Fluhr JW, Wiora G, Nikolaeva DG, Miséry L, Darlenski R. In vivo transepidermal water loss: Validation of a new multi-sensor open chamber water evaporation system Tewameter TM Hex. Skin Res Technol 2023; 29:e13307. [PMID: 37113085 PMCID: PMC10234168 DOI: 10.1111/srt.13307] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/12/2023] [Accepted: 03/02/2023] [Indexed: 04/03/2023]
Abstract
BACKGROUND Instrumentation technology for transepidermal water loss measurements has not been substantially modified since its introduction by Nilsson in 1977. Recent progress in sensor development allowed a new sensor arrangement using a matrix of 30 sensors. Raw measurement values are processed with spatial statistical analysis. We aimed to compare the new, multi-sensor probe (Tewameter TM Hex) with the established Tewameter TM 300 probe and to gain reference data for the new parameters of transepidermal energy loss and water vapor concentration on skin. MATERIAL AND METHODS Baseline measurements and repeated measurements on the volar forearm and assessment on eight different anatomical locations were performed on 24 healthy volunteers (both gender) with the TM Hex and the TM 300. RESULTS A significant correlation (p < 0.001; R-coefficient = 0.9) between TM Hex and the TM 300 with a low coefficient of variance (CV) 11% for TM Hex and 19% for TM 300, could be assessed. The CV ranged between 7% (right inner upper arm) and 14% (palms). Average transepidermal heat loss ranged from 12 W/m2 on the lower leg to 38.8 W/m2 on the palm. CONCLUSION The correlation between TM Hex and TM 300 along with the robustness of the measurements with the TM Hex shows that the new probe for assessment of epidermal barrier function is comparable to the TM 300. In most conditions, TM Hex provides more accurate measurements than TM 300. New parameters open the field to studying skin's water and energy balance.
Collapse
Affiliation(s)
- Joachim W. Fluhr
- Charité – Universitätsmedizin BerlinInstitute of AllergologyBerlinGermany
- Fraunhofer Institute for Translational Medicine and Pharmacology ITMPAllergology and ImmunologyGermany
| | - Georg Wiora
- Courage + Khazaka electronic GmbHCologneGermany
| | - Dessyslava G. Nikolaeva
- Department of Dermatology and VenereologyAcibadem City Clinic Tokuda Hospital, SofiaBulgaria
- EuroDerma ClinicSofiaBulgaria
| | | | - Razvigor Darlenski
- Department of Dermatology and VenereologyAcibadem City Clinic Tokuda Hospital, SofiaBulgaria
- Department of Dermatology and VenereologyMedical FacultyTrakia University, Stara ZagoraBulgaria
| |
Collapse
|
2
|
Anushree U, Shetty S, Kumar R, Bharati S. Adjunctive Diagnostic Methods for Skin Cancer Detection: A Review of Electrical Impedance-Based Techniques. Bioelectromagnetics 2022; 43:193-210. [PMID: 35181899 DOI: 10.1002/bem.22396] [Citation(s) in RCA: 4] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/21/2021] [Revised: 12/06/2021] [Accepted: 02/04/2022] [Indexed: 12/15/2022]
Abstract
Skin cancer is among the fastest-growing cancers with an excellent prognosis, if detected early. However, the current method of diagnosis by visual inspection has several disadvantages such as overlapping tumor characteristics, subjectivity, low sensitivity, and specificity. Hence, several adjunctive diagnostic techniques such as thermal imaging, optical imaging, ultrasonography, tape stripping methods, and electrical impedance imaging are employed along with visual inspection to improve the diagnosis. Electrical impedance-based skin cancer detection depends upon the variations in electrical impedance characteristics of the transformed cells. The information provided by this technique is fundamentally different from other adjunctive techniques and thus has good prospects. Depending on the stage, type, and location of skin cancer, various impedance-based devices have been developed. These devices when used as an adjunct to visual methods have increased the sensitivity and specificity of skin cancer detection up to 100% and 87%, respectively, thus demonstrating their potential to minimize unnecessary biopsies. In this review, the authors track the advancements and progress made in this technique for the detection of skin cancer, focusing mainly on the advantages and limitations in the clinical setting. © 2022 Bioelectromagnetics Society.
Collapse
Affiliation(s)
- U Anushree
- Department of Nuclear Medicine, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, Karnataka, India
| | - Sachin Shetty
- Department of Nuclear Medicine, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, Karnataka, India
| | - Rajesh Kumar
- Department of Nuclear Medicine, All India Institute of Medical Sciences, Jodhpur, Rajasthan, India
| | - Sanjay Bharati
- Department of Nuclear Medicine, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, Karnataka, India
| |
Collapse
|
3
|
Chaturvedi P, Worsley PR, Zanelli G, Kroon W, Bader DL. Quantifying skin sensitivity caused by mechanical insults: A review. Skin Res Technol 2021; 28:187-199. [PMID: 34708455 PMCID: PMC9298205 DOI: 10.1111/srt.13104] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/09/2021] [Accepted: 08/21/2021] [Indexed: 11/29/2022]
Abstract
BACKGROUND Skin sensitivity (SS) is a commonly occurring response to a range of stimuli, including environmental conditions (e.g., sun exposure), chemical irritants (e.g., soaps and cosmetics), and mechanical forces (e.g., while shaving). From both industry and academia, many efforts have been taken to quantify the characteristics of SS in a standardised manner, but the study is hindered by the lack of an objective definition. METHODS A review of the scientific literature regarding different parameters attributed to the loss of skin integrity and linked with exhibition of SS was conducted. Articles included were screened for mechanical stimulation of the skin, with objective quantification of tissue responses using biophysical or imaging techniques. Additionally, studies where cohorts of SS and non-SS individuals were reported have been critiqued. RESULTS The findings identified that the structure and function of the stratum corneum and its effective barrier properties are closely associated with SS. Thus, an array of skin tissue responses has been selected for characterization of SS due to mechanical stimuli, including: transepidermal water loss, hydration, redness, temperature, and sebum index. Additionally, certain imaging tools allow quantification of the superficial skin layers, providing structural characteristics underlying SS. CONCLUSION This review proposes a multimodal approach for identification of SS, providing a means to characterise skin tissue responses objectively. Optical coherence tomography (OCT) has been suggested as a suitable tool for dermatological research with clinical applications. Such an approach would enhance the knowledge underlying the multifactorial nature of SS and aid the development of personalised solutions in medical and consumer devices.
Collapse
Affiliation(s)
- Pakhi Chaturvedi
- Philips Consumer Lifestyle B.V., Drachten, The Netherlands.,School of Health Sciences, University of Southampton, Southampton, UK
| | - Peter R Worsley
- School of Health Sciences, University of Southampton, Southampton, UK
| | - Giulia Zanelli
- Philips Consumer Lifestyle B.V., Drachten, The Netherlands
| | - Wilco Kroon
- Philips Consumer Lifestyle B.V., Drachten, The Netherlands
| | - Dan L Bader
- School of Health Sciences, University of Southampton, Southampton, UK
| |
Collapse
|
4
|
Antioxidant Activity of Hydrogen Water Mask Pack Composed of Gel-Type Emulsion and Hydrogen Generation Powder. Int J Mol Sci 2020; 21:ijms21249731. [PMID: 33419292 PMCID: PMC7766410 DOI: 10.3390/ijms21249731] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/24/2020] [Revised: 12/14/2020] [Accepted: 12/18/2020] [Indexed: 01/19/2023] Open
Abstract
In this study, hydrogen generation powder samples were prepared using zinc carbonate as a precursor, at a temperature varying from 400 to 700 °C in H2 atmosphere, and were characterized in terms of antioxidant activity. The concentration of dissolved hydrogen obtained by the powder samples was measured using a dissolved hydrogen meter as a function of time. In addition, the antioxidant activity of the samples was evaluated based on the Oyaizu’s method, removal rate of
·OH radicals, and ferric reducing antioxidant power. Finally, the hydrogen mask pack was fabricated using the hydrogen generation powder sample and gel-type emulsion. In the clinical test on the mask pack, the effect of the mask on skin aging was characterized and compared to that of a commercial sample. The skin densities of the participants in the experimental group and the control group increased by 18.41% and 9.93% after 4 weeks, respectively. The improved skin density of the participants who used the hydrogen mask pack in the experimental group, might be attributed to the recovery effect of the hydrogen molecule in the mask pack on the denatured thick skin layer.
Collapse
|
5
|
Logger JGM, Driessen RJB, de Jong EMGJ, van Erp PEJ. Value of GPSkin for the measurement of skin barrier impairment and for monitoring of rosacea treatment in daily practice. Skin Res Technol 2020; 27:15-23. [PMID: 32573826 PMCID: PMC7984125 DOI: 10.1111/srt.12900] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/03/2020] [Accepted: 06/04/2020] [Indexed: 12/23/2022]
Abstract
Background Stratum corneum hydration (SCH) and transepidermal water loss (TEWL) provide useful information about skin barrier function. This study aimed to determine the value of GPSkin Pro, a new handheld device determining both SCH and TEWL, to measure skin barrier impairment and to monitor barrier function in rosacea in daily practice. Materials and Methods Two pilots were performed. Pilot 1: in 27 healthy participants, GPSkin SCH and TEWL were compared to Aquaflux® and Epsilon® values at the forearm before and after skin barrier perturbation via tapestripping. Moreover, GPSkin values were measured at both cheeks without intervention. Pilot 2: in 16 rosacea patients, GPSkin measurements were performed at the forearm, and at both cheeks before and during anti‐inflammatory treatment. They were compared to clinical symptoms and to GPSkin values from pilot 1. Results Pilot 1: after merging data from before and after tapestripping, a strong correlation was observed between GPSkin TEWL and Aquaflux® (Rs = 0.9256), and GPSkin SCH and Epsilon® (Rs = 0.8798). Pilot 2: SCH was significantly lower at the cheeks of rosacea patients compared to controls, with a normalizing trend during successful treatment. TEWL was comparable among patients and controls and did not change during treatment at all locations. Conclusion The GPSkin determines TEWL and SCH accurately in healthy and impaired skin barrier state and can monitor skin barrier function in rosacea during treatment. The GPSkin device is much more practical compared to previous skin barrier tools when used in clinical practice. Its further validation in other inflammatory skin diseases is recommended.
Collapse
Affiliation(s)
- Jade G M Logger
- Department of Dermatology, Radboud University Medical Center, Nijmegen, The Netherlands
| | - Rieke J B Driessen
- Department of Dermatology, Radboud University Medical Center, Nijmegen, The Netherlands
| | - Elke M G J de Jong
- Department of Dermatology, Radboud University Medical Center, Nijmegen, The Netherlands
| | - Piet E J van Erp
- Department of Dermatology, Radboud University Medical Center, Nijmegen, The Netherlands
| |
Collapse
|
6
|
Choe C, Ri J, Schleusener J, Lademann J, Darvin ME. The non-homogenous distribution and aggregation of carotenoids in the stratum corneum correlates with the organization of intercellular lipids in vivo. Exp Dermatol 2019; 28:1237-1243. [PMID: 31400168 DOI: 10.1111/exd.14018] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/05/2019] [Revised: 07/09/2019] [Accepted: 08/01/2019] [Indexed: 12/30/2022]
Abstract
The human stratum corneum (SC) contains an abundant amount of carotenoid antioxidants, quenching free radicals and thereby protecting the skin. For the precise measurements of the depth-dependent carotenoid concentration, confocal Raman microscopy is a suitable method. The quantitative concentration can be determined by the carotenoid-related peak intensity of a Gaussian function approached at ≈1524 cm-1 using non-linear regression. Results show that the carotenoid concentration is higher at the superficial layers of the SC then decreases to a minimum at 20% SC depth and increases again towards the bottom of the SC. In the present work, two carotenoid penetration pathways into the SC are postulated. The first pathway is from the stratum granulosum to the bottom of the SC, while in the second pathway, the carotenoids are delivered to the skin surface by sweat and/or sebum secretion and penetrate from outside. The carotenoids are aggregated at the superficial layers, which are shown by high correlation between the aggregation states of carotenoids and the lateral organization of lipids. At the 30%-40% SC depths, the ordered and dense lipid molecules intensify the lipid-carotenoid interactions and weaken the carotenoid-carotenoid interaction and thus exhibit the disaggregation of carotenoids. At 90%-100% SC depths, the carotenoid-lipid interaction is weakened and the carotenoids have a tendency to be aggregated. Thus, the molecular structural correlation of carotenoid and SC lipid might be reserved in the intercellular space of the SC and also serves as the skeleton of the intercellular lipids.
Collapse
Affiliation(s)
- ChunSik Choe
- Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Berlin Institute of Health, Charité - Universitätsmedizin Berlin, Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin, Germany.,Kim Il Sung University, Pyongyang, Korea
| | | | - Johannes Schleusener
- Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Berlin Institute of Health, Charité - Universitätsmedizin Berlin, Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin, Germany
| | - Juergen Lademann
- Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Berlin Institute of Health, Charité - Universitätsmedizin Berlin, Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin, Germany
| | - Maxim E Darvin
- Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Berlin Institute of Health, Charité - Universitätsmedizin Berlin, Freie Universität Berlin, Humboldt-Universität zu Berlin, Berlin, Germany
| |
Collapse
|
7
|
Affiliation(s)
- Leopold Eckhart
- Research Division of Biology and Pathobiology of the Skin, Department of Dermatology, Medical University of Vienna, Vienna, Austria
| | - Patrick L J M Zeeuwen
- Department of Dermatology, Radboud Institute for Molecular Life Sciences (RIMLS), Radboud University Nijmegen Medical Center (Radboudumc), Nijmegen, The Netherlands
| |
Collapse
|
8
|
Darvin ME, Choe CS, Schleusener J, Lademann J. Non-invasive depth profiling of the stratum corneum in vivo using confocal Raman microscopy considering the non-homogeneous distribution of keratin. BIOMEDICAL OPTICS EXPRESS 2019; 10:3092-3103. [PMID: 31259076 PMCID: PMC6583359 DOI: 10.1364/boe.10.003092] [Citation(s) in RCA: 15] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/26/2019] [Revised: 05/02/2019] [Accepted: 05/02/2019] [Indexed: 06/09/2023]
Abstract
Confocal Raman microscopy has a number of advantages in investigating the human stratum corneum (SC) in vivo and ex vivo. The penetration profiles of xenobiotics in the SC, as well as depth profiles of the physiological parameters of the SC, such as the concentration of water depending on the strength of hydrogen bonds, total water concentration, the hydrogen bonding state of water molecules, concentration of intercellular lipids, the lamellar and lateral packing order of intercellular lipids, the concentration of natural moisturizing factor molecules, carotenoids, and the secondary and tertiary structure properties of keratin are well investigated. To consider the depth-dependent Raman signal attenuation, in most cases a normalization procedure is needed, which uses the main SC's protein keratin-related Raman peaks, based on the assumption that keratin is homogeneously distributed in the SC. We found that this assumption is not accurate for the bottom part of the SC, where the water concentration is considerably increased, thus, reducing the presence of keratin. Our results demonstrate that the bottom part of the SC depth profile should be multiplied by 0.94 in average in order to match this non-homogeneity, which result in a decrease of the uncorrected values in these depths. The correctly normalized depth profiles of the concentration of lipids, water, natural moisturizing factor and carotenoids are presented in this work. The obtained results should be taken into consideration in future skin research using confocal Raman microscopy.
Collapse
Affiliation(s)
- Maxim E. Darvin
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Charitéplatz 1, 10117 Berlin, Germany
| | - Chun-Sik Choe
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Charitéplatz 1, 10117 Berlin, Germany
- Kim Il Sung University, Ryongnam-Dong, Taesong District, Pyongyang, DPR Korea
| | - Johannes Schleusener
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Charitéplatz 1, 10117 Berlin, Germany
| | - Jürgen Lademann
- Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Department of Dermatology, Venerology and Allergology, Center of Experimental and Applied Cutaneous Physiology, Charitéplatz 1, 10117 Berlin, Germany
| |
Collapse
|
9
|
Abstract
The epidermal barrier function is disrupted in various inflammatory skin diseases. Accurate methods to measure skin barrier function are needed to assess the effect of therapeutic agents. Therefore, we developed a noninvasive multiparametric approach to measure four different parameters regarding the skin barrier. In the current pilot study, we evaluate this method in 14 healthy volunteers. We assessed erythema, transepidermal water loss (TEWL), water content, and epidermal thickness at both cheeks before and 30 min after application of Lanette and Vaseline-Lanette cream. For this, we used spectrophotometry, the Aquaflux device, the Epsilon device, and reflection confocal microscopy, respectively. Stratum corneum (SC) thickness was significantly increased after application of both creams (p < 0.05), and this increase was larger after Lanette cream compared to after Vaseline-Lanette cream (p = 0.035). Erythema, TEWL, and water content did not significantly change after cream application. Our multiparametric approach is promising and offers a feasible and practical way to quickly obtain multifaceted information about skin barrier function. Further exploration of this approach after prolonged use of cream and in conditions of disrupted skin barrier are recommended areas for future research.