, Bethesda, MD
| | | | - Zia Fazili
- National Center for Environmental Health, CDC, Atlanta, GA
| | - Mindy Zhang
- National Center for Environmental Health, CDC, Atlanta, GA
| | - Per M Ueland
- Department of Clinical Science, Univeristy of Bergen, Bergen, Norway
| | - Anne M Molloy
- Institute of Molecular Medicine, Trinity College, Dublin, Ireland
| | - Marie A Caudill
- Division of Nutritional Sciences, Cornell University, Ithaca, NY
| | - Barry Shane
- Department of Nutritional Sciences and Toxicology, University of California-Berkeley, Berkeley, CA
| | - Robert J Berry
- National Center on Birth Defects and Developmental Disabilities, CDC, Atlanta, GA; and
| | | | - Dorothy B Hausman
- Department of Foods and Nutrition, University of Georgia, Athens, GA
| | - Ramkripa Raghavan
- Eunice Kennedy Shriver National Institute of Child Health and Human Development, NIH, Bethesda, MD
| | - Daniel J Raiten
- Eunice Kennedy Shriver National Institute of Child Health and Human Development, NIH, Bethesda, MD;
| |
Collapse
|
8
|
Wang X, Zhang T, Zhao X, Guan Z, Wang Z, Zhu Z, Xie Q, Wang J, Niu B. Quantification of folate metabolites in serum using ultraperformance liquid chromatography tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci 2014; 962:9-13. [PMID: 24878879 DOI: 10.1016/j.jchromb.2014.05.023] [Citation(s) in RCA: 18] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/13/2014] [Revised: 05/07/2014] [Accepted: 05/08/2014] [Indexed: 10/25/2022]
Abstract
Folate deficiency is considered a risk factor for many diseases such as cancer, congenital heart disease and neural tube defects (NTDs). There is a pressing need for more methods of detecting folate and its main metabolites in the human body. Here, we developed a simple, fast and sensitive ultraperformance liquid chromatography tandem mass spectrometry (UPLC/MS/MS) method for the simultaneous quantifications of folate metabolites including folic acid, 5-methyltetrahydrofolate (5-MeTHF), 5-formyltetrahydrofolate (5-FoTHF), homocysteine (Hcy), S-adenosylmethionine (SAM) and S-adenosylhomocysteine (SAH). The method was validated by determining the linearity (r(2)>0.998), sensitivity (limit of detection ranged from 0.05 to 0.200ng/mL), intra- and inter-day precision (both CV<6%) and recovery (each analyte was >90%). The total analysis time was 7min. Serum samples of NTD-affected pregnancies and controls from a NTD high-risk area in China were analyzed by this method, the NTD serum samples showed lower concentrations of 5-MeTHF (P<0.05) and 5-FoTHF (P<0.05), and higher concentrations of Hcy (P<0.05) and SAH (P<0.05) compared with serum samples from controls, consistent with a previous study. These results showed that the method is sensitive and reliable for simultaneous determination of six metabolites, which might indicate potential risk factors for NTDs, aid early diagnosis and provide more insights into the pathogenesis of NTDs.
Collapse
Affiliation(s)
- Xiuwei Wang
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Ting Zhang
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Xin Zhao
- Chines Academy of Inspection & Quarantine, Beijing 100023, China
| | - Zhen Guan
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Zhen Wang
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Zhiqiang Zhu
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Qiu Xie
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China
| | - Jianhua Wang
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China.
| | - Bo Niu
- Beijing Municipal Key Laboratory of Child Development and Nutriomics, Capital Institute of Pediatrics, Beijing 100020, China; Department of Biochemistry and Molecular Biology, Shanxi Medical University, Taiyuan 030001, China.
| |
Collapse
|
9
|
Pfeiffer CM, Hughes JP, Durazo-Arvizu RA, Lacher DA, Sempos CT, Zhang M, Yetley EA, Johnson CL. Changes in measurement procedure from a radioassay to a microbiologic assay necessitate adjustment of serum and RBC folate concentrations in the U.S. population from the NHANES 1988-2010. J Nutr 2012; 142:894-900. [PMID: 22437557 PMCID: PMC3327748 DOI: 10.3945/jn.111.156901] [Citation(s) in RCA: 23] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/21/2023] Open
Abstract
The NHANES measured serum and RBC folate concentrations by using a radioassay during prefortification (1988-1994) and postfortification (1999-2006) periods followed by the use of a microbiologic assay (MBA) from 2007-2010. The MBA produces higher concentrations than does the radioassay and is considered to be more accurate. To allow for accurate long-term trending (1988-2010), we evaluated different regression models (linear, piecewise linear, and fractional polynomial) to assay-adjust the radioassay results to be comparable to the MBA results. The data used to derive the regression models originated from 2 crossover studies in which the 2 assays were applied to a set of 325 serum and 171 whole-blood samples. Fractional polynomial regression of logarithmically transformed data provided the best fit for serum folate. Linear regression of logarithmically transformed whole-blood data provided an equally good fit compared with the other models and was the simplest to apply for RBC folate. Prefortification serum and RBC folate geometric mean concentrations increased after adjustment from 13.0 to 16.7 nmol/L and from 403 to 747 nmol/L, respectively. Postfortification serum folate concentrations increased from ~30 to ~43 nmol/L, and RBC folate concentrations increased from ~600 to ~1100 nmol/L after adjustment, with some variation across survey cycles. The presented regression equations allow the estimation of more accurate prevalence estimates and long-term trends in blood folate concentrations in the U.S. population by using results that are equivalent to the MBA. This information will be useful to public health officials in the United States who are dealing with folic acid fortification issues.
Collapse
Affiliation(s)
- Christine M. Pfeiffer
- National Center for Environmental Health, CDC, Atlanta, GA,To whom correspondence should be addressed. E-mail:
| | | | | | | | | | - Mindy Zhang
- National Center for Environmental Health, CDC, Atlanta, GA
| | | | | |
Collapse
|
10
|
Cobalamin and folic Acid status in relation to the etiopathogenesis of pancytopenia in adults at a tertiary care centre in north India. Anemia 2012; 2012:707402. [PMID: 22545211 PMCID: PMC3321527 DOI: 10.1155/2012/707402] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/20/2011] [Revised: 12/06/2011] [Accepted: 12/06/2011] [Indexed: 11/17/2022] Open
Abstract
Background. Pancytopenia has multiple etiologies like megaloblastic anemia, aplastic anemia, leukemia, and various infections. We investigated the clinical, etiological and hematological profile including bone marrow morphology of patients with pancytopenia in relation to their vitamin B12 and folic acid status at a tertiary care referral hospital in north India. Methods. A total of 140 consecutive patients with pancytopenia were selected from June 2007 to December 2008. Bone marrow examination and other tests were carried out as warranted, including serum cobalamin and folate assays using liquid chromatography mass spectroscopy (LC MS/MS). Results. The study population consisted of 92 males and 48 females with a mean age of 32.8 years. Megaloblastic anemia 60.7%, aplastic anemia (7.8%), and leukemia (9.2%) were common causes. Infectious causes (16.4% of all cases) included leishmaniasis, HIV–AIDS, malaria and tuberculosis. Severe cobalamin deficiency (B12 < 100 pg/mL) was seen in 81% of all patients including 91.6% of patients with MA. In contrast, only 7.14% of all pancytopenic patients were folate deficient. Folate deficiency (<5 ng/mL) was seen in just 5% MA patients. Combined cobalamin and folate deficiency was seen in 5 patients (3.51%). Conclusion. Cobalamin deficiency was found to be more common in our setting and is largely underdiagnosed in the age of folate supplementation. Infectious diseases like tuberculosis, leishmaniasis, and increasingly HIV are important and treatable causes of pancytopenia. This is in contrast with the developed nations where the bulk of disease is due to malignancy or marrow aplasia.
Collapse
|
11
|
NAKAZATO M, MAEDA T, EMURA K, MAEDA M, TAMURA T. Blood Folate Concentrations Analyzed by Microbiological Assay and Chemiluminescent Immunoassay Methods. J Nutr Sci Vitaminol (Tokyo) 2012; 58:59-62. [DOI: 10.3177/jnsv.58.59] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/27/2022]
|
12
|
Bock JL, Eckfeldt JH. Advances in standardization of laboratory measurement procedures: implications for measuring biomarkers of folate and vitamin B-12 status in NHANES. Am J Clin Nutr 2011; 94:332S-336S. [PMID: 21562088 PMCID: PMC3127510 DOI: 10.3945/ajcn.111.013359] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/22/2023] Open
Abstract
Population studies such as NHANES analyze large numbers of laboratory measurements and are often performed in different laboratories using different measurement procedures and over an extended period of time. Correct clinical and epidemiologic interpretations of the results depend on the accuracy of those measurements. Unfortunately, considerable variability has been observed among assays for folate, vitamin B-12, and related biomarkers. In the past few decades, the science of metrology has advanced considerably, with the development of improved primary reference measurement procedures and high-level reference materials, which can serve as the basis for accurate measurement. A rigorous approach has been established for making field methods traceable to the highest-level reference measurement procedures and reference materials. This article reviews some basic principles of metrology and describes their recent application to measurements of folate and vitamin B-12.
Collapse
Affiliation(s)
- Jay L Bock
- Department of Pathology, Stony Brook University Medical Center, Stony Brook, NY, USA.
| | | |
Collapse
|
13
|
Blackmore S, Pfeiffer CM, Lee A, Fazili Z, Hamilton MS. Isotope Dilution–LC-MS/MS Reference Method Assessment of Serum Folate Assay Accuracy and Proficiency Testing Consensus Mean. Clin Chem 2011; 57:986-94. [DOI: 10.1373/clinchem.2010.160135] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
BACKGROUND
Current methods for measuring folates in clinical laboratories are competitive folate binding protein assays. These assays show a considerable lack of agreement that has implications for the comparability of data across studies as well as for long-term population studies. The development of isotope dilution–liquid chromatography–tandem mass spectrometry (ID-LC-MS/MS) reference methods permitted the evaluation of method accuracy and consistency over time.
METHODS
We measured 3 pools of human serum by ID-LC-MS/MS, calculated values for total folate, and distributed the same pools to participants in a national External Quality Assessment scheme. We used linear regression to compare the all-laboratory and method data with reference method values. The exercise was repeated after 18 months to assess the stability of the all-laboratory and method data.
RESULTS
The distributed serum pools had mass spectrometry values for folate species typical of those found in healthy individuals from populations not receiving dietary folic acid fortification. There was good agreement of the all-laboratory data set with the reference method (y =0.86x + 0.91 μg/L) at both time points. Linear regression demonstrated that the Abbott Architect showed the closest agreement with the reference method. The Roche Elecsys method was nonlinear and showed a calibration offset of 2.6 μg/L (4.57 nmol/L).
CONCLUSIONS
Calibration of serum folate assays traceable to higher-order reference methods increases method accuracy and improves consistency. The all-laboratory consensus mean proved sufficiently accurate and stable to be used as the target for monitoring laboratory performance.
Collapse
Affiliation(s)
- Sheena Blackmore
- UK NEQAS Haematinics Scheme for Haematinics, Good Hope Hospital, Sutton Coldfield, UK
| | - Christine M Pfeiffer
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| | - Anne Lee
- UK NEQAS Haematinics Scheme for Haematinics, Good Hope Hospital, Sutton Coldfield, UK
| | - Zia Fazili
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA
| | - Malcolm S Hamilton
- UK NEQAS Haematinics Scheme for Haematinics, Good Hope Hospital, Sutton Coldfield, UK
| |
Collapse
|
14
|
Yetley EA, Pfeiffer CM, Phinney KW, Fazili Z, Lacher DA, Bailey RL, Blackmore S, Bock JL, Brody LC, Carmel R, Curtin LR, Durazo-Arvizu RA, Eckfeldt JH, Green R, Gregory JF, Hoofnagle AN, Jacobsen DW, Jacques PF, Molloy AM, Massaro J, Mills JL, Nexo E, Rader JI, Selhub J, Sempos C, Shane B, Stabler S, Stover P, Tamura T, Tedstone A, Thorpe SJ, Coates PM, Johnson CL, Picciano MF. Biomarkers of folate status in NHANES: a roundtable summary. Am J Clin Nutr 2011; 94:303S-312S. [PMID: 21593502 PMCID: PMC3127517 DOI: 10.3945/ajcn.111.013011] [Citation(s) in RCA: 92] [Impact Index Per Article: 7.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/21/2023] Open
Abstract
A roundtable to discuss the measurement of folate status biomarkers in NHANES took place in July 2010. NHANES has measured serum folate since 1974 and red blood cell (RBC) folate since 1978 with the use of several different measurement procedures. Data on serum 5-methyltetrahydrofolate (5MTHF) and folic acid (FA) concentrations in persons aged ≥60 y are available in NHANES 1999-2002. The roundtable reviewed data that showed that folate concentrations from the Bio-Rad Quantaphase II procedure (Bio-Rad Laboratories, Hercules, CA; used in NHANES 1991-1994 and NHANES 1999-2006) were, on average, 29% lower for serum and 45% lower for RBC than were those from the microbiological assay (MA), which was used in NHANES 2007-2010. Roundtable experts agreed that these differences required a data adjustment for time-trend analyses. The roundtable reviewed the possible use of an isotope-dilution liquid chromatography-tandem mass spectrometry (LC-MS/MS) measurement procedure for future NHANES and agreed that the close agreement between the MA and LC-MS/MS results for serum folate supported conversion to the LC-MS/MS procedure. However, for RBC folate, the MA gave 25% higher concentrations than did the LC-MS/MS procedure. The roundtable agreed that the use of the LC-MS/MS procedure to measure RBC folate is premature at this time. The roundtable reviewed the reference materials available or under development at the National Institute of Standards and Technology and recognized the challenges related to, and the scientific need for, these materials. They noted the need for a commutability study for the available reference materials for serum 5MTHF and FA.
Collapse
Affiliation(s)
- Elizabeth A Yetley
- Office of Dietary Supplements, National Institutes of Health, Bethesda, MD 20892, USA.
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
15
|
Yetley EA, Johnson CL. Folate and vitamin B-12 biomarkers in NHANES: history of their measurement and use. Am J Clin Nutr 2011; 94:322S-331S. [PMID: 21593508 PMCID: PMC3127520 DOI: 10.3945/ajcn.111.013300] [Citation(s) in RCA: 36] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/21/2023] Open
Abstract
NHANES measured folate and vitamin B-12 status biomarkers, starting with serum folate from NHANES I (1974-1975) through 2010. Subsequent NHANES measured additional biomarkers [eg, red blood cell folate, serum vitamin B-12, total homocysteine (tHcy), methylmalonic acid, serum folic acid, and 5-methyltetrahydrofolic acid]. Examples of the uses of these data are wide ranging and include public policy applications, the derivation of reference intervals, and research. Periodically, the National Center for Health Statistics and its federal partners convene expert panels to review the use of the folate- and vitamin B-12-related biomarkers in NHANES. These panels have evaluated the need for results to be comparable across time and with published data and the use of crossover studies and adjustment equations to ensure comparability. With the recent availability of reference methods and materials for serum folate and tHcy, NHANES has started to use traceability approaches to enhance the accuracy and comparability of its results. A major user concern over the years has been the use of cutoffs to estimate the prevalence of inadequate folate and vitamin B-12 status. Because these cutoffs depend on the measurement procedure, several expert panels suggested approaches for dealing with cutoff challenges. This review summarizes the history and use of folate- and vitamin B-12-related biomarkers beginning with NHANES I (1974-1975) through 2010.
Collapse
Affiliation(s)
- Elizabeth A Yetley
- Office of Dietary Supplements, National Institutes of Health, Bethesda, MD 20892, USA.
| | | |
Collapse
|
16
|
Pfeiffer CM, Zhang M, Lacher DA, Molloy AM, Tamura T, Yetley EA, Picciano MF, Johnson CL. Comparison of serum and red blood cell folate microbiologic assays for national population surveys. J Nutr 2011; 141:1402-9. [PMID: 21613453 PMCID: PMC3113292 DOI: 10.3945/jn.111.141515] [Citation(s) in RCA: 43] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/14/2022] Open
Abstract
Three laboratories participated with their laboratory-specific microbiologic growth assays (MA) in the NHANES 2007-2008 to assess whether the distributions of serum (n = 2645) and RBC folate (n = 2613) for the same one-third sample of participants were comparable among laboratories. Laboratory (L) 2 produced the highest and L1 the lowest serum and RBC folate geometric means (nmol/L) in the NHANES sample (serum: L1, 39.5; L2, 59.2; L3, 47.7; and RBC: L1, 1120; L2, 1380; L3, 1380). Each laboratory produced different reference intervals for the central 95% of the population. Pearson correlation coefficients were highest between L3 and L1 (serum, r = 0.95; RBC, r = 0.92) and lowest between L2 and L1 (serum, r = 0.81; RBC, r = 0.65). Notable procedural differences among the laboratories were the Lactobacillus rhamnosus microorganism (L1 and L3: chloramphenicol resistant, L2: wild type) and the calibrator [L1: [6S]5-methyltetrahydrofolate (5-methylTHF), L2: [6R,S] 5-formyltetrahydrofolate ([6R,S] 5-formylTHF), L3: folic acid (FA)]. Compared with 5-methylTHF as calibrator, the folate results were 22-32% higher with FA as calibrator and 8% higher with 5-formylTHF as calibrator, regardless of the matrix (n = 30 serum, n = 28 RBC). The use of different calibrators explained most of the differences in results between L3 and L1 but not between L2 and L1. The use of the wild-type L. rhamnosus by L2 appeared to be the main reason for the differences in results between L2 and the other 2 laboratories. These findings indicate how assay variations influence MA folate results and how those variations can affect population data. To ensure data comparability, better assay harmonization is needed.
Collapse
Affiliation(s)
- Christine M. Pfeiffer
- National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA 30341,To whom correspondence should be addressed. E-mail:
| | - Mindy Zhang
- National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA 30341
| | - David A. Lacher
- National Center for Health Statistics, Centers for Disease Control and Preventio, Hyattsville, MD 20782
| | - Anne M. Molloy
- Institute of Molecular Medicine, Trinity College, Dublin Ireland
| | | | - Elizabeth A. Yetley
- Office of Dietary Supplements, National Institutes of Health, Bethesda, MD 20892
| | | | - Clifford L. Johnson
- National Center for Health Statistics, Centers for Disease Control and Preventio, Hyattsville, MD 20782
| |
Collapse
|
17
|
Yetley EA, Pfeiffer CM, Phinney KW, Bailey RL, Blackmore S, Bock JL, Brody LC, Carmel R, Curtin LR, Durazo-Arvizu RA, Eckfeldt JH, Green R, Gregory JF, Hoofnagle AN, Jacobsen DW, Jacques PF, Lacher DA, Molloy AM, Massaro J, Mills JL, Nexo E, Rader JI, Selhub J, Sempos C, Shane B, Stabler S, Stover P, Tamura T, Tedstone A, Thorpe SJ, Coates PM, Johnson CL, Picciano MF. Biomarkers of vitamin B-12 status in NHANES: a roundtable summary. Am J Clin Nutr 2011; 94:313S-321S. [PMID: 21593512 PMCID: PMC3127527 DOI: 10.3945/ajcn.111.013243] [Citation(s) in RCA: 132] [Impact Index Per Article: 10.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/11/2022] Open
Abstract
A roundtable to discuss the measurement of vitamin B-12 (cobalamin) status biomarkers in NHANES took place in July 2010. NHANES stopped measuring vitamin B-12-related biomarkers after 2006. The roundtable reviewed 3 biomarkers of vitamin B-12 status used in past NHANES--serum vitamin B-12, methylmalonic acid (MMA), and total homocysteine (tHcy)--and discussed the potential utility of measuring holotranscobalamin (holoTC) for future NHANES. The roundtable focused on public health considerations and the quality of the measurement procedures and reference methods and materials that past NHANES used or that are available for future NHANES. Roundtable members supported reinstating vitamin B-12 status measures in NHANES. They noted evolving concerns and uncertainties regarding whether subclinical (mild, asymptomatic) vitamin B-12 deficiency is a public health concern. They identified the need for evidence from clinical trials to address causal relations between subclinical vitamin B-12 deficiency and adverse health outcomes as well as appropriate cutoffs for interpreting vitamin B-12-related biomarkers. They agreed that problems with sensitivity and specificity of individual biomarkers underscore the need for including at least one biomarker of circulating vitamin B-12 (serum vitamin B-12 or holoTC) and one functional biomarker (MMA or tHcy) in NHANES. The inclusion of both serum vitamin B-12 and plasma MMA, which have been associated with cognitive dysfunction and anemia in NHANES and in other population-based studies, was preferable to provide continuity with past NHANES. Reliable measurement procedures are available, and National Institute of Standards and Technology reference materials are available or in development for serum vitamin B-12 and MMA.
Collapse
Affiliation(s)
- Elizabeth A Yetley
- Office of Dietary Supplements, National Institutes of Health, Bethesda, MD 20892-7517, USA.
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
18
|
Ihara H, Watanabe T, Hashizume N, Totani M, Kamioka K, Onda K, Sunahara S, Suzuki T, Itabashi M, Aoki Y, Ishibashi M, Ito S, Ohashi K, Enomoto T, Saito K, Saeki K, Nagamura Y, Nobori T, Hirota K, Fujishiro K, Maekawa M, Miura M, Ohta Y. Commutability of National Institute of Standards and Technology standard reference material 1955 homocysteine and folate in frozen human serum for total folate with automated assays. Ann Clin Biochem 2010; 47:541-8. [PMID: 20926465 DOI: 10.1258/acb.2010.010094] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Abstract
BACKGROUND The aim of the present study was to evaluate standard reference material (SRM) 1955 commutability as a reference material for serum folate using automated methods. We also designed so as to reduce the intermethod variability present in different automated methods. METHODS Using a microbiological assay related to the 'information value' of SRM 1955 as a comparison method, we investigated the possibility of standardization for the assay values of serum folate as measured by the automated methods (Access, Centaur and Elecsys). In the assay of 50 patient sera by these automated methods, we corrected observed values by the SRM 1955 and compared with comparison values. RESULTS The observed values of SRM 1955 Levels I, II and III were within or outside (but near) a 95% prediction interval obtained from patient sera by the automated methods. The normalized residuals obtained from SRM 1955 were within ±3.0 (in SD units), which enabled us to conclude that the SRM 1955 had a physicochemical characterization similar to native serum. Twelve patients were assessed as hypofolataemia (<6.0 ng/mL) and 38 patients as normal (≥6.0 ng/mL). Before correction, folate levels in six of 12 patients were lower than 6.0 ng/mL, and those in seven of 38 patients were higher than 6.0 ng/mL with the automated methods. After correction, low levels were found in four of 12 patients, and normal levels were found in 33 of 38 patients. CONCLUSIONS The use of SRM 1955 would help to reduce the intermethod variability present in different automated methods for serum folate measurement.
Collapse
Affiliation(s)
- Hiroshi Ihara
- Department of Laboratory Medicine, Toho University Ohashi Medical Center, Ohashi, Meguro, Tokyo, Japan.
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
19
|
Tai SSC, Prendergast JL, Sniegoski LT, Welch MJ, Phinney KW, Zhang NF. Certification of drugs of abuse in a human serum standard reference material: SRM 1959. Anal Bioanal Chem 2010; 397:501-9. [PMID: 20333365 DOI: 10.1007/s00216-010-3603-1] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/07/2009] [Revised: 02/19/2010] [Accepted: 02/19/2010] [Indexed: 11/25/2022]
Abstract
A new standard reference material (SRM) for drugs of abuse in human serum (SRM 1959) has been developed. This SRM is intended to be used as a control material for laboratories performing analysis of drugs of abuse in blood to evaluate the accuracy of their methods. SRM 1959 is a frozen human serum material fortified with seven compounds for which analyses are performed to determine evidence of illegal drug use: benzoylecgonine (BZE), methadone (METH), methamphetamine (MAMP), morphine (MOR), nordiazepam (NOR), phencyclidine (PCP), and 11-nor-Delta(9)-tetrahydrocannabinol-9-carboxylic acid (THC-9-COOH). Two independent methods involving isotope dilution (ID)-gas chromatography/mass spectrometry (GC/MS) and ID-liquid chromatography/mass spectrometry (LC/MS) were used for the value assignment. For THC-9-COOH, an additional measurement using LC/tandem mass spectrometry (LC/MS/MS) was also included. All methods used isotopically labeled compounds as internal standards and solid-phase extractions to isolate the analytes from the serum. The GC/MS methods used different clean-up procedures from those used for the LC/MS-based methods. Repeatability with within-set coefficients of variation (CVs) ranged from 0.5% to 4.3% for the GC/MS methods and from 0.2% to 1.2% for the LC/MS-based methods. Intermediate precision with between-set CVs for all the methods ranged from 0.1% to 1.1%. Agreement between the GC/MS and LC/MS methods ranged from 0.8% to 8.8%. The results from the methods were combined to obtain the certified concentrations and their expanded uncertainties.
Collapse
Affiliation(s)
- Susan S-C Tai
- Analytical Chemistry Division, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.
| | | | | | | | | | | |
Collapse
|
20
|
Berry RJ, Bailey L, Mulinare J, Bower C, Dary O. Fortification of Flour with Folic Acid. Food Nutr Bull 2010; 31:S22-35. [DOI: 10.1177/15648265100311s103] [Citation(s) in RCA: 123] [Impact Index Per Article: 8.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/10/2023]
Abstract
Background After randomized, controlled trials established that consumption of folic acid before pregnancy and during the early weeks of gestation reduces the risk of a neural tube defect (NTD)-affected pregnancy, the United States Public Health Service recommended in 1992 that all women capable of becoming pregnant consume 400 μg folic acid daily. In 1998, folic acid fortification of all enriched cereal grain product flour was fully implemented in the United States and Canada. Objective To provide guidance on national fortification of wheat and maize flours to prevent 50 to 70% of the estimated 300,000 NTD-affected pregnancies worldwide. Methods An expert workgroup reviewed the latest evidence of effectiveness of folic acid flour fortification and the safety of folic acid. Results Recent estimates show that in the United States and Canada, the additional intake of about 100 to 150 μg/day of folic acid through food fortification has been effective in reducing the prevalence of NTDs at birth and increasing blood folate concentrations in both countries. Most potential adverse effects associated with folic acid are associated with extra supplement use not mandatory fortification. Fortification of wheat flour has a proven record of prevention in other developed countries. In 2009, 51 countries had regulations written for mandatory wheat flour fortification programs that included folic acid. Conclusions NTDs remain an important cause of perinatal mortality and infantile paralysis worldwide. Mandatory fortification of flour with folic acid has proved to be one of the most successful public health interventions in reducing the prevalence of NTD-affected pregnancies. Most developing countries have few, if any, common sources of folic acid, unlike many developed countries, which have folic acid available from ready-to-eat cereals and supplements. Expanding the number of developed and developing countries with folic acid flour fortification has tremendous potential to safely eliminate most folic acid-preventable NTDs.
Collapse
|
21
|
Tomaiuolo M, Vecchione G, Margaglione M, Pisanelli D, Grandone E. Stable-isotope dilution LC–ESI-MS/MS techniques for the quantification of total homocysteine in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci 2009; 877:3292-9. [DOI: 10.1016/j.jchromb.2009.07.024] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/30/2009] [Revised: 07/03/2009] [Accepted: 07/15/2009] [Indexed: 11/25/2022]
|
22
|
Simultaneous analysis of multiple aminothiols in human plasma by high performance liquid chromatography with fluorescence detection. J Chromatogr B Analyt Technol Biomed Life Sci 2009; 877:3274-81. [DOI: 10.1016/j.jchromb.2009.05.046] [Citation(s) in RCA: 76] [Impact Index Per Article: 5.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/02/2009] [Revised: 05/18/2009] [Accepted: 05/21/2009] [Indexed: 11/18/2022]
|
23
|
Ma YY, Shek CC, Wong MCK, Yip KC, Ng RMK, Nguyen DGH, Poon TK. Homocysteine level in schizophrenia patients. Aust N Z J Psychiatry 2009; 43:760-5. [PMID: 19629798 DOI: 10.1080/00048670903001935] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 12/15/2022]
Abstract
OBJECTIVE Elevated serum levels of homocysteine have been shown to be associated with schizophrenia in some studies, but the evidence is still limited and mixed. The aim of the present study was therefore to evaluate the serum homocysteine level in Chinese in schizophrenia patients and assess its association with clinical phenotypes of the disease. METHOD A total of 250 inpatients and 250 healthy controls were identified at Kowloon Hospital and the Red Cross Blood Donation Centre, respectively. Each subject was evaluated with a structured diagnostic interview. Demographic data were collected and blood was analysed for homocysteine level. The Positive and Negative Syndrome Scale was used to grade clinical symptoms of schizophrenia. Confounding factors affecting homocysteine levels were controlled by strict exclusion criteria or statistical methods. RESULTS Serum homocysteine level was elevated in Chinese schizophrenia patients. Both male (mean, 12.26 micromol L(-1); control, 11.40 micromol L(-1), p = 0.026) and female (mean, 9.61 micromol L(-1); control, 8.33 micromol L(-1), p < 0.001) patients had a higher serum homocysteine level than the healthy controls, and these differences persisted after controlling for age. There was no significant association between serum homocysteine level and duration of illness, clinical symptoms or age of onset of schizophrenia. CONCLUSION The serum homocysteine levels were elevated in both male and female Chinese schizophrenia patients. The levels were not apparently related to other clinical parameters.
Collapse
Affiliation(s)
- Yin Ying Ma
- Department of Psychiatry, Kowloon Hospital, Kowloon, Hong Kong.
| | | | | | | | | | | | | |
Collapse
|
24
|
Zhang HY, Luo GA, Liang QL, Wang Y, Yang HH, Wang YM, Zheng XY, Song XM, Chen G, Zhang T, Wu JX. Neural tube defects and disturbed maternal folate- and homocysteine-mediated one-carbon metabolism. Exp Neurol 2008; 212:515-21. [DOI: 10.1016/j.expneurol.2008.04.044] [Citation(s) in RCA: 48] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/02/2008] [Revised: 04/09/2008] [Accepted: 04/28/2008] [Indexed: 11/16/2022]
|
25
|
Huang Y, Khartulyari S, Morales ME, Stanislawska-Sachadyn A, Von Feldt JM, Whitehead AS, Blair IA. Quantification of key red blood cell folates from subjects with defined MTHFR 677C>T genotypes using stable isotope dilution liquid chromatography/mass spectrometry. RAPID COMMUNICATIONS IN MASS SPECTROMETRY : RCM 2008; 22:2403-12. [PMID: 18634122 PMCID: PMC4400668 DOI: 10.1002/rcm.3624] [Citation(s) in RCA: 21] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/26/2023]
Abstract
Red blood cell (RBC) folate levels are established at the time of erythropoiesis and therefore provide a surrogate biomarker for the average folate status of an individual over the preceding four months. Folates are present as folylpolyglutamates, highly polar molecules that cannot be secreted from the RBCs, and must be converted into their monoglutamate forms prior to analysis. This was accomplished using an individual's plasma pteroylpolyglutamate hydrolase by lysing the RBCs in whole blood at pH 5 in the presence of ascorbic acid. Quantitative conversion of formylated tetrahydrofolate derivatives into the stable 5,10-methenyltetrahydrofolate (5,10-MTHF) form was conducted at pH 1.5 in the presence of [(13)C(5)]-5-formyltetrahydrofolate. The resulting [(13)C(5)]-5,10-MTHF was then used as an internal standard for the formylated forms of tetrahydrofolate that had been converted into 5,10-MTHF as well any 5,10-MTHF that had been present in the original sample. A stable isotope dilution liquid chromatography-multiple reaction monitoring/mass spectrometry method was validated and then used for the accurate and precise quantification of RBC folic acid, 5-methyltetrahydrofolate (5-MTHF), tetrahydrofolate (THF), and 5,10-MTHF. The method was sensitive and robust and was used to assess the relationship between different methylenetetrahydrofolate reductase (MTHFR) 677C>T genotypes and RBC folate phenotypes. Four distinct RBC folate phenotypes could be identified. These were classified according to the relative amounts of individual RBC folates as type I (5-MTHF >95%; THF <5%; 5,10-MTHF <5%), type II (5-MTHF <95%; THF 5% to 20%; 5,10-MTHF <5%), type III (5-MTHF >55%; THF >20%; 5,10-MTHF >5%), and type IV (5-MTHF <55%; THF >20%; 5,10-MTHF >5%).
Collapse
Affiliation(s)
- Yuehua Huang
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Stefanie Khartulyari
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Megan E. Morales
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Anna Stanislawska-Sachadyn
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Joan M. Von Feldt
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Alexander S. Whitehead
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| | - Ian A. Blair
- Centers for Cancer Pharmacology, Pharmacogenetics, and Excellence in Environmental Toxicology, Department of Pharmacology, and Division of Rheumatology, Department of Medicine, University of Pennsylvania School of Medicine, Philadelphia, PA, USA
| |
Collapse
|
26
|
Commutability of NIST SRM 1955 Homocysteine and Folate in Frozen Human Serum with selected total homocysteine immunoassays and enzymatic assays. Clin Chim Acta 2008; 395:99-105. [PMID: 18565331 DOI: 10.1016/j.cca.2008.05.016] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/04/2008] [Revised: 05/16/2008] [Accepted: 05/21/2008] [Indexed: 11/24/2022]
Abstract
BACKGROUND The National Institute of Standards and Technology (NIST) has recently developed Standard Reference Material (SRM) 1955 Homocysteine and Folate in Frozen Human Serum with certified values for total homocysteine (tHcy) and 5-methyl-tetrahydrofolic acid. NIST has performed an international, interlaboratory assessment of SRM 1955 commutability; results are reported for tHcy only. METHODS Total Hcy was measured in 20 patient sera and in 3 levels of SRM 1955 using 14 immunoassays and/or enzymatic assays. Liquid chromatography/tandem mass spectrometry was utilized as the reference assay. An "errors-in-variables" statistical model was utilized to assess the commutability of SRM 1955. RESULTS Normalized residuals ranged from -2.65 to 2.19 for SRM 1955. The median interlaboratory/interassay imprecision (CV) was approximately 4% for patient specimens and ranged from approximately 3% to approximately 7% for SRM 1955. The median intra-assay imprecision ranged from approximately 1% to approximately 13%. Orthogonal residuals, as a descriptor of assay accuracy, ranged from 0.29 to 7.71 and from 0.20 to 2.22 for patient specimens and SRM 1955 samples, respectively. CONCLUSION The current study suggests that SRM 1955 is commutable with the investigated tHcy assays; however, a broader specimen set needs to be evaluated to completely substantiate this conclusion.
Collapse
|
27
|
Pfeiffer CM, Osterloh JD, Kennedy-Stephenson J, Picciano MF, Yetley EA, Rader JI, Johnson CL. Trends in circulating concentrations of total homocysteine among US adolescents and adults: findings from the 1991-1994 and 1999-2004 National Health and Nutrition Examination Surveys. Clin Chem 2008; 54:801-13. [PMID: 18375482 DOI: 10.1373/clinchem.2007.100214] [Citation(s) in RCA: 42] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
BACKGROUND The National Health and Nutrition Examination Survey (NHANES) has monitored total homocysteine (tHcy) concentrations in a nationally-representative sample of the US population since 1991. Until recently, however, data could not be compared across survey periods because of changes in analytical methods and specimen matrices. Such an analysis of these data could supplement current knowledge regarding whether the US folic acid fortification program has modified national plasma tHcy concentrations. METHODS We examined tHcy data in the prefortification NHANES III survey (phase II, 1991-1994) and in 3 postfortification survey periods (1999-2000, 2001-2002, and 2003-2004). We applied method adjustment equations to the survey data based on method comparison studies of separate samples. Persons with chronic kidney disease were excluded from the analyses. RESULTS Mean plasma tHcy concentrations decreased by 8%, 9%, and 10% for adolescent, adult, and older men and by 6%, 3%, and 13% for women, respectively, from before to after fortification. Concentrations remained unchanged between the first and third postfortification survey periods. Prevalence estimates of increased plasma tHcy concentrations (>13 micromol/L) for older men and women decreased from prefortification (32% and 20%, respectively) to postfortification (14% and 5%, respectively) but remained unchanged thereafter (16% and 14%, respectively [males] and 5% and 9%, respectively [females]). CONCLUSIONS After adjusting for method changes, we quantified a prefortification to postfortification decrease in circulating tHcy concentrations of about 10% in a national sample of the US population. This change is similar to effects seen in intervention trials with folic acid and in smaller observational studies.
Collapse
Affiliation(s)
- Christine M Pfeiffer
- National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA 30341, USA.
| | | | | | | | | | | | | |
Collapse
|
28
|
Ogbonna G, Gottermeier G, Fox LS. Non-fasting plasma total homocysteine reference interval using the Vitros® homocysteine assay standardized against NIST SRM® 1955. Clin Chim Acta 2008; 388:225-7. [DOI: 10.1016/j.cca.2007.10.016] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/13/2007] [Revised: 10/09/2007] [Accepted: 10/10/2007] [Indexed: 10/22/2022]
|
29
|
Automated immunoassay methods for ferritin: recovery studies to assess traceability to an international standard. Clin Chem Lab Med 2008; 46:1450-7. [DOI: 10.1515/cclm.2008.304] [Citation(s) in RCA: 26] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
|
30
|
|
31
|
Dodder NG, Tai SSC, Sniegoski LT, Zhang NF, Welch MJ. Certification of Creatinine in a Human Serum Reference Material by GC-MS and LC-MS. Clin Chem 2007; 53:1694-9. [PMID: 17660272 DOI: 10.1373/clinchem.2007.090027] [Citation(s) in RCA: 62] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
Abstract
Background: To meet recommendations given by the Laboratory Working Group of the National Kidney Disease Education Program for improving serum creatinine measurements, NIST developed standard reference material (SRM) 967 Creatinine in Frozen Human Serum. SRM 967 is intended for use by laboratories and in vitro diagnostic equipment manufacturers for the calibration and evaluation of routine clinical methods.
Methods: The SRM was produced from 2 serum pools with different creatinine concentrations. The concentrations were certified using a higher-order isotope-dilution GC-MS method and an isotope-dilution LC-MS method. The LC-MS method is a potential higher-order reference measurement procedure.
Results: The GC-MS mean (CV) concentrations were 67.0 (0.9%) μmol/L for serum pool 1 and 346.1 (0.45%) μmol/L for serum pool 2. The LC-MS results were 66.1 (0.2%) μmol/L and 346.3 (0.2%) μmol/L, respectively. For serum pool 1, there was a 1.4% difference between the mean GC-MS and LC-MS measurements, and a 0.10% difference for serum pool 2. The results from the 2 methods were combined to give the certified concentrations and expanded uncertainties.
Conclusions: The certified concentration (expanded uncertainty) of SRM 967 was 66.5 (1.8) μmol/L for serum pool 1 (a value close to the diagnostically important concentration of 88.4 μmol/L) and 346.2 (7.4) μmol/L for serum pool 2 (a concentration corresponding to that expected in a patient with chronic kidney disease).
Collapse
Affiliation(s)
- Nathan G Dodder
- Analytical Chemistry Division and Statistical Engineering Division, National Institute of Standards and Technology, Gaithersburg, MD, USA.
| | | | | | | | | |
Collapse
|
32
|
Rafii M, Elango R, Courtney-Martin G, House JD, Fisher L, Pencharz PB. High-throughput and simultaneous measurement of homocysteine and cysteine in human plasma and urine by liquid chromatography-electrospray tandem mass spectrometry. Anal Biochem 2007; 371:71-81. [PMID: 17727809 DOI: 10.1016/j.ab.2007.07.026] [Citation(s) in RCA: 109] [Impact Index Per Article: 6.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/28/2007] [Revised: 07/17/2007] [Accepted: 07/19/2007] [Indexed: 10/23/2022]
Abstract
Total homocysteine (tHcy) and cysteine (tCys) concentrations in biological fluids are routinely used in the clinical diagnosis of genetic and metabolic diseases, and this necessitates the development of rapid and sensitive methods for quantification. Liquid chromatography-electrospray tandem mass spectrometry (LC-MS/MS) was used to measure tHcy and tCys in 23 plasma and 21 urine samples from healthy adults and 14 urine samples from healthy children. The results were compared with a standard high-performance liquid chromatography (HPLC) method. The coefficient of variation (CV) for the LC-MS/MS method ranged from 2.9% to 6.1% for the intraassay and 4.8% to 6.4% for the interassay. Mean recoveries were close to 100% for both plasma and urinary tHcy and tCys. The mean plasma tHcy and tCys concentrations in healthy adults were 8.62 and 261.40 micromol/L, respectively. The mean urinary tHcy and tCys in adults were 0.98 and 22.60 micromol/mmol creatinine, respectively. The mean urinary tHcy and tCys in children were 1.17 and 27.43 micromol/mmol creatinine, respectively. Bland-Altman difference plots of method comparison between LC-MS/MS and HPLC showed good agreement in plasma and urinary tHcy and tCys concentrations. Our method is suitable for rapid measurements, and the reported urinary values in children will help to develop a pediatric reference range for clinical use.
Collapse
Affiliation(s)
- Mahroukh Rafii
- Research Institute, Hospital for Sick Children, Toronto, ON, Canada M5G 1X8
| | | | | | | | | | | |
Collapse
|
33
|
Fazili Z, Pfeiffer CM, Zhang M. Comparison of serum folate species analyzed by LC-MS/MS with total folate measured by microbiologic assay and Bio-Rad radioassay. Clin Chem 2007; 53:781-4. [PMID: 17272488 DOI: 10.1373/clinchem.2006.078451] [Citation(s) in RCA: 64] [Impact Index Per Article: 3.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
BACKGROUND The Bio-Rad QuantaPhase II radioassay (BR), used for 25 years to measure total folate (TFOL) concentrations for the National Health and Nutrition Examination Survey (NHANES), will be discontinued in 2007. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) or a microbiologic assay (MA) will be used in the future. METHODS We measured folate species by LC-MS/MS and TFOL by MA and BR in 327 serum samples. RESULTS LC-MS/MS measured 5-methyltetrahydrofolic acid (5CH(3)THF; 82%), folic acid (FA; 8%), 5-formyltetrahydrofolic acid (5CHOTHF; 6%), tetrahydrofolic acid (THF; 4%), and 5,10-methenyltetrahydrofolic acid (5,10CH=THF; 0%). The sum of the folate species correlated well with TFOL measured by MA (R(2) = 0.97) and BR (R(2) = 0.91). Compared with LC-MS/MS results, MA and BR values were significantly lower (-6% and -29%, respectively); however, these differences were concentration dependent. The MA almost completely recovered folates added to serum samples except for FA [69% (3%)] and THF [36% (10%)]. The BR underrecovered 5CH(3)THF [61% (9%)] and 5CHOTHF [38% (14%)] and overrecovered 5,10CH=THF [234% (32%)]. Multiple linear regression models with log-transformed data yielded a good fit for converting BR data to MA or LC-MS/MS data and MA data to LC-MS/MS data. CONCLUSIONS The good correspondence between the sum of folate species determined by LC-MS/MS and TFOL determined by MA makes these 2 assays interchangeable. The BR produces much lower results, on average, probably because of 5CH(3)THF underrecovery. The conversion equations provided could be used for future NHANES time trend analyses.
Collapse
Affiliation(s)
- Zia Fazili
- Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, Georgia 30345, USA
| | | | | |
Collapse
|
34
|
Thorpe SJ, Heath A, Blackmore S, Lee A, Hamilton M, O'broin S, Nelson BC, Pfeiffer C. International Standard for serum vitamin B12 and serum folate: international collaborative study to evaluate a batch of lyophilised serum for B12 and folate content. Clin Chem Lab Med 2007; 45:380-6. [PMID: 17378737 DOI: 10.1515/cclm.2007.072] [Citation(s) in RCA: 36] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
BACKGROUND Vitamin B(12) and folate measurements in serum show wide inter-methodology variability. This variability appears to be due in part to the lack of standardisation against internationally accepted reference materials. Pooled human serum, lyophilised in ampoules and designated 03/178, was therefore evaluated by 24 laboratories in seven countries for its suitability to serve as an International Standard (IS) for B(12) and folate. METHODS IS 03/178 was assayed using a range of commercial analysers, microbiological assays and, for folate, candidate reference methods based on liquid chromatography coupled to isotope-dilution tandem mass spectrometry (LC/MS/MS). RESULTS Mean vitamin B(12) and folate values for reconstituted 03/178 across all laboratories and methods were 480 pg/mL [coefficient of variation (CV) 12.8%] and 5.52 ng/mL (CV 17.1%), respectively. The total folate content of reconstituted 03/178, determined using LC/MS/MS, was 12.1 nmol/L (equivalent to 5.33 ng/mL), made up of 9.75 nmol/L 5-methyl tetrahydrofolic acid (5MeTHF; CV 5.5%), 1.59 nmol/L 5-formyl tetrahydrofolic acid (5FoTHF; CV 4.2%) and 0.74 nmol/L folic acid (FA; CV 31.6%). The inclusion of three serum samples in the study with different B(12) and folate levels demonstrated a considerable reduction in inter-laboratory variability when the B(12) and folate content of the samples was determined relative to the IS 03/178 rather than to the analyser calibration. IS 03/178 demonstrated satisfactory long-term stability in accelerated degradation studies. CONCLUSIONS Use of IS 03/178 to standardise serum B(12) and folate assays reduced inter-laboratory variability. The World Health Organization (WHO) Expert Committee on Biological Standardisation established 03/178 as the first IS for serum vitamin B(12) and serum folate, with assigned values of 480 pg/mL of vitamin B(12) and 12.1 nmol/L folate when the lyophilised contents of the ampoule are reconstituted with 1 mL of water.
Collapse
Affiliation(s)
- Susan J Thorpe
- National Institute for Biological Standards and Control, Potters Bar, Hertfordshire, UK.
| | | | | | | | | | | | | | | |
Collapse
|