van Minnen BS, Heesterbeek PJC, Defoort KC, Emans PJ, van Arkel ERA, Struik T, Jutten LM, Susan S, van de Groes SAW, Verdonschot N, van Tienen TG. High failure rate but promising clinical performance after implantation of a flexible medial meniscus prosthesis at 1-year follow-up. Knee Surg Sports Traumatol Arthrosc 2025; 33:1594-1604. [PMID: 39224030 PMCID: PMC12022821 DOI: 10.1002/ksa.12454] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/05/2024] [Revised: 08/23/2024] [Accepted: 08/25/2024] [Indexed: 09/04/2024]
Abstract
PURPOSE After unsatisfactory results in a first-in-human clinical investigation with an anatomically shaped medial meniscus prosthesis, the prosthesis and its fixation technique were altered. This interim analysis of a prospective single-arm clinical investigation aims to evaluate safety and clinical performance in a first-in-human study with the redesigned meniscus prosthesis system. METHODS Ten patients suffering from medial postmeniscectomy pain syndrome were treated with the meniscus prosthesis. Patient-reported outcome measures were obtained at baseline and at 6-week, 3-, 6-, and 12-month follow-up. Radiographs and magnetic resonance imaging scans were obtained to evaluate joint degeneration and prosthesis location. RESULTS The device alterations resolved the issues occurring with the previous design, but four prostheses were explanted after fixation failure or subluxation of the prosthesis. Five out of six patients who reached 1-year follow-up reported a clinically significant improvement of the knee injury and osteoarthritis outcome score pain subscale. Imaging revealed no adverse effects on joint degeneration. CONCLUSION The failure mechanisms that occurred with the previous design have been resolved, but the new fixation technique introduced new safety issues. Improvement of positioning and fixation techniques are considered essential future adaptations to reduce the risk of failure. The good clinical outcomes reported by the patients reaching 1-year follow-up indicate that the medial meniscus prosthesis is a potential solution for patients suffering from postmeniscectomy pain syndrome. LEVEL OF EVIDENCE Level II.
Collapse
Affiliation(s)
- Branco S. van Minnen
- Radboud University Medical CentreNijmegenThe Netherlands
- ATRO Medical B.V.UdenThe Netherlands
| | | | | | - Pieter J. Emans
- Maastricht University Medical CentreMaastrichtThe Netherlands
| | | | - Thijmen Struik
- Radboud University Medical CentreNijmegenThe Netherlands
- ATRO Medical B.V.UdenThe Netherlands
| | | | - Saskia Susan
- Sint Maartenskliniek NijmegenUbbergenThe Netherlands
| | | | | | - Tony G. van Tienen
- Radboud University Medical CentreNijmegenThe Netherlands
- ATRO Medical B.V.UdenThe Netherlands
| |
Collapse
|
2
|
van Hugten PPW, Jeuken RM, Asik EE, Oevering H, Welting TJM, van Donkelaar CC, Thies JC, Emans PJ, Roth AK. In vitro and in vivo evaluation of the osseointegration capacity of a polycarbonate-urethane zirconium-oxide composite material for application in a focal knee resurfacing implant. J Biomed Mater Res A 2024; 112:1424-1435. [PMID: 38465895 DOI: 10.1002/jbm.a.37691] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/25/2023] [Revised: 02/12/2024] [Accepted: 02/14/2024] [Indexed: 03/12/2024]
Abstract
Currently available focal knee resurfacing implants (FKRIs) are fully or partially composed of metals, which show a large disparity in elastic modulus relative to bone and cartilage tissue. Although titanium is known for its excellent osseointegration, the application in FKRIs can lead to potential stress-shielding and metal implants can cause degeneration of the opposing articulating cartilage due to the high resulting contact stresses. Furthermore, metal implants do not allow for follow-up using magnetic resonance imaging (MRI).To overcome the drawbacks of using metal based FKRIs, a biomimetic and MRI compatible bi-layered non-resorbable thermoplastic polycarbonate-urethane (PCU)-based FKRI was developed. The objective of this preclinical study was to evaluate the mechanical properties, biocompatibility and osteoconduction of a novel Bionate® 75D - zirconium oxide (B75D-ZrO2) composite material in vitro and the osseointegration of a B75D-ZrO2 composite stem PCU implant in a caprine animal model. The tensile strength and elastic modulus of the B75D-ZrO2 composite were characterized through in vitro mechanical tests under ambient and physiological conditions. In vitro biocompatibility and osteoconductivity were evaluated by exposing human mesenchymal stem cells to the B75D-ZrO2 composite and culturing the cells under osteogenic conditions. Cell activity and mineralization were assessed and compared to Bionate® 75D (B75D) and titanium disks. The in vivo osseointegration of implants containing a B75D-ZrO2 stem was compared to implants with a B75D stem and titanium stem in a caprine large animal model. After a follow-up of 6 months, bone histomorphometry was performed to assess the bone-to-implant contact area (BIC). Mechanical testing showed that the B75D-ZrO2 composite material possesses an elastic modulus in the range of the elastic modulus reported for trabecular bone. The B75D-ZrO2 composite material facilitated cell mediated mineralization to a comparable extent as titanium. A significantly higher bone-to-implant contact (BIC) score was observed in the B75D-ZrO2 implants compared to the B75D implants. The BIC of B75D-ZrO2 implants was not significantly different compared to titanium implants. A biocompatible B75D-ZrO2 composite approximating the elastic modulus of trabecular bone was developed by compounding B75D with zirconium oxide. In vivo evaluation showed an significant increase of osseointegration for B75D-ZrO2 composite stem implants compared to B75D polymer stem PCU implants. The osseointegration of B75D-ZrO2 composite stem PCU implants was not significantly different in comparison to analogous titanium stem metal implants.
Collapse
Affiliation(s)
- Pieter P W van Hugten
- Department of Orthopedic Surgery, Research School CAPHRI, Maastricht University, Maastricht, The Netherlands
- Department of Orthopedic Surgery, Joint Preservation Clinic, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Ralph M Jeuken
- Department of Orthopedic Surgery, Research School CAPHRI, Maastricht University, Maastricht, The Netherlands
- Department of Orthopedic Surgery, Joint Preservation Clinic, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Erkan E Asik
- Department of Biomedical Engineering, Orthopaedic Biomechanics, Eindhoven University of Technology, Eindhoven, The Netherlands
- Avalanche Medical BV, Maastricht, The Netherlands
| | | | - Tim J M Welting
- Department of Orthopedic Surgery, Research School CAPHRI, Maastricht University, Maastricht, The Netherlands
| | - Corrinus C van Donkelaar
- Department of Biomedical Engineering, Orthopaedic Biomechanics, Eindhoven University of Technology, Eindhoven, The Netherlands
| | | | - Peter J Emans
- Department of Orthopedic Surgery, Research School CAPHRI, Maastricht University, Maastricht, The Netherlands
- Department of Orthopedic Surgery, Joint Preservation Clinic, Maastricht University Medical Center, Maastricht, The Netherlands
- Avalanche Medical BV, Maastricht, The Netherlands
| | - Alex K Roth
- Department of Orthopedic Surgery, Research School CAPHRI, Maastricht University, Maastricht, The Netherlands
- Avalanche Medical BV, Maastricht, The Netherlands
| |
Collapse
|
3
|
Pensado-López A, Ummarino A, Khan S, Guildford A, Allan IU, Santin M, Chevallier N, Varaillon E, Kon E, Allavena P, Torres Andón F. Synthetic peptides of IL-1Ra and HSP70 have anti-inflammatory activity on human primary monocytes and macrophages: Potential treatments for inflammatory diseases. NANOMEDICINE : NANOTECHNOLOGY, BIOLOGY, AND MEDICINE 2024; 55:102719. [PMID: 37977510 DOI: 10.1016/j.nano.2023.102719] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/22/2023] [Revised: 10/05/2023] [Accepted: 10/23/2023] [Indexed: 11/19/2023]
Abstract
Chronic inflammatory diseases are increasing in developed societies, thus new anti-inflammatory approaches are needed in the clinic. Synthetic peptides complexes can be designed to mimic the activity of anti-inflammatory mediators, in order to alleviate inflammation. Here, we evaluated the anti-inflammatory efficacy of tethered peptides mimicking the interleukin-1 receptor antagonist (IL-1Ra) and the heat-shock protein 70 (HSP70). We tested their biocompatibility and anti-inflammatory activity in vitro in primary human monocytes and differentiated macrophages activated with two different stimuli: the TLR agonists (LPS + IFN-γ) or Pam3CSK4. Our results demonstrate that IL-1Ra and HSP70 synthetic peptides present a satisfactory biocompatible profile and significantly inhibit the secretion of several pro-inflammatory cytokines (IL-6, IL-8, IL-1β and TNFα). We further confirmed their anti-inflammatory activity when peptides were coated on a biocompatible material commonly employed in surgical implants. Overall, our findings support the potential use of IL-1Ra and HSP70 synthetic peptides for the treatment of inflammatory conditions.
Collapse
Affiliation(s)
- Alba Pensado-López
- Department of Biomedical Sciences, Humanitas University, 20072 Pieve Emanuele, Milan, Italy.
| | - Aldo Ummarino
- Department of Biomedical Sciences, Humanitas University, 20072 Pieve Emanuele, Milan, Italy; IRCCS Humanitas Research Hospital, Rozzano, 20089, Milan, Italy.
| | - Sophia Khan
- Tissue Click Ltd, The Knoll Business Centre, Old Shoreham Rd, Hove, BN3 7GS, UK.
| | - Anna Guildford
- Tissue Click Ltd, The Knoll Business Centre, Old Shoreham Rd, Hove, BN3 7GS, UK.
| | - Iain U Allan
- Tissue Click Ltd, The Knoll Business Centre, Old Shoreham Rd, Hove, BN3 7GS, UK.
| | - Matteo Santin
- Centre for Regenerative Medicine and Devices, School of Applied Sciences, University of Brighton, Huxley Building Lewes Road, Brighton BN 24GJ, UK.
| | - Nathalie Chevallier
- IMRB, U955, INSERM, Unite d'Ingenierie et de Therapie Cellulaire-Etablissement Français du Sang, Universite Paris-EST Créteil, 94017 Créteil, France.
| | - Elina Varaillon
- IMRB, U955, INSERM, Unite d'Ingenierie et de Therapie Cellulaire-Etablissement Français du Sang, Universite Paris-EST Créteil, 94017 Créteil, France.
| | - Elizaveta Kon
- Department of Biomedical Sciences, Humanitas University, 20072 Pieve Emanuele, Milan, Italy; IRCCS Humanitas Research Hospital, Rozzano, 20089, Milan, Italy.
| | - Paola Allavena
- IRCCS Humanitas Research Hospital, Rozzano, 20089, Milan, Italy.
| | - Fernando Torres Andón
- Department of Biomedical Sciences, Humanitas University, 20072 Pieve Emanuele, Milan, Italy; Instituto de Investigación Biomédica de A Coruña (INIBIC), Medical Oncology Unit, Complexo Hospitalario de A Coruña (CHUAC), 15006 A Coruña, Spain.
| |
Collapse
|
4
|
Singh S, Kumar Paswan K, Kumar A, Gupta V, Sonker M, Ashhar Khan M, Kumar A, Shreyash N. Recent Advancements in Polyurethane-based Tissue Engineering. ACS APPLIED BIO MATERIALS 2023; 6:327-348. [PMID: 36719800 DOI: 10.1021/acsabm.2c00788] [Citation(s) in RCA: 8] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/01/2023]
Abstract
In tissue engineering, polyurethane-based implants have gained significant traction because of their high compatibility and inertness. The implants therefore show fewer side effects and lasts longer. Also, the mechanical properties can be tuned and morphed into a particular shape, owing to which polyurethanes show immense versatility. In the last 3 years, scientists have devised methods to enhance the strength of and induce dynamic properties in polyurethanes, and these developments offer an immense opportunity to use them in tissue engineering. The focus of this review is on applications of polyurethane implants for biomedical application with detailed analysis of hard tissue implants like bone tissues and soft tissues like cartilage, muscles, skeletal tissues, and blood vessels. The synthetic routes for the preparation of scaffolds have been discussed to gain a better understanding of the issues that arise regarding toxicity. The focus here is also on concerns regarding the biocompatibility of the implants, given that the precursors and byproducts are poisonous.
Collapse
Affiliation(s)
- Sukriti Singh
- Department of Chemical and Biochemical Engineering, Rajiv Gandhi Institute of Petroleum Technology, Mubarakpur Mukhatiya, Uttar Pradesh 229304, India
| | - Karan Kumar Paswan
- Department of Chemical and Biochemical Engineering, Rajiv Gandhi Institute of Petroleum Technology, Mubarakpur Mukhatiya, Uttar Pradesh 229304, India
| | - Alok Kumar
- Department of Chemical and Biochemical Engineering, Rajiv Gandhi Institute of Petroleum Technology, Mubarakpur Mukhatiya, Uttar Pradesh 229304, India
| | - Vishwas Gupta
- Department of Petroleum Engineering, Rajiv Gandhi Institute of Petroleum Technology, Mubarakpur Mukhatiya, Uttar Pradesh 229304, India
| | - Muskan Sonker
- Department of Chemical and Biomolecular Engineering, Georgia Institute of Technology, Atlanta, Georgia 30332, United States
| | - Mohd Ashhar Khan
- Department of Chemical and Biological Engineering, State University of New York at Buffalo, Buffalo, New York 14260, United States
| | - Amrit Kumar
- Indian Oil Corporation Limited, Panipat Refinery, Panipat, Odisha 132140, India
| | - Nehil Shreyash
- Department of Chemical and Biomolecular Engineering, University of Houston, Houston, Texas 77204, United States
| |
Collapse
|
5
|
Kanca Y, Özkahraman B. An investigation on tribological behavior of methacrylated κ-carrageenan and gellan gum hydrogels as a candidate for chondral repair. J Biomater Appl 2023; 37:1271-1285. [PMID: 36473707 DOI: 10.1177/08853282221144235] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/12/2022]
Abstract
Natural polysaccharides have recently attracted attention as structural biomaterials to replace focal chondral defects. In the present study, in-vitro tribological performance of methacrylated κ-carrageenan and gellan gum hydrogels (KA-MA and GG-MA) was evaluated under physiological conditions. Coefficient of friction (COF) was continuously recorded over testing whilst worn area was measured post-testing. The findings help improve our understanding of KA-MA-H and GG-MA-H tribological performance under various physiological conditions. The friction and wear performance of the hydrogels improved in bovine calf serum lubricant at lower applied loads. Adhesion was the dominant wear mechanism detected by SEM. Among the proposed hydrogels GG-MA-H found robust mechanical properties, increased wear resistance and considerably low COF, which may suggest its potential usage as a cartilage substitute.
Collapse
Affiliation(s)
- Yusuf Kanca
- Department of Mechanical Engineering, Faculty of Engineering, 162313Hitit University, Çorum, Turkey
| | - Bengi Özkahraman
- Department of Polymer Materials Engineering, Faculty of Engineering, 162313Hitit University, Çorum, Turkey
| |
Collapse
|
6
|
Coefficient of Friction and Height Loss: Two Criteria Used to Determine the Mechanical Property and Stability of Regenerated Versus Natural Articular Cartilage. Biomedicines 2022; 10:biomedicines10112685. [DOI: 10.3390/biomedicines10112685] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2022] [Revised: 09/28/2022] [Accepted: 10/12/2022] [Indexed: 11/17/2022] Open
Abstract
Background: The coefficient of friction (CoF) serves as an indicator for the mechanical properties of natural and regenerated articular cartilage (AC). After tribological exposure, a height loss (HL) of the cartilage pair specimens can be measured. Our aim was to determine the CoF and HL of regenerated AC tissue and compare them with those of natural AC from non-operated joints and AC from joints where the regenerated tissues had been created after different treatments. Methods: In partial-thickness defects of the trochleae of the stifle joints of 60 Göttingen Minipigs, regenerated AC was created. In total, 40 animals received a Col I matrix, 20 laden with autologous chondrocytes, and 20 without. The defects of 20 animals were left empty. The healing periods were 24 and 48 weeks. A total of 10 not-operated animals, delivered the “external” control specimens. Osteochondral pins were harvested from defect and non-defect areas, the latter serving as “internal” controls. Using a pin-on-plate tribometer, we measured the CoF and the HL. Results: The CoF of the regenerated AC ranged from 0.0393 to 0.0688, and the HL, from 0.22 mm to 0.3 mm. The differences between the regenerated AC of the six groups and the “external” controls were significant. The comparison with the “internal” controls revealed four significant differences for the CoF and one for the HL in the operated groups. No differences were seen within the operated groups. Conclusions: The mechanical quality of the regenerated AC tissue showed inferior behavior with regard to the CoF and HL in comparison with natural AC. The comparison of regenerated AC tissue with AC from untreated joints was more promising than with AC from the treated joints.
Collapse
|
7
|
Vanaclocha-Saiz A, Vanaclocha V, Atienza C, Jorda-Gomez P, Primo-Capella V, Barrios C, Vanaclocha L. Bionate Biocompatibility: In Vivo Study in Rabbits. ACS OMEGA 2022; 7:29647-29654. [PMID: 36061708 PMCID: PMC9435029 DOI: 10.1021/acsomega.2c01690] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 03/20/2022] [Accepted: 08/10/2022] [Indexed: 06/15/2023]
Abstract
Response to foreign materials includes local tissue reaction, osteolysis, implant loosening, and migration to lymph nodes and organs. Bionate 80A human explants show minor wear and slight local tissue reaction, but we do not know the response at the spinal cord, nerve roots, lymph nodes, or distant organs. This study aims to figure out reactions against Bionate 80A when implanted at the spinal epidural space of 24 20-week-old New Zealand white rabbits. In one group of 12 rabbits, we implanted Bionate 80A on the spinal epidural space, and another group of 12 rabbits was used as the control group. We studied tissues, organs, and tissue damage markers on blood biochemistry, urine tests, and necropsy. The animals' clinical parameters and weight showed no statistically significant differences. At 3 months, the basophils increased slightly in the implant group, platelets decreased in all, and at 6 months, implanted animals showed slight eosinophilia, but none of these changes was statistically significant. External, organ, and spinal tissue examination showed neither toxic reaction, inflammatory changes, or noticeable differences between groups or survival periods. Under microscopic examination, the Bionate 80A particles induced a chronic granulomatous response always outside the dura mater, with giant multinucleated cells holding phagocytized particles and no particle migration to lymph nodes or organs. Thus, it was concluded that Bionate particles, when implanted in the rabbit lumbar epidural space, do not generate a significant reaction limited to the surrounding soft tissues with giant multinucleated cells. In addition, the particles did not cross the dura mater or migrate to lymph nodes or organs.
Collapse
Affiliation(s)
- Amparo Vanaclocha-Saiz
- Instituto de Biomecánica (IBV), Universitat Politècnica de Valencia, Valencia 46022, Spain
| | | | - Carlos Atienza
- Instituto de Biomecánica (IBV), Universitat Politècnica de Valencia, Valencia 46022, Spain
| | - Pablo Jorda-Gomez
- Hospital General Universitario de Castellón, Castellón de la Plana 12004, Spain
| | - Víctor Primo-Capella
- Instituto de Biomecánica (IBV), Universitat Politècnica de Valencia, Valencia 46022, Spain
| | - Carlos Barrios
- Catholic University of Valencia, Saint Vincent Martyr, Valencia 46001, Spain
| | - Leyre Vanaclocha
- Medius Klinik, Ostfildern-Ruit Klinik für Urologie, Hedelfinger Strasse 166, 73760 Ostfildern, Germany
| |
Collapse
|
8
|
Oliveira AS, Ferreira I, Branco AC, Silva JC, Costa C, Nolasco P, Marques AC, Silva D, Colaço R, Figueiredo-Pina CG, Serro AP. Development of polycarbonate urethane-based materials with controlled diclofenac release for cartilage replacement. J Biomed Mater Res B Appl Biomater 2022; 110:1839-1852. [PMID: 35226412 DOI: 10.1002/jbm.b.35042] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/21/2021] [Revised: 12/04/2021] [Accepted: 02/09/2022] [Indexed: 11/05/2022]
Abstract
Hydrogels are very promising human cartilage replacement materials since they are able to mimic its structure and properties. Besides, they can be used as platforms for drug delivery to reduce inflammatory postsurgical reactions. Polycarbonate urethane (PCU) has been used in orthopedic applications due to its long-term biocompatibility and bio-durability. In this work, PCU-based hydrogels with the ability to release an anti-inflammatory (diclofenac) were developed, for the first time, for such purpose. The materials were reinforced with different amounts of cellulose acetate (CA, 10%, 15%, and 25% w/w) or carbon nanotubes (CNT, 1% and 2% w/w) in order to improve their mechanical properties. Samples were characterized in terms of compressive and tensile mechanical behavior. It was found that 15% CA and 2% CNT reinforcement led to the best mechanical properties. Thus, these materials were further characterized in terms of morphology, wettability, and friction coefficient (CoF). Contrarily to CNTs, the addition of CA significantly increased the material's porosity. Both materials became more hydrophilic, and the CoF slightly increased for PCU + 15%CA. The materials were loaded by soaking with diclofenac, and drug release experiments were conducted. PCU, PCU + 15%CA and PCU + 2%CNT presented similar release profiles, being able to ensure a controlled release of DFN for at least 4 days. Finally, in vitro cytotoxicity tests using human chondrocytes were also performed and confirmed a high biocompatibility for the three studied materials.
Collapse
Affiliation(s)
- Andreia S Oliveira
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,IDMEC e Departamento de Engenharia Mecânica, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal
| | - Inês Ferreira
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal
| | - Ana C Branco
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal.,CDP2T, Escola Superior de Tecnologia de Setúbal, Instituto Politécnico de Setúbal, Setúbal, Portugal
| | - João C Silva
- IBB - Instituto de Bioengenharia e Biociências e Departamento de Bioengenharia, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Carolina Costa
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Pedro Nolasco
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Ana C Marques
- CERENA, DEQ, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Diana Silva
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal
| | - Rogério Colaço
- IDMEC e Departamento de Engenharia Mecânica, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Célio G Figueiredo-Pina
- CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal.,CDP2T, Escola Superior de Tecnologia de Setúbal, Instituto Politécnico de Setúbal, Setúbal, Portugal.,CeFEMA, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal
| | - Ana P Serro
- CQE, Instituto Superior Técnico, Universidade de Lisboa, Lisbon, Portugal.,CiiEM, Escola Superior de Saúde Egas Moniz, Monte de Caparica, Portugal
| |
Collapse
|
9
|
Engelhardt JP, Schütte A, Hetjens S, Reisig G, Schwarz ML. Resilience to height loss of articular cartilage of osteoarthritic stifle joints of old pigs, compared with healthy cartilage from young pigs in a tribological pin-on-plate exposure, revealing similar friction forces. PLoS One 2021; 16:e0250244. [PMID: 33891624 PMCID: PMC8064609 DOI: 10.1371/journal.pone.0250244] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/07/2020] [Accepted: 04/04/2021] [Indexed: 12/04/2022] Open
Abstract
Introduction We saw a lack of data on the biomechanical behavior of degenerated articular cartilage (OA) compared with that of healthy cartilage, even though the susceptibility to wear and tear of articular cartilage plays a key role in the progression of osteoarthritis (OA). Therefore, we performed a comparison between naturally occurring OA and healthy cartilage from pigs, before and after tribological stress. Aim The aim of the study was to compare OA-cartilage with healthy cartilage and to analyze the resilience to tribological shear stress, which will be measured as height loss (HL), and to friction forces of the cartilage layers. The findings will be substantiated in macro- and microscopical evaluations before and after tribological exposure. Methods We assessed stifle joints of fifteen old and sixteen young pigs from the local abattoir radiologically, macroscopically and histologically to determine possible OA alterations. We put pins from the femoral part of the joints and plates from the corresponding tibial plateaus in a pin-on-plate tribometer under stress for about two hours with about 1108 reciprocating cycles under a pressure of approximately 1 MPa. As a surrogate criterion of wear and tear, the HL was recorded in the tribometer. The heights of the cartilage layers measured before and after the tribological exposure were compared histologically. The condition of the cartilage before and after the tribological exposure was analyzed both macroscopically with an adapted ICRS score and microscopically according to Little et al. (2010). We assessed the friction forces acting between the surfaces of the cartilage pair–specimens. Results Articular cartilage taken from old pigs showed significant degenerative changes compared to that taken from the young animals. The macroscopic and microscopic scores showed strong alterations of the cartilage after the tribological exposure. There was a noticeable HL of the cartilage specimens after the first 100 to 300 cycles. The HL after tribological exposure was lower in the group of the old animals with 0.52 mm ± 0.23 mm than in the group of the young animals with 0.86 mm ± 0.26 mm (p < 0.0001). The data for the HL was validated by the histological height measurements with 0.50 mm ± 0.82 mm for the old and 0.79 mm ±0.53 mm for the young animals (p = 0.133). The friction forces measured at the cartilage of the old animals were 2.25 N ± 1.15 N and 1.89 N ± 1.45 N of the young animals (p = 0.3225). Conclusion Unlike articular cartilage from young pigs, articular cartilage from old pigs showed OA alterations. Tribological shear stress exposure revealed that OA cartilage showed less HL than healthy articular cartilage. Tribological stress exposure in a pin–on–plate tribometer seemed to be an appropriate way to analyze the mechanical stability of articular cartilage, and the applied protocol could reveal weaknesses of the assessed cartilage tissue. Friction and HL seemed to be independent parameters when degenerated and healthy articular cartilage were assessed under tribological exposure in a pin–on- plate tribometer.
Collapse
Affiliation(s)
- Jan P. Engelhardt
- Department of Experimental Orthopedics and Trauma Surgery, Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
| | - Andy Schütte
- Department of Experimental Orthopedics and Trauma Surgery, Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
| | - Svetlana Hetjens
- Department of Medical Statistics, University Medicine Mannheim, Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
| | - Gregor Reisig
- Department of Experimental Orthopedics and Trauma Surgery, Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
| | - Markus L. Schwarz
- Department of Experimental Orthopedics and Trauma Surgery, Medical Faculty Mannheim of the University of Heidelberg, Mannheim, Germany
- * E-mail:
| |
Collapse
|
10
|
Ford A, Hua Z, Ferguson SJ, Pruitt LA, Gao L. A 3D-transient elastohydrodynamic lubrication hip implant model to compare ultra high molecular weight polyethylene with more compliant polycarbonate polyurethane acetabular cups. J Mech Behav Biomed Mater 2021; 119:104472. [PMID: 33813334 DOI: 10.1016/j.jmbbm.2021.104472] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/14/2020] [Revised: 10/20/2020] [Accepted: 03/14/2021] [Indexed: 10/21/2022]
Abstract
Wear remains a significant challenge in the design of orthopedic implants such as total hip replacements. Early elastohydrodynamic lubrication modeling has predicted thicker lubrication films in hip replacement designs with compliant polycarbonate polyurethane (PCU) bearing materials compared to stiffer materials like ultra-high molecular weight polyethylene (UHMWPE). The predicted thicker lubrication films suggest improved friction and wear performance. However, when compared to the model predictions, experimental wear studies showed mixed results. The mismatch between the model and experimental results may lie in the simplifying assumptions of the early models such as: steady state conditions, one dimensional rotation and loading, and high viscosities. This study applies a 3D-transient elastohydrodynamic model based on an ISO standard gait cycle to better understand the interaction between material stiffness and film thickness in total hip arthroplasty material couples. Similar to previous, simplified models, we show that the average and central film thickness of PCU (∼0.4μm) is higher than that of UHMWPE (∼0.2μm). However, in the 3D-transient model, the film thickness distribution was largely asymmetric and the minimum film thickness occurred outside of the central axis. Although the overall film thickness of PCU was higher than UHMWPE, the minimum film thickness of PCU was lower than UHMPWE for the majority of the gait cycle. The minimum film thickness of PCU also had a larger range throughout the gait cycle. Both materials were found to be operating between boundary and mixed lubrication regimes. This 3D-transient model reveals a more nuanced interaction between bearing material stiffness and film thickness that supports the mixed results found in experimental wear studies of PCU hip implant designs.
Collapse
Affiliation(s)
- Audrey Ford
- Department of Mechanical Engineering, 6141 Etcheverry Hall, Mail Code 1740, University of California Berkeley, Berkeley, CA 94720-1740, USA
| | - Zikai Hua
- School of Mechatronics Engineering and Automation, Shanghai University, 266 Jufengyuan Rd, Baoshan, Shanghai, China
| | - Stephen J Ferguson
- Institute for Biomechanics, ETH Zurich, Leopold-Ruzicka-Weg 4, 8093 Zurich, Switzerland
| | - Lisa A Pruitt
- Department of Mechanical Engineering, 6141 Etcheverry Hall, Mail Code 1740, University of California Berkeley, Berkeley, CA 94720-1740, USA
| | - Leiming Gao
- Engineering Department, School of Science and Technology, Nottingham Trent University, 50 Shakespeare Street, Nottingham, NG1 4FG, United Kingdom.
| |
Collapse
|
11
|
Link JM, Salinas EY, Hu JC, Athanasiou KA. The tribology of cartilage: Mechanisms, experimental techniques, and relevance to translational tissue engineering. Clin Biomech (Bristol, Avon) 2020; 79:104880. [PMID: 31676140 PMCID: PMC7176516 DOI: 10.1016/j.clinbiomech.2019.10.016] [Citation(s) in RCA: 21] [Impact Index Per Article: 4.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 12/14/2018] [Revised: 10/03/2019] [Accepted: 10/17/2019] [Indexed: 02/07/2023]
Abstract
Diarthrodial joints, found at the ends of long bones, function to dissipate load and allow for effortless articulation. Essential to these functions are cartilages, soft hydrated tissues such as hyaline articular cartilage and the knee meniscus, as well as lubricating synovial fluid. Maintaining adequate lubrication protects cartilages from wear, but a decrease in this function leads to tissue degeneration and pathologies such as osteoarthritis. To study cartilage physiology, articular cartilage researchers have employed tribology, the study of lubrication and wear between two opposing surfaces, to characterize both native and engineered tissues. The biochemical components of synovial fluid allow it to function as an effective lubricant that exhibits shear-thinning behavior. Although tribological properties are recognized to be essential to native tissue function and a critical characteristic for translational tissue engineering, tribology is vastly understudied when compared to other mechanical properties such as compressive moduli. Further, tribometer configurations and testing modalities vary greatly across laboratories. This review aims to define commonly examined tribological characteristics and discuss the structure-function relationships of biochemical constituents known to contribute to tribological properties in native tissue, address the variations in experimental set-ups by suggesting a move toward standard testing practices, and describe how tissue-engineered cartilages may be augmented to improve their tribological properties.
Collapse
Affiliation(s)
- Jarrett M. Link
- 3131 Engineering Hall, Department of Biomedical Engineering, University of California, Irvine, California 92617, USA
| | - Evelia Y. Salinas
- 3131 Engineering Hall, Department of Biomedical Engineering, University of California, Irvine, California 92617, USA
| | - Jerry C. Hu
- 3131 Engineering Hall, Department of Biomedical Engineering, University of California, Irvine, California 92617, USA
| | - Kyriacos A. Athanasiou
- 3131 Engineering Hall, Department of Biomedical Engineering, University of California, Irvine, California 92617, USA
| |
Collapse
|
12
|
Berkmortel C, Langohr GDG, King G, Johnson J. Hemiarthroplasty implants should have very low stiffness to optimize cartilage contact stress. J Orthop Res 2020; 38:1719-1726. [PMID: 32017162 DOI: 10.1002/jor.24610] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/12/2019] [Revised: 11/15/2019] [Accepted: 01/21/2020] [Indexed: 02/04/2023]
Abstract
Hemiarthroplasty is often preferred to total arthroplasty as it preserves native tissue; however, accelerated wear of the opposing cartilage is problematic. This is thought to be caused by the stiffness mismatch between the implant and cartilage-bone construct. Reducing the stiffness of the implant by changing the material has been hypothesized as a potential solution. This study employs a finite element model to study a concave-convex hemiarthroplasty articulation for various implant materials (cobalt-chrome, pyrolytic carbon, polyether ether ketone, ultra-high-molecular-weight polyethylene, Bionate-55D, Bionate-75D, and Bionate-80A). The effect of the radius of curvature and the degree of flexion-extension was also investigated to ensure any relationships found between materials were generalizable. The implant material had a significant effect (P < .001) for both contact area and maximum contact pressure on the cartilage surface. All of the materials were different from the native state except for Bionate-80A at two of the different flexion angles. Bionate-80A and Bionate-75D, the materials with the lowest stiffnesses, were the closest to the native state for all flexion angles and radii of curvature. No evident difference between materials occurred unless the modulus was below that of Bionate-55D (288 MPa), suggesting that hemiarthroplasty materials need to be less stiff than this material if they are to protect the opposing cartilage. This is clinically significant as the findings suggest that the development of new hemiarthroplasty implants should use materials with stiffnesses much lower than currently available devices.
Collapse
Affiliation(s)
- Carolyn Berkmortel
- Department of Mechanical and Materials Engineering, Western University, London, Ontario, Canada
| | - G Daniel G Langohr
- Department of Mechanical and Materials Engineering, Western University, London, Ontario, Canada.,Bioengineering Laboratory, Roth McFarlane Hand and Upper Limb Centre, Lawson Health Research Institute, St Joseph's Health Care, Western University, Bone and Joint Institute, London, Ontario, Canada.,Department of Surgery, Western University, London, Ontario, Canada
| | - Graham King
- Bioengineering Laboratory, Roth McFarlane Hand and Upper Limb Centre, Lawson Health Research Institute, St Joseph's Health Care, Western University, Bone and Joint Institute, London, Ontario, Canada.,Department of Surgery, Western University, London, Ontario, Canada
| | - James Johnson
- Department of Mechanical and Materials Engineering, Western University, London, Ontario, Canada.,Bioengineering Laboratory, Roth McFarlane Hand and Upper Limb Centre, Lawson Health Research Institute, St Joseph's Health Care, Western University, Bone and Joint Institute, London, Ontario, Canada.,Department of Surgery, Western University, London, Ontario, Canada
| |
Collapse
|
13
|
Jeuken RM, Roth AK, Peters MJM, Welting TJM, van Rhijn LW, Koenen J, Peters RJRW, Thies JC, Emans PJ. In vitro and in vivo study on the osseointegration of BCP-coated versus uncoated nondegradable thermoplastic polyurethane focal knee resurfacing implants. J Biomed Mater Res B Appl Biomater 2020; 108:3370-3382. [PMID: 32614486 PMCID: PMC7586808 DOI: 10.1002/jbm.b.34672] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/25/2019] [Revised: 04/18/2020] [Accepted: 06/03/2020] [Indexed: 11/10/2022]
Abstract
Focal knee resurfacing implants (FKRIs) are intended to treat cartilage defects in middle-aged patients. Most FKRIs are metal-based, which hampers follow-up of the joint using magnetic resonance imaging and potentially leads to damage of the opposing cartilage. The purpose of this study was to develop a nondegradable thermoplastic polyurethane (TPU) FKRI and investigate its osseointegration. Different surface roughness modifications and biphasic calcium phosphate (BCP) coating densities were first tested in vitro on TPU discs. The in vivo osseointegration of BCP-coated TPU implants was subsequently compared to uncoated TPU implants and the titanium bottom layer of metal control implants in a caprine model. Implants were implanted bilaterally in stifle joints and animals were followed for 12 weeks, after which the bone-to-implant contact area (BIC) was assessed. Additionally, 18F-sodium-fluoride (18F-NaF) positron emission tomography PET/CT-scans were obtained at 3 and 12 weeks to visualize the bone metabolism over time. The BIC was significantly higher for the BCP-coated TPU implants compared to the uncoated TPU implants (p = .03), and did not significantly differ from titanium (p = .68). Similar 18F-NaF tracer uptake patterns were observed between 3 and 12 weeks for the BCP-coated TPU and titanium implants, but not for the uncoated implants. TPU FKRIs with surface modifications could provide the answer to the drawbacks of metal FKRIs.
Collapse
Affiliation(s)
- Ralph M Jeuken
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Alex K Roth
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Marloes J M Peters
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Tim J M Welting
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Lodewijk W van Rhijn
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| | - Jac Koenen
- DSM Biomedical BV, Geleen, The Netherlands
| | | | | | - Pieter J Emans
- Department of Orthopaedic Surgery, Research School CAPHRI, Maastricht University Medical Center, Maastricht, The Netherlands
| |
Collapse
|
14
|
Ajdari N, Tempelaere C, Masouleh MI, Abel R, Delfosse D, Emery R, Dini D, Hansen U. Hemiarthroplasties: the choice of prosthetic material causes different levels of damage in the articular cartilage. J Shoulder Elbow Surg 2020; 29:1019-1029. [PMID: 31948834 DOI: 10.1016/j.jse.2019.09.041] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/17/2019] [Revised: 09/24/2019] [Accepted: 09/30/2019] [Indexed: 02/01/2023]
Abstract
BACKGROUND Hemiarthroplasty has clear advantages over alternative procedures and is used in 20% of all shoulder joint replacements. Because of cartilage wear, the clinical outcome of hemiarthroplasty is unreliable and controversial. This paper suggests that the optimal choice of prosthetic material may reduce cartilage degeneration and improve the reliability of the procedure. The specific objectives were to assess 3 materials and assess how the severity of arthritis might affect the choice of prosthetic material. METHODS A CoCr alloy, an AL2O3 ceramic, and a polycarbonate urethane polymer (PCU) were mechanically tested against 5 levels of human osteoarthritic cartilage (from intact to severely arthritic, n = 45). A high friction coefficient, a decrease in Young's modulus, an increase in permeability, a decrease in relaxation time, an increase in surface roughness, and a disrupted appearance of the cartilage after testing were used as measures of cartilage damage. The biomaterial that caused minimal cartilage damage was defined as superior. RESULTS The CoCr caused the most damage. This was followed by the AL2O3 ceramic, whereas the PCU caused the least amount of damage. Although the degree of arthritis had an effect on the results, it did not change the trend that CoCr performed worst and PCU the best. DISCUSSION AND CONCLUSION This study indicates that ceramic implants may be a better choice than metals, and the articulating surface should be as smooth as possible. Although our results indicate that the degree of arthritis should not affect the choice of prosthetic material, this suggestion needs to be further investigated.
Collapse
Affiliation(s)
- Niloofar Ajdari
- Department of Mechanical Engineering, Imperial College London, South Kensington Campus, London, UK
| | - Christine Tempelaere
- Department of Mechanical Engineering, Imperial College London, South Kensington Campus, London, UK; Department of Orthopaedic Surgery, Bicetre Hospital, Le Kremlin-Bicêtre, France
| | - Maryam Imani Masouleh
- Department of Mechanical Engineering, Imperial College London, South Kensington Campus, London, UK
| | - Richard Abel
- Department of Surgery and Cancer, Charing Cross Campus, London, UK
| | | | - Roger Emery
- Department of Surgery and Cancer, St. Mary's Hospital, London, UK
| | - Daniele Dini
- Department of Mechanical Engineering, Imperial College London, South Kensington Campus, London, UK
| | - Ulrich Hansen
- Department of Mechanical Engineering, Imperial College London, South Kensington Campus, London, UK.
| |
Collapse
|