García-Moreno A, Comerma-Montells A, Tudela-Pi M, Minguillon J, Becerra-Fajardo L, Ivorra A. Wireless networks of injectable microelectronic stimulators based on rectification of volume conducted high frequency currents. J Neural Eng 2022; 19. [PMID: 36041421 DOI: 10.1088/1741-2552/ac8dc4] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/15/2022] [Accepted: 08/30/2022] [Indexed: 11/12/2022]
Abstract
OBJECTIVE To develop and in vivo demonstrate threadlike wireless implantable neuromuscular microstimulators that are digitally addressable. APPROACH These devices perform, through its two electrodes, electronic rectification of innocuous high frequency current bursts delivered by volume conduction via epidermal textile electrodes. By avoiding the need of large components to obtain electrical energy, this approach allows the development of thin devices that can be intramuscularly implanted by minimally invasive procedures such as injection. For compliance with electrical safety standards, this approach requires a minimum distance, in the order of millimeters or a very few centimeters, between the implant electrodes. Additionally, the devices must cause minimal mechanical damage to tissues, avoid dislocation and be adequate for long-term implantation. Considering these requirements, the implants were conceived as tubular and flexible devices with two electrodes at opposite ends and, at the middle section, a hermetic metallic capsule housing the electronics. MAIN RESULTS The developed implants have a submillimetric diameter (0.97 mm diameter, 35 mm length) and consist of a microcircuit, which contains a single custom-developed integrated circuit, housed within a titanium capsule (0.7 mm diameter, 6.5 mm length), and two platinum-iridium coils that form two electrodes (3 mm length) located at opposite ends of a silicone body. These neuromuscular stimulators are addressable, allowing to establish a network of microstimulators that can be controlled independently. Their operation was demonstrated in an acute study by injecting a few of them in the hind limb of anesthetized rabbits and inducing controlled and independent contractions. SIGNIFICANCE These results show the feasibility of manufacturing threadlike wireless addressable neuromuscular stimulators by using fabrication techniques and materials well established for chronic electronic implants. Although long-term operation still must be demonstrated, the obtained results pave the way to the clinical development of advanced motor neuroprostheses formed by dense networks of such wireless devices.
Collapse
Affiliation(s)
- Aracelys García-Moreno
- Department of Information and Communications Technologies, Pompeu Fabra University Department of Information and Communication Technologies, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| | - Albert Comerma-Montells
- Department of Information and Communications Technologies, Universitat Pompeu Fabra, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| | - Marc Tudela-Pi
- Department of Information and Communications Technologies, Universitat Pompeu Fabra, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| | - Jesus Minguillon
- Department of Information and Communications Technologies, Universitat Pompeu Fabra, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| | - Laura Becerra-Fajardo
- Department of Information and Communications Technologies, Universitat Pompeu Fabra, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| | - Antoni Ivorra
- Department of Information and Communications Technologies, Universitat Pompeu Fabra, Carrer Roc Boronat 138, Barcelona, Barcelona, 08018, SPAIN
| |
Collapse
|
2
|
Eickhoff S, Jarvis JC. Pulse Shaping Strategies for Electroceuticals: A Comprehensive Survey of the Use of Interphase Gaps in Miniature Stimulation Systems. IEEE Trans Biomed Eng 2021; 68:1658-1667. [PMID: 33651679 DOI: 10.1109/tbme.2021.3063029] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
OBJECTIVE Interphase gaps (IPGs) are among the most commonly suggested pulse shape variations to try to enhance neural stimulation efficiency by reducing the action potential (AP) suppressing effect of an early anodic hyperpolarization. The majority of published literature on the effect of IPGs is based on investigations of monopolar stimulation configurations. However, many contemporary neuromodulation applications including the emerging field of electroceutical devices operate in a bipolar electrode configuration. METHODS We investigated the effect of IPGs and asymmetric biphasic current controlled pulses with reduced anodic amplitude on neural activation in both principal electrode configurations in a rodent in-vivo nerve muscle preparation. RESULTS In the monopolar electrode configuration, our findings of 10.9 ± 1.5% decreased stimulation amplitude with 200 μs IPGs in biphasic pulses of 40 μs phase width are in agreement with published literature in this configuration. Surprisingly, using the bipolar configuration, opposite effects of IPGs were observed and neural activation required up to 18.6 ± 3.1% (phase width 100 μs, IPG = 1000 μs) higher amplitudes. Electroneurogram recordings of the stimulated nerve revealed temporal differences in AP generation between the monopolar and bipolar configuration. In the bipolar configuration excitation first occurred in response to the middle field transition of biphasic pulses. CONCLUSION This is the first study to report consistently increased amplitude requirements with IPGs in bipolar stimulation configurations. SIGNIFICANCE Our findings must be taken into consideration when designing stimulation waveforms for neuromodulation devices that operate in a bipolar mode to avoid increased amplitude requirements that result in increased energy consumption.
Collapse
|
3
|
Abiri P, Yousefi A, Abiri A, Gudapati V, Ding Y, Nguyen KL, Abiri A, Markovic D, Tai YC, Hsiai TK. A Multi-Dimensional Analysis of a Novel Approach for Wireless Stimulation. IEEE Trans Biomed Eng 2020; 67:3307-3316. [PMID: 32248088 DOI: 10.1109/tbme.2020.2983443] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The elimination of integrated batteries in biomedical implants holds great promise for improving health outcomes in patients with implantable devices. However, despite extensive research in wireless power transfer, achieving efficient power transfer and effective operational range have remained a hindering challenge within anatomical constraints. OBJECTIVE We hereby demonstrate an intravascular wireless and batteryless microscale stimulator, designed for (1) low power dissipation via intermittent transmission and (2) reduced fixation mechanical burden via deployment to the anterior cardiac vein (ACV, ∼3.8 mm in diameter). METHODS We introduced a unique coil design circumferentially confined to a 3 mm diameter hollow-cylinder that was driven by a novel transmitter-based control architecture with improved power efficiency. RESULTS We examined wireless capacity using heterogenous bovine tissue, demonstrating >5 V stimulation threshold with up to 20 mm transmitter-receiver displacement and 20° of misalignment. Feasibility for human use was validated using Finite Element Method (FEM) simulation of the cardiac cycle, guided by pacer phantom-integrated Magnetic Resonance Images (MRI). CONCLUSION This system design thus enabled sufficient wireless power transfer in the face of extensive stimulator miniaturization. SIGNIFICANCE Our successful feasibility studies demonstrated the capacity for minimally invasive deployment and low-risk fixation.
Collapse
|
4
|
Abiri P, Abiri A, Gudapati V, Chang CC, Roustaei M, Bourenane H, Anwar U, Markovic D, Hsiai TK. Wireless Pacing Using an Asynchronous Three-Tiered Inductive Power Transfer System. Ann Biomed Eng 2020; 48:1368-1381. [PMID: 31974869 DOI: 10.1007/s10439-020-02450-y] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/06/2019] [Accepted: 01/06/2020] [Indexed: 11/29/2022]
Abstract
Despite numerous advancements in pacemaker technology for the treatment of cardiac arrhythmias and conduction disorders, lead-related complications associated with these devices continue to compromise patient safety and survival. In this work, we present a system architecture that has the capacity to deliver power to a wireless, batteryless intravascular pacer. This was made possible through a three-tiered, dual-sub-system, four-coil design, which operates on two different frequencies through intermittent remote-controlled inductive power transfer. System efficiency was enhanced using coil design optimization, and validated using numerical simulations and experimental analysis. Our pacemaker design was concepted to achieve inductive power transfer over a 55 mm range to a microscale pacer with a 3 mm diameter. Thus, the proposed system design enabled long-range wireless power transfer to a small implanted pacer with the capacity for intravascular deployment to the anterior cardiac vein. This proposed stent-like fixation mechanism can bypass the multitude of complications associated with pacemaker wires while wireless power can eliminate the need for repeated procedures for battery replacement.
Collapse
Affiliation(s)
- Parinaz Abiri
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA.,Department of Medicine, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Arash Abiri
- Department of Medicine, University of California, Irvine, Irvine, CA, 92697, USA
| | - Varun Gudapati
- Department of Medicine, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Chih-Chiang Chang
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Mehrdad Roustaei
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Hamed Bourenane
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Usama Anwar
- Department of Electrical Engineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Dejan Markovic
- Department of Electrical Engineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA
| | - Tzung K Hsiai
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, CA, 90095, USA. .,Department of Medicine, University of California, Los Angeles, Los Angeles, CA, 90095, USA.
| |
Collapse
|
5
|
Levine YA, Faltys M, Chernoff D. Harnessing the Inflammatory Reflex for the Treatment of Inflammation-Mediated Diseases. Cold Spring Harb Perspect Med 2020; 10:cshperspect.a034330. [PMID: 30833463 DOI: 10.1101/cshperspect.a034330] [Citation(s) in RCA: 9] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/31/2022]
Abstract
Treating diseases nonpharmacologically, using targeted neurostimulation instead of systemic drugs, is a hallmark of the burgeoning field of bioelectronic medicine. In this review, we provide a brief overview of the discovery and function of the prototypical neuroimmune reflex, the "inflammatory reflex." We discuss various biomarkers developed and used to translate early physiological discoveries into dosing parameters used in experimental settings, from the treatment of animal models of disease through a proof-of-concept clinical study in rheumatoid arthritis (RA). Finally, we relate how unique aspects of this form of therapy enabled the design of a next-generation implanted pulse generator using integrated electrodes, currently under evaluation in a U.S.-based clinical study for patients with drug refractory RA.
Collapse
|
6
|
Ni L, KC P, Zhang G, Zhe J. Enabling single cell electrical stimulation and response recording via a microfluidic platform. BIOMICROFLUIDICS 2019; 13:064126. [PMID: 31867086 PMCID: PMC6910869 DOI: 10.1063/1.5128884] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/23/2019] [Accepted: 11/30/2019] [Indexed: 05/12/2023]
Abstract
Electrical stimulation (ES) has been recognized to play important roles in regulating cell behaviors. A microfluidic device was developed for the electrical stimulation of single cells and simultaneous recording of extracellular field potential (EFP). Each single cell was trapped onto an electrode surface by a constriction channel for ES testing and was then driven to the outlet by the pressure afterward. This design allows the application of ES on and detection of EFP of single cells continuously in a microfluidic channel. Human cardiomyocytes and primary rat cortex neurons were tested with specific ES with the device. Each cell's EFP signal was detected and analyzed during the ES process. Results have shown that after applying specific ES on the excitable single cells, the cells evoked electrical responses. In addition, increased secretion of glutamic acid was detected from the stimulated neurons. Altogether, these results indicated that the developed device can be used to continuously apply ES on and accurately determine cell responses of single cells with shorter probing time. The throughput of the measurement can achieve 1 cell per minute, which is higher than the traditional ES methods that need culturing cells or manually positioning the cells onto the electrode surface. Before and after the application of ES, the cell viability had no significant change. Such a device can be used to study the biological process of various types of cells under electrical stimulation.
Collapse
Affiliation(s)
- Liwei Ni
- Department of Mechanical Engineering, University of Akron, Akron, Ohio 44325, USA
| | - Pawan KC
- Department of Biomedical Engineering, University of Akron, Akron, Ohio 44325, USA
| | - Ge Zhang
- Department of Biomedical Engineering, University of Akron, Akron, Ohio 44325, USA
- Authors to whom correspondence should be addressed: and
| | - Jiang Zhe
- Department of Mechanical Engineering, University of Akron, Akron, Ohio 44325, USA
- Authors to whom correspondence should be addressed: and
| |
Collapse
|
7
|
Wong YT, Ahnood A, Maturana MI, Kentler W, Ganesan K, Grayden DB, Meffin H, Prawer S, Ibbotson MR, Burkitt AN. Feasibility of Nitrogen Doped Ultrananocrystalline Diamond Microelectrodes for Electrophysiological Recording From Neural Tissue. Front Bioeng Biotechnol 2018; 6:85. [PMID: 29988378 PMCID: PMC6024013 DOI: 10.3389/fbioe.2018.00085] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/29/2017] [Accepted: 06/05/2018] [Indexed: 01/19/2023] Open
Abstract
Neural prostheses that can monitor the physiological state of a subject are becoming clinically viable through improvements in the capacity to record from neural tissue. However, a significant limitation of current devices is that it is difficult to fabricate electrode arrays that have both high channel counts and the appropriate electrical properties required for neural recordings. In earlier work, we demonstrated nitrogen doped ultrananocrystalline diamond (N-UNCD) can provide efficacious electrical stimulation of neural tissue, with high charge injection capacity, surface stability and biocompatibility. In this work, we expand on this functionality to show that N-UNCD electrodes can also record from neural tissue owing to its low electrochemical impedance. We show that N-UNCD electrodes are highly flexible in their application, with successful recordings of action potentials from single neurons in an in vitro retina preparation, as well as local field potential responses from in vivo visual cortex tissue. Key properties of N-UNCD films, combined with scalability of electrode array fabrication with custom sizes for recording or stimulation along with integration through vertical interconnects to silicon based integrated circuits, may in future form the basis for the fabrication of versatile closed-loop neural prostheses that can both record and stimulate.
Collapse
Affiliation(s)
- Yan T. Wong
- Department of Physiology and Department of Electrical and Computer Systems Engineering, Biomedicine Discovery Institute, Monash University, Melbourne, VIC, Australia
| | - Arman Ahnood
- School of Physics, University of Melbourne, Melbourne, VIC, Australia
| | - Matias I. Maturana
- Department of Biomedical Engineering, University of Melbourne, Melbourne, VIC, Australia
- National Vision Research Institute, Australian College of Optometry, Carlton, VIC, Australia
| | - William Kentler
- Department of Biomedical Engineering, University of Melbourne, Melbourne, VIC, Australia
| | | | - David B. Grayden
- Department of Biomedical Engineering, University of Melbourne, Melbourne, VIC, Australia
| | - Hamish Meffin
- National Vision Research Institute, Australian College of Optometry, Carlton, VIC, Australia
- Department of Optometry and Vision Science University of Melbourne, Melbourne, VIC, Australia
| | - Steven Prawer
- School of Physics, University of Melbourne, Melbourne, VIC, Australia
| | - Michael R. Ibbotson
- National Vision Research Institute, Australian College of Optometry, Carlton, VIC, Australia
- Department of Optometry and Vision Science University of Melbourne, Melbourne, VIC, Australia
| | - Anthony N. Burkitt
- Department of Biomedical Engineering, University of Melbourne, Melbourne, VIC, Australia
| |
Collapse
|
8
|
Charthad J, Chang TC, Liu Z, Sawaby A, Weber MJ, Baker S, Gore F, Felt SA, Arbabian A. A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS 2018; 12:257-270. [PMID: 29578414 DOI: 10.1109/tbcas.2018.2799623] [Citation(s) in RCA: 84] [Impact Index Per Article: 14.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/05/2023]
Abstract
A wireless electrical stimulation implant for peripheral nerves, achieving >10× improvement over state of the art in the depth/volume figure of merit, is presented. The fully integrated implant measures just 2 mm × 3 mm × 6.5 mm (39 mm3, 78 mg), and operates at a large depth of 10.5 cm in a tissue phantom. The implant is powered using ultrasound and includes a miniaturized piezoelectric receiver (piezo), an IC designed in 180 nm HV BCD process, an off-chip energy storage capacitor, and platinum stimulation electrodes. The package also includes an optional blue light-emitting diode for potential applications in optogenetic stimulation in the future. A system-level design strategy for complete operation of the implant during the charging transient of the storage capacitor, as well as a unique downlink command/data transfer protocol, is presented. The implant enables externally programmable current-controlled stimulation of peripheral nerves, with a wide range of stimulation parameters, both for electrical (22 to 5000 μA amplitude, ∼14 to 470 μs pulse-width, 0 to 60 Hz repetition rate) and optical (up to 23 mW/mm2 optical intensity) stimulation. Additionally, the implant achieves 15 V compliance voltage for chronic applications. Full integration of the implant components, end-to-end in vitro system characterizations, and results for the electrical stimulation of a sciatic nerve, demonstrate the feasibility and efficacy of the proposed stimulator for peripheral nerves.
Collapse
|
9
|
|
10
|
Agarwal K, Jegadeesan R, Guo YX, Thakor NV. Wireless Power Transfer Strategies for Implantable Bioelectronics. IEEE Rev Biomed Eng 2017; 10:136-161. [PMID: 28328511 DOI: 10.1109/rbme.2017.2683520] [Citation(s) in RCA: 78] [Impact Index Per Article: 11.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
Neural implants have emerged over the last decade as highly effective solutions for the treatment of dysfunctions and disorders of the nervous system. These implants establish a direct, often bidirectional, interface to the nervous system, both sensing neural signals and providing therapeutic treatments. As a result of the technological progress and successful clinical demonstrations, completely implantable solutions have become a reality and are now commercially available for the treatment of various functional disorders. Central to this development is the wireless power transfer (WPT) that has enabled implantable medical devices (IMDs) to function for extended durations in mobile subjects. In this review, we present the theory, link design, and challenges, along with their probable solutions for the traditional near-field resonant inductively coupled WPT, capacitively coupled short-ranged WPT, and more recently developed ultrasonic, mid-field, and far-field coupled WPT technologies for implantable applications. A comparison of various power transfer methods based on their power budgets and WPT range follows. Power requirements of specific implants like cochlear, retinal, cortical, and peripheral are also considered and currently available IMD solutions are discussed. Patient's safety concerns with respect to electrical, biological, physical, electromagnetic interference, and cyber security from an implanted neurotech device are also explored in this review. Finally, we discuss and anticipate future developments that will enhance the capabilities of current-day wirelessly powered implants and make them more efficient and integrable with other electronic components in IMDs.
Collapse
|
11
|
Ivorra A, Becerra-Fajardo L, Castellví Q. In vivo demonstration of injectable microstimulators based on charge-balanced rectification of epidermically applied currents. J Neural Eng 2015; 12:066010. [PMID: 26447945 DOI: 10.1088/1741-2560/12/6/066010] [Citation(s) in RCA: 8] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
OBJECTIVE It is possible to develop implantable microstimulators whose actuation principle is based on rectification of high-frequency (HF) current bursts supplied through skin electrodes. This has been demonstrated previously by means of devices consisting of a single diode. However, previous single diode devices caused dc currents which made them impractical for clinical applications. Here flexible thread-like stimulation implants which perform charge balance are demonstrated in vivo. APPROACH The implants weigh 40.5 mg and they consist of a 3 cm long tubular silicone body with a diameter of 1 mm, two electrodes at opposite ends, and, within the central section of the body, an electronic circuit made up of a diode, two capacitors, and a resistor. In the present study, each implant was percutaneously introduced through a 14 G catheter into either the gastrocnemius muscle or the cranial tibial muscle of a rabbit hindlimb. Then stimulation was performed by delivering HF bursts (amplitude <60 V, frequency 1 MHz, burst repetition frequency from 10 Hz to 200 Hz, duration = 200 μs) through a pair of textile electrodes strapped around the hindlimb and either isometric plantarflexion or dorsiflexion forces were recorded. Stimulation was also assayed 1, 2 and 4 weeks after implantation. MAIN RESULTS The implants produced bursts of rectified current whose mean value was of a few mA and were capable of causing local neuromuscular stimulation. The implants were well-tolerated during the 4 weeks. SIGNIFICANCE Existing power supply methods, and, in particular inductive links, comprise stiff and bulky parts. This hinders the development of minimally invasive implantable devices for neuroprostheses based on electrical stimulation. The proposed methodology is intended to relieving such bottleneck. In terms of mass, thinness, and flexibility, the demonstrated implants appear to be unprecedented among the intramuscular stimulation implants ever assayed in vertebrates.
Collapse
|
12
|
Li X, Serdijn WA, Zheng W, Tian Y, Zhang B. The injectable neurostimulator: an emerging therapeutic device. Trends Biotechnol 2015; 33:388-94. [PMID: 25921343 DOI: 10.1016/j.tibtech.2015.04.001] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/06/2015] [Revised: 03/31/2015] [Accepted: 04/01/2015] [Indexed: 12/14/2022]
Abstract
Injectable neurostimulators are currently applied in clinical trials to minimize the side effects such as discomfort, risk of infection, and post-surgery trauma, which can be pronounced with conventional, bulky implantable neurostimulators. Owing to its smaller size, wireless-updatable software, and wireless power supply, the injectable neurostimulator is presumably less invasive, 'smarter', and has a longer lifetime. We discuss the concept and development of the injectable neurostimulator, persistent implementation challenges, and obstacles to be overcome in its evolution. We survey the use of new materials, technologies, and design methods for injectable electrodes, batteries, antennas, and packaging to enhance reliability and other features. These advances in the field are accompanied by progress in electrophysiology, neuroscience, neurology, clinical trials, and treatments.
Collapse
Affiliation(s)
- Xiaolong Li
- School of Electronics and Information, Jiangsu University of Science and Technology, Zhenjiang, China.
| | - Wouter A Serdijn
- Section of Bioelectronics, Delft University of Technology, Delft, the Netherlands
| | - Wei Zheng
- School of Electronics and Information, Jiangsu University of Science and Technology, Zhenjiang, China
| | - Yubo Tian
- School of Electronics and Information, Jiangsu University of Science and Technology, Zhenjiang, China
| | - Bing Zhang
- School of Electronics and Information, Jiangsu University of Science and Technology, Zhenjiang, China
| |
Collapse
|
13
|
Chan YC, Ting S, Lee YK, Ng KM, Zhang J, Chen Z, Siu CW, Oh SKW, Tse HF. Electrical stimulation promotes maturation of cardiomyocytes derived from human embryonic stem cells. J Cardiovasc Transl Res 2013; 6:989-99. [PMID: 24081385 DOI: 10.1007/s12265-013-9510-z] [Citation(s) in RCA: 126] [Impact Index Per Article: 11.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/29/2013] [Accepted: 09/02/2013] [Indexed: 11/26/2022]
Abstract
While human embryonic stem cells (hESCs) can differentiate into functional cardiomyocytes, their immature phenotypes limit their therapeutic application for myocardial regeneration. We sought to determine whether electrical stimulation could enhance the differentiation and maturation of hESC-derived cardiomyocytes. Cardiac differentiation was induced in a HES3 hESC line via embryoid bodies formation treated with a p38 MAP kinase inhibitor. Detailed molecular and functional analysis were performed in those hESC-derived cardiomyocytes cultured for 4 days in the absence or presence of electrical field stimulation (6.6 V/cm, 1 Hz, and 2 ms pulses) using an eight-channel C-Pace stimulator (Ion-Optics Co., MA). Upon electrical stimulation, quantitative polymerase chain reaction demonstrated significant upregulation of cardiac-specific gene expression including HCN1, MLC2V, SCN5A, SERCA, Kv4.3, and GATA4; immunostaining and flow cytometry analysis revealed cellular elongation and an increased proportion of troponin-T positive cells (6.3 ± 1.2% vs. 15.8 ± 2.1%; n = 3, P < 0.01). Electrophysiological studies showed an increase in the proportion of ventricular-like hESC-derived cardiomyocytes (48 vs. 29%, P < 0.05) with lengthening of their action potential duration at 90% repolarization (387.7 ± 35.35; n = 11 vs. 291.8 ± 20.82; n = 10, P < 0.05) and 50% repolarization (313.9 ± 27.94; n = 11 vs. 234.0 ± 16.10; n = 10, P < 0.05) after electrical stimulation. Nonetheless, the membrane diastolic potentials and action potential upstrokes of different hESC-derived cardiomyocyte phenotypes, and the overall beating rate remained unchanged (all P > 0.05). Fluorescence confocal imaging revealed that electrical stimulation significantly increased both spontaneous and caffeine-induced calcium flux in the hESC-derived cardiomyocytes (approximately 1.6-fold for both cases; P < 0.01). In conclusion, electrical field stimulation increased the expression of cardiac-specific genes and the yield of differentiation, promoted ventricular-like phenotypes, and improved the calcium handling of hESC-derived cardiomyocytes.
Collapse
Affiliation(s)
- Yau-Chi Chan
- Cardiology Division, Department of Medicine, Queen Mary Hospital, The University of Hong Kong, Hong Kong, Hong Kong SAR, China
| | | | | | | | | | | | | | | | | |
Collapse
|
14
|
Merrill DR, Davis R, Turk R, Burridge JH. A personalized sensor-controlled microstimulator system for arm rehabilitation poststroke. Part 1: System architecture. Neuromodulation 2011; 14:72-9; discussion 79. [PMID: 21992166 DOI: 10.1111/j.1525-1403.2010.00309.x] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
Abstract
OBJECTIVES For rehabilitation of the poststroke upper limb in seven subjects, an external sensor-based system controls the timing of five to seven microstimulators implanted near radial nerve branches or their motor points to sequentially extend the elbow, wrist, and fingers with thumb extension and abduction, enabled at the subject's own pace. We hypothesize this system will support sequential activation of affected upper limb muscles intended to improve functional recovery. MATERIALS AND METHODS Presented here is a personalized sensor-controlled stimulation system, including its architecture, sensor design, and testing of equipment specific to this study, including coils and sensors. RESULTS All electrical and magnetic tests, and safety tests per International Electrotechnical Commission 60601-1 passed. One sensor type displayed a vulnerability to drop. CONCLUSIONS The new control system tested safe, met requirements, and allowed each subject to activate the system at their own pace, making the rehabilitation process more acceptable and efficient.
Collapse
|
15
|
Remote electrical stimulation by means of implanted rectifiers. PLoS One 2011; 6:e23456. [PMID: 21850274 PMCID: PMC3151300 DOI: 10.1371/journal.pone.0023456] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/11/2011] [Accepted: 07/18/2011] [Indexed: 11/19/2022] Open
Abstract
Miniaturization of active implantable medical devices is currently compromised by the available means for electrically powering them. Most common energy supply techniques for implants – batteries and inductive couplers – comprise bulky parts which, in most cases, are significantly larger than the circuitry they feed. Here, for overcoming such miniaturization bottleneck in the case of implants for electrical stimulation, it is proposed to make those implants act as rectifiers of high frequency bursts supplied by remote electrodes. In this way, low frequency currents will be generated locally around the implant and these low frequency currents will perform stimulation of excitable tissues whereas the high frequency currents will cause only innocuous heating. The present study numerically demonstrates that low frequency currents capable of stimulation can be produced by a miniature device behaving as a diode when high frequency currents, neither capable of thermal damage nor of stimulation, flow through the tissue where the device is implanted. Moreover, experimental evidence is provided by an in vivo proof of concept model consisting of an anesthetized earthworm in which a commercial diode was implanted. With currently available microelectronic techniques, very thin stimulation capsules (diameter <500 µm) deliverable by injection are easily conceivable.
Collapse
|
16
|
BION microstimulators: A case study in the engineering of an electronic implantable medical device. Med Eng Phys 2011; 33:7-16. [PMID: 21087890 DOI: 10.1016/j.medengphy.2010.08.010] [Citation(s) in RCA: 29] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/13/2010] [Revised: 08/14/2010] [Accepted: 08/18/2010] [Indexed: 11/20/2022]
|
17
|
Hart RL, Bhadra N, Montague FW, Kilgore KL, Peckham PH. Design and testing of an advanced implantable neuroprosthesis with myoelectric control. IEEE Trans Neural Syst Rehabil Eng 2010; 19:45-53. [PMID: 20876029 DOI: 10.1109/tnsre.2010.2079952] [Citation(s) in RCA: 29] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/09/2022]
Abstract
An implantable stimulator-telemeter (IST-12) was developed for applications in neuroprosthetic restoration of limb function in paralyzed individuals. The IST-12 provides 12 stimulation channels and two myoelectric signal (MES) channels. The MES circuitry includes a two-channel multiplexer, preamplifier, variable gain amplifier/bandpass filter, full-wave rectifier, and bin integrator. Power and control signals are transmitted from an external control unit to the IST-12 through an inductive link. Recorded MES signals are telemetered back to the external control unit through the same inductive link. Following bench testing, one device was implanted chronically in a dog for 15 months and evaluated. Conditions were identified in which MES could be recorded with minimal stimulus artifact. The ability to record MES in the presence of stimulation was verified, confirming the potential of the IST-12 to be used as a myoelectric controlled neuroprosthesis.
Collapse
Affiliation(s)
- Ronald L Hart
- Louis Stokes Veterans Affairs Medical Center, Cleveland, OH 44109, USA.
| | | | | | | | | |
Collapse
|
18
|
Kim JA, Roy RR, Kim SJ, Zhong H, Haddad F, Baldwin KM, Edgerton VR. Electromechanical modulation of catabolic and anabolic pathways in chronically inactive, but neurally intact, muscles. Muscle Nerve 2010; 42:410-21. [PMID: 20658566 DOI: 10.1002/mus.21720] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022]
Abstract
The extent and mechanisms by which neural input regulates skeletal muscle mass remain largely unknown. Adult spinal cord isolated (SI) rats were implanted unilaterally with a microstimulator, whereas the contralateral limb served as SI control (SI-C). A 100-HZ, 1-s stimulus was delivered every 30 s for 5 min, followed by a 5-min rest. This was repeated six times consecutively (SI-Stim1) or with a 9-h interval after the third bout (SI-Stim2) for 30 days (1 min of daily activity). SI-Stim1 and SI-Stim2 paradigms attenuated plantaris atrophy by 20% and 38%, respectively, whereas only SI-Stim2 blunted soleus atrophy (24%) relative to SI-C. Muscle mass changes occurred independent of the IGF-1/PI3K/Akt pathway. No relationships between SI or electromechanical stimulation and expression of several atrophy markers were observed. These data suggest that regulatory mechanisms for maintaining muscle mass previously shown in acute states of atrophy differ substantially from those observed in chronic states.
Collapse
Affiliation(s)
- Jung A Kim
- Department of Physiological Science, University of California, Los Angeles, Los Angeles, California 90095, USA
| | | | | | | | | | | | | |
Collapse
|
19
|
|
20
|
Chen MQ, Xie X, Wilson KD, Sun N, Wu JC, Giovangrandi L, Kovacs GTA. Current-Controlled Electrical Point-Source Stimulation of Embryonic Stem Cells. Cell Mol Bioeng 2009; 2:625-635. [PMID: 20652088 DOI: 10.1007/s12195-009-0096-0] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022] Open
Abstract
Stem cell therapy is emerging as a promising clinical approach for myocardial repair. However, the interactions between the graft and host, resulting in inconsistent levels of integration, remain largely unknown. In particular, the influence of electrical activity of the surrounding host tissue on graft differentiation and integration is poorly understood. In order to study this influence under controlled conditions, an in vitro system was developed. Electrical pacing of differentiating murine embryonic stem (ES) cells was performed at physiologically relevant levels through direct contact with microelectrodes, simulating the local activation resulting from contact with surrounding electroactive tissue. Cells stimulated with a charged balanced voltage-controlled current source for up to 4 days were analyzed for cardiac and ES cell gene expression using real-time PCR, immunofluorescent imaging, and genome microarray analysis. Results varied between ES cells from three progressive differentiation stages and stimulation amplitudes (nine conditions), indicating a high sensitivity to electrical pacing. Conditions that maximally encouraged cardiomyocyte differentiation were found with Day 7 EBs stimulated at 30 microA. The resulting gene expression included a sixfold increase in troponin-T and a twofold increase in beta-MHCwithout increasing ES cell proliferation marker Nanog. Subsequent genome microarray analysis revealed broad transcriptome changes after pacing. Concurrent to upregulation of mature gene programs including cardiovascular, neurological, and musculoskeletal systems is the apparent downregulation of important self-renewal and pluripotency genes. Overall, a robust system capable of long-term stimulation of ES cells is demonstrated, and specific conditions are outlined that most encourage cardiomyocyte differentiation.
Collapse
Affiliation(s)
- Michael Q Chen
- Department of Bioengineering, Stanford University, 330 Serra Mall, CISX-206X, Stanford, CA 94305, USA
| | | | | | | | | | | | | |
Collapse
|
21
|
The Bion® Microstimulator and its Clinical Applications. BIOLOGICAL AND MEDICAL PHYSICS, BIOMEDICAL ENGINEERING 2009. [DOI: 10.1007/978-0-387-77261-5_8] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 02/04/2023]
|
22
|
MRI Safety and Neuromodulation Systems. Neuromodulation 2009. [DOI: 10.1016/b978-0-12-374248-3.00021-5] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register]
|
23
|
Kim SJ, Roy RR, Kim JA, Zhong H, Haddad F, Baldwin KM, Edgerton VR. Gene expression during inactivity-induced muscle atrophy: effects of brief bouts of a forceful contraction countermeasure. J Appl Physiol (1985) 2008; 105:1246-54. [PMID: 18653749 DOI: 10.1152/japplphysiol.90668.2008] [Citation(s) in RCA: 36] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/22/2022] Open
Abstract
Anabolic and catabolic markers of muscle protein metabolism were examined in inactivity-induced atrophying muscles with and without daily short-duration, high-resistance isometric contractions. Inactivity was achieved via spinal cord isolation (SI), which results in near inactivity of the hindlimb musculature without compromising the motoneuron-muscle connectivity. Adult rats were assigned to a control (Con) or SI group in which one limb was stimulated (SI-Stim, 5 consecutive days of brief bouts of high-load isometric contractions) while the other served as a SI control (SI). Both the medial gastrocnemius (MG) and soleus weights (relative to body weight) were approximately 71% of Con in the SI, but maintained at Con in the SI-Stim group. Activity of the IGF-1/phosphatidylinositol 3-kinase (PI3K)/Akt pathway of protein synthesis was similar among all groups in the MG. Expression of atrogin-1 and muscle RING finger-1 (MuRF-1), markers of protein degradation, were higher in the MG and soleus of the SI than Con and maintained at Con in the SI-Stim. Compared with Con, the anti-growth factor myostatin was unaffected in the MG and soleus in the SI but was lower in the MG of the SI-Stim. These results demonstrate that upregulation of specific protein catabolic pathways plays a critical role in SI-induced atrophy, while this response was blunted by 4 min of daily high-resistance electromechanical stimulation and was able to preserve most of the muscle mass. Although the protein anabolic pathway (IGF-1/PI3K/Akt) appears to play a minor role in regulating mass in the SI model, increased translational capacity may have contributed to mass preservation in response to isometric contractions.
Collapse
Affiliation(s)
- Soo J Kim
- Brain Research Institute, University of California-Los Angeles, CA 90095-1761, USA
| | | | | | | | | | | | | |
Collapse
|
24
|
Schuettler M, Franke M, Krueger TB, Stieglitz T. A voltage-controlled current source with regulated electrode bias-voltage for safe neural stimulation. J Neurosci Methods 2008; 171:248-52. [PMID: 18471890 DOI: 10.1016/j.jneumeth.2008.03.016] [Citation(s) in RCA: 33] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/06/2008] [Revised: 03/24/2008] [Accepted: 03/26/2008] [Indexed: 11/29/2022]
Affiliation(s)
- Martin Schuettler
- University of Freiburg, Department of Microsystems Engineering-IMTEK, Laboratory for Biomedical Microtechnology, Georges-Koehler-Allee 102, 79110 Freiburg, Germany.
| | | | | | | |
Collapse
|
25
|
Loeb GE, Richmond FJR, Singh J, Peck RA, Tan W, Zou Q, Sachs N. RF-powered BIONs for stimulation and sensing. CONFERENCE PROCEEDINGS : ... ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL CONFERENCE 2007; 2004:4182-5. [PMID: 17271225 DOI: 10.1109/iembs.2004.1404167] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
Virtually all bodily functions are controlled by electrical signals in nerves and muscles. Electrical stimulation can restore missing signals but this has been difficult to achieve practically because of limitations in the bioelectric interfaces. Wireless, injectable microdevices are versatile, robust and relatively inexpensive to implant in a variety of sites and applications. Several variants are now in clinical use or under development to perform stimulation and/or sensing functions and to operate autonomously or with continuous coordination and feedback control.
Collapse
Affiliation(s)
- G E Loeb
- A.E. Mann Institute for Biomedical Engineering, University of Southern California, Los Angeles, CA 90089, USA
| | | | | | | | | | | | | |
Collapse
|
26
|
Carbunaru R, Whitehurst T, Jaax K, Koff J, Makous J. Rechargeable battery-powered bion microstimulators for neuromudulation. CONFERENCE PROCEEDINGS : ... ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL CONFERENCE 2007; 2004:4193-6. [PMID: 17271228 DOI: 10.1109/iembs.2004.1404170] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
The rechargeable battery-powered bion microstimulator is designed to bring neurostimulation to mainstream medicine and improve the quality of life of those suffering from neurological disorders. The bion, through direct electrical stimulation of excitable tissues, is intended to treat numerous disorders including urinary urge incontinence, fecal incontinence, chronic headaches, peripheral pain, angina and obstructive sleep apnea. The rechargeable bion microstimulator, manufactured by Advanced Bionics Corporation, is designed to be implanted using a minimally invasive approach that would appeal to clinicians of different specialties as well as patients suffering from a wide variety of debilitating conditions. The bion is currently in clinical trials for the treatment of urinary urge incontinence and headaches. Other clinical trials are expected to follow in the future.
Collapse
|
27
|
Kim SJ, Roy RR, Zhong H, Suzuki H, Ambartsumyan L, Haddad F, Baldwin KM, Edgerton VR. Electromechanical stimulation ameliorates inactivity-induced adaptations in the medial gastrocnemius of adult rats. J Appl Physiol (1985) 2007; 103:195-205. [PMID: 17431083 DOI: 10.1152/japplphysiol.01427.2006] [Citation(s) in RCA: 27] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/22/2022] Open
Abstract
The efficacy of high-load, short-duration isometric contractions, delivered as one vs. two sessions per day, on blunting inactivity-induced adaptations in the medial gastrocnemius (MG) were compared. Adult rats were assigned to a control (Con) or spinal cord-isolated (SI) group where one limb was stimulated (SI-Stim) while the other served as a SI control (SI-C). One bout of stimulation (BION microstimulator) consisted of a 100-Hz, 1-s stimulus, delivered every 30 s for 5 min with a 5-min rest period. This bout was repeated six times consecutively (SI-Stim1) or with a 9-h rest interval after the third bout (SI-Stim2) for 30 consecutive days. MG weights (relative to body weight) were 63, 72, and 79% of Con in SI-C, SI-Stim1, and SI-Stim2, respectively. Mean fiber size was 56% smaller in SI-C than in Con, and it was 19 and 31% larger in SI-Stim1 and SI-Stim2, respectively, compared with SI-C. Maximum tetanic tension was 42, 60, and 73% of Con in SI-C, SI-Stim1, and SI-Stim2, respectively. Specific tension was 77% of Con in SI-C, and at Con levels in both SI-Stim groups. SI increased the percent IIb myosin heavy chain composition (from 49 to 77%) and IIb+ fibers (from 63 to 79%): these adaptations were prevented by both Stim paradigms. These results demonstrate that 1) brief periods of high-load isometric contractions are effective in reducing inactivity-induced atrophy, functional deficits, and phenotypic adaptations in a fast hindlimb extensor, and 2) the same amount of stimulation distributed in two compared with one session per day is more effective in ameliorating inactivity-related adaptations.
Collapse
Affiliation(s)
- Soo J Kim
- Brain Research Institute, University of California, Los Angeles, Los Angeles, CA 90095-1761, USA
| | | | | | | | | | | | | | | |
Collapse
|
28
|
Weber DJ, Stein RB, Chan KM, Loeb GE, Richmond FJR, Rolf R, James K, Chong SL. BIONic WalkAide for correcting foot drop. CONFERENCE PROCEEDINGS : ... ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL CONFERENCE 2007; 2004:4189-92. [PMID: 17271227 DOI: 10.1109/iembs.2004.1404169] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/13/2023]
Abstract
The goal of this study was to test the feasibility and efficacy of using microstimulators (BIONs) to correct foot drop, the first human application of BIONs in functional electrical stimulation (FES). A prototype BIONic foot drop stimulator was developed by modifying a WalkAide2 stimulator to control BION stimulation of the ankle dorsiflexor muscles. BION stimulation was compared with surface stimulation of the common peroneal nerve provided by a normal WalkAide2 foot drop stimulator. Compared to surface stimulation, we found that BION stimulation of the deep peroneal nerve produces a more balanced ankle flexion movement without everting the foot. A 3-D motion analysis was performed to measure the ankle and foot kinematics with and without stimulation. Without stimulation, the toe on the affected leg drags across the ground. The BIONic WalkAide elevates the foot such that the toe clears the ground by 3 cm, which is equivalent to the toe clearance in the unaffected leg. The physiological cost index (PCI) was used to measure effort during walking. The PCI is high without stimulation (2.29 +/- 0.37; mean +/- S.D.) and greatly reduced with surface (1.29 +/- 0.10) and BION stimulation (1.46 +/- 0.24). Also, walking speed is increased from 9.4 +/- 0.4 m/min. without stimulation to 19.6 +/- 2.0 m/min. with surface and 17.8 +/- 0.7 m/min. with BION stimulation. We conclude that functional electrical stimulation with BIONs is a practical alternative to surface stimulation and provides more selective control of muscle activation.
Collapse
Affiliation(s)
- D J Weber
- Faculty of Physical Education and Recreation, University of Alberta, Edmonton, AB, Canada
| | | | | | | | | | | | | | | |
Collapse
|
29
|
Davoodi R, Urata C, Todorov E, Loeb GE. Development of clinician-friendly software for musculoskeletal modeling and control. CONFERENCE PROCEEDINGS : ... ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL CONFERENCE 2007; 2004:4622-5. [PMID: 17271337 DOI: 10.1109/iembs.2004.1404281] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/13/2023]
Abstract
Research and development in various fields dealing with human movement has been hampered by the lack of adequate software tools. We have formed a core development team to organize a collective effort by the research community to develop musculoskeletal modeling software that satisfies the requirements of both researchers and clinicians. We have identified initial requirements and have developed some of the basic components. We are developing common standards to facilitate sharing and reuse of musculoskeletal models and their component parts. Free distribution of the software and its source code will allow users to contribute to further development of the software as new models and data become available in the future.
Collapse
Affiliation(s)
- R Davoodi
- A.E. Mann Institute and Biomedical Engineering Department, University of Southern California, Los Angeles, CA 90089, USA
| | | | | | | |
Collapse
|
30
|
Loeb GE, Peck RA, Singh J, Kim YH, Deshpande S, Baker LL, Bryant JT. Mechanical loading of rigid intramuscular implants. Biomed Microdevices 2006; 9:901-10. [PMID: 17177107 DOI: 10.1007/s10544-006-9031-5] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/27/2022]
Abstract
Several groups are developing different versions of a new class of leadless, permanently implanted electronic devices with a size and form factor that allows them to be injected into muscles (BIONs). Their circuitry is protected from body fluids by thin-walled hermetic capsules made from rigid and brittle materials (glass or ceramic) that include feedthroughs to their electrodes. These packages experience repetitive stresses from the very contractions that they excite. We here provide a worst-case analysis of such stresses and methods for testing and validation of devices intended for such usage, along with the failure analysis and remediation strategy for a design that experienced unanticipated failures in vivo.
Collapse
Affiliation(s)
- Gerald E Loeb
- Alfred Mann Institute for Biomedical Engineering, University of Southern California, Los Angeles, CA, USA.
| | | | | | | | | | | | | |
Collapse
|
31
|
|
32
|
Abstract
Retinal prostheses represent the best near-term hope for individuals with incurable, blinding diseases of the outer retina. On the basis of the electrical activation of nerves, prototype retinal prostheses have been tested in blind humans and have demonstrated the capability to elicit the sensation of light and to give test subjects the ability to detect motion. To improve the visual function in implant recipients, a more sophisticated device is required. Simulations suggest that 600-1000 pixels will be required to provide visual function such as face recognition and reading. State-of-the-art implantable stimulator technology cannot produce such a device, which mandates the advancement of the state of the art in areas such as analog microelectronics, wireless power and data transfer, packaging, and stimulating electrodes.
Collapse
Affiliation(s)
- James D Weiland
- Doheny Retina Institute, Department of Ophthalmology, Keck School of Medicine, University of Southern California, Los Angeles, CA 90089, USA.
| | | | | |
Collapse
|
33
|
Weber DJ, Stein RB, Chan KM, Loeb G, Richmond F, Rolf R, James K, Chong SL. BIONic WalkAide for correcting foot drop. IEEE Trans Neural Syst Rehabil Eng 2005; 13:242-6. [PMID: 16003906 DOI: 10.1109/tnsre.2005.847385] [Citation(s) in RCA: 64] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
The goal of this study was to test the feasibility and efficacy of using microstimulators (BIONs) to correct foot drop, the first human application of BIONs in functional electrical stimulation (FES). A prototype BIONic foot drop stimulator was developed by modifying a WalkAide2 stimulator to control BION stimulation of the ankle dorsiflexor muscles. BION stimulation was compared with surface stimulation of the common peroneal nerve provided by a normal WalkAide2 foot drop stimulator. Compared to surface stimulation, we found that BION stimulation of the deep peroneal nerve produces a more balanced ankle flexion movement without everting the foot. A three-dimensional motion analysis was performed to measure the ankle and foot kinematics with and without stimulation. Without stimulation, the toe on the affected leg drags across the ground. The BIONic WalkAide elevates the foot such that the toe clears the ground by 3 cm, which is equivalent to the toe clearance in the unaffected leg. The physiological cost index (PCI) was used to measure effort during walking. The PCI is high without stimulation (2.29 +/- 0.37; mean +/- S.D.) and greatly reduced with surface (1.29 +/- 0.10) and BION stimulation (1.46 +/- 0.24). Also, walking speed is increased from 9.4 +/- 0.4 m/min without stimulation to 19.6 +/- 2.0 m/min with surface and 17.8 +/- 0.7 m/min with BION stimulation. We conclude that functional electrical stimulation with BIONs is a practical alternative to surface stimulation and provides more selective control of muscle activation.
Collapse
Affiliation(s)
- Douglas J Weber
- Centre for Neuroscience, University of Alberta, Edmonton, AB T6G 2H9, Canada.
| | | | | | | | | | | | | | | |
Collapse
|
34
|
Shellock FG, Cosendai G, Park SM, Nyenhuis JA. Implantable Microstimulator: Magnetic Resonance Safety at 1.5 Tesla. Invest Radiol 2004; 39:591-9. [PMID: 15377938 DOI: 10.1097/01.rli.0000138090.43450.ec] [Citation(s) in RCA: 32] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
RATIONALE AND OBJECTIVE Ex vivo testing is necessary to characterize implants to determine if it is safe for the patient to undergo a magnetic resonance imaging (MRI) examination. Therefore, the objective of this study was to evaluate MR safety for an implantable microstimulator in association with a 1.5 Tesla MR system. METHODS A microstimulator (RF BION, Alfred E. Mann Foundation for Scientific Research, Valencia, CA) was evaluated for magnetic field interactions and MRI-related heating. The functional aspects of this implant were assessed immediately before and after exposure to MRI (15 different pulse sequences). Artifacts were also characterized. RESULTS Magnetic field interactions exhibited by the microstimulator will not pose a hazard after a suitable postimplantation period has elapsed. Temperature changes will not pose a risk. The function of the microstimulator was unaffected by MRI. Artifacts will only create a problem if the area of interest is in proximity to this implant (largest artifact area: T1-weighted spin echo, 2291 mm2; gradient echo, 3310 mm2). CONCLUSION The overall findings indicated that it is safe for a patient with the microstimulator to undergo MRI at 1.5 Tesla by following specific safety guidelines described herein.
Collapse
|
35
|
Branner A, Stein RB, Fernandez E, Aoyagi Y, Normann RA. Long-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve. IEEE Trans Biomed Eng 2004; 51:146-57. [PMID: 14723504 DOI: 10.1109/tbme.2003.820321] [Citation(s) in RCA: 165] [Impact Index Per Article: 8.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
We studied the consequences of long-term implantation of a penetrating microelectrode array in peripheral nerve over the time course of 4-6 mo. Electrode arrays without lead wires were implanted to test the ability of different containment systems to protect the array and nerve during contractions of surrounding muscles. Treadmill walking was monitored and the animals showed no functional deficits as a result of implantation. In a different set of experiments, electrodes with lead wires were implanted for up to 7 mo and the animals were tested at 2-4 week intervals at which time stimulation thresholds and recorded sensory activity were monitored for every electrode. It was shown that surgical technique highly affected the long-term stimulation results. Results between measurement sessions were compared, and in the best case, the stimulation properties stabilized in 80% of the electrodes over the course of the experiment (162 days). The recorded sensory signals, however, were not stable over time. A histological analysis performed on all implanted tissues indicated that the morphology and fiber density of the nerve around the electrodes were normal.
Collapse
Affiliation(s)
- Almut Branner
- Center for Neural Interfaces, Department of Bioengineering, University of Utah, Salt Lake City, UT 84112, USA.
| | | | | | | | | |
Collapse
|
36
|
Abstract
The stability behavior of a multi-joint limb with electrically activated muscles provides important clues for postural control of motor tasks. The stability property of the musculoskeletal system can be characterized with its eigenvalues evaluated at operating postures in the workspace. A planar arm model with shoulder and elbow joints and three pairs of antagonistic muscles was constructed in ADAMS. Stability behavior of shoulder and elbow joints was analyzed using the loci of eigenvalues in the s-plane. In the analysis of open-loop cocontraction of antagonist muscles with increasing activation from 5% to 100%, the eigenvalues of the shoulder and elbow joints were confined within the left half of the s-plane in a stripe of +/- j 0.5, and moved toward left onto the real axis. The shoulder eigenvalues were generally nearer to the imaginary axis than the elbow ones, indicating a more oscillatory behavior at the shoulder joint than that at the elbow joint. The effects of joint configuration evaluated within the workspace from 40 degrees to 110 degrees for the elbow and from 40 degrees to 120 degrees for the shoulder showed that the elbow eigenvalues were more prone to configuration changes, particularly elbow angles. We also developed a simulation paradigm for sampled data FES control systems that contain a mixture of continuous time components and sampling and hold effects. This simulation paradigm is useful for realistic simulation of local feedback controller performance.
Collapse
Affiliation(s)
- Ning Lan
- Alfred E. Mann Institute for Biomedical Engineering, Department of Biomedical Engineering, University of Southern California, Los Angeles, CA 90089, USA.
| |
Collapse
|
37
|
Davoodi R, Andrews BJ, Wheeler GD, Lederer R. Development of an indoor rowing machine with manual FES controller for total body exercise in paraplegia. IEEE Trans Neural Syst Rehabil Eng 2002; 10:197-203. [PMID: 12503785 DOI: 10.1109/tnsre.2002.802880] [Citation(s) in RCA: 52] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/09/2022]
Abstract
Concept 2 indoor rowing machine (Concept 2 Inc., USA) was modified for functional electrical stimulation (FES) rowing exercise in paraplegia. A new seating system provides trunk stability and constrains the leg motion to the sagittal plane. A 4-channel electrical stimulator activates the quadriceps and hamstrings in Drive and Recovery phases of the rowing cycle, respectively. Two force-sensing resistors (FSR) on the handle measure the thumb press as the command signal to the electrical stimulator. Optical encoders measure the positions of the seat and handle during rowing. To synchronize the voluntarily controlled upper body movement with the FES controlled leg movement, a novel manual control system was developed. It uses the voluntary thumb presses to control the timing of the stimulation to the paralyzed leg muscles. The manual control system was intuitive and easy to learn and resulted in well-coordinated rowing. Evaluation of the modified rower by paraplegic volunteers showed that it is effective, safe, and affordable exercise alternative for paraplegics.
Collapse
Affiliation(s)
- Rahman Davoodi
- A. E. Mann Institute for Biomedical Engineering, University of Southern California, Los Angeles, CA 90089, USA.
| | | | | | | |
Collapse
|
38
|
Arcos I, Davis R, Fey K, Mishler D, Sanderson D, Tanacs C, Vogel MJ, Wolf R, Zilberman Y, Schulman J. Second-generation microstimulator. Artif Organs 2002; 26:228-31. [PMID: 11940019 DOI: 10.1046/j.1525-1594.2002.06938.x] [Citation(s) in RCA: 28] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
The first-generation injectable microstimulator was glass encased with an external tantalum capacitor electrode. This second-generation device uses a hermetically sealed ceramic case with platinum electrodes. Zener diodes protect the electronics from defibrillation shocks and from electrostatic discharge. The capacitor is sealed inside the case so that it cannot be inadvertently damaged by surgical instruments. This microstimulator, referred to as BION, is the main component of a 255-channel wireless stimulating system. BION devices have been implanted in rats for periods of up to 5 months. Results show benign tissue reactions resulting in identical encapsulation around BION and controls. Stimulation threshold levels did not change significantly over time and ranged between 0.81 to 1.35 mA for all the animals at a 60 micros pulse width. All of the tests performed to date indicate that the BION is safe and effective for long-term human implant. We have elected to develop BION applications by seeking collaboration with the research community through our BION Technology Partnership.
Collapse
Affiliation(s)
- Isabel Arcos
- Alfred E. Mann Foundation, Valencia, California 91355, USA.
| | | | | | | | | | | | | | | | | | | |
Collapse
|
39
|
Abstract
The development of man-made systems to restore functional vision in the profoundly blind has recently undergone a renaissance that has been fueled by a combination of celebrity and government interest, advances in the field of bioengineering, and successes with existing neuroprosthetic systems. This chapter presents the underlying physiologic principles of artificial vision, discusses three contemporary approaches to restoring functional vision in the blind, and concludes by presenting several relevant questions to vision prostheses. While there has been significant progress in the individual components constituting an artificial vision system, the remaining challenge of integrating these components with each other and the nervous system does not lie strictly in the realm of neuroscience, medicine, or engineering but at the interface of all three. In spite of the apparent complexity of an artificial vision system, it is not unreasonable to be optimistic about its eventual success.
Collapse
Affiliation(s)
- E M Maynard
- Center for Neural Interfaces, Department of Bioengineering, University of Utah, Salt Lake City, Utah 84112, USA.
| |
Collapse
|
40
|
Abstract
Historically, electronic devices such as pacemakers and neuromuscular stimulators have been surgically implanted into animals and humans. A new class of implants made possible by advances in monolithic electronic design and implant packaging is small enough to be implanted by percutaneous injection through large-gauge hypodermic needles and does not require surgical implantation. Among these, commercially available implants, known as radio frequency identification (RFID) tags, are used for livestock, pet, laboratory animal, and endangered-species identification. The RFID tag is a subminiature glass capsule containing a solenoidal coil and an integrated circuit. Acting as the implanted half of a transcutaneous magnetic link, the RFID tag is powered by and communicates with an extracorporeal magnetic reader. The tag transmits a unique identification code that serves the function of identifying the animal. Millions of RFID tags have been sold since the early 1980s. Based on the success of the RFID tags, research laboratories have developed injectable medical implants, known as micromodules. One type of micromodule, the microstimulator, is designed for use in functional-neuromuscular stimulation. Each microstimulator is uniquely addressable and could comprise one channel of a multichannel functional-neuromuscular stimulation system. Using bidirectional telemetry and commands, from a single extracorporeal transmitter, as many as 256 microstimulators could form the hardware basis for a complex functional-neuromuscular stimulation feedback-control system. Uses include stimulation of paralyzed muscle, therapeutic functional-neuromuscular stimulation, and neuromodulatory functions such as laryngeal stimulation and sleep apnea.
Collapse
Affiliation(s)
- P R Troyk
- Department of Electrical and Computer Engineering, Pritzker Institute of Medical Engineering, Illinois Institute of Technology, Chicago, Illinois 60616, USA.
| |
Collapse
|
41
|
Cheney PD, Hill-Karrer J, Belhaj-Saïf A, McKiernan BJ, Park MC, Marcario JK. Cortical motor areas and their properties: implications for neuroprosthetics. PROGRESS IN BRAIN RESEARCH 2001; 128:135-60. [PMID: 11105675 DOI: 10.1016/s0079-6123(00)28013-8] [Citation(s) in RCA: 13] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 02/18/2023]
Affiliation(s)
- P D Cheney
- Department of Molecular and Integrative Physiology, University of Kansas Medical Center, Kansas City 66160-7336, USA.
| | | | | | | | | | | |
Collapse
|
42
|
Abstract
We have developed the first in a planned series of neural prosthetic interfaces that allow multichannel systems to be assembled from single-channel micromodules called BIONs (BIOnic Neurons). Multiple BION implants can be injected directly into the sites requiring stimulating or sensing channels, where they receive power and digital commands by inductive coupling to an externally generated radio-frequency magnetic field. This article describes some of the novel technology required to achieve the required microminiaturization, hermeticity, power efficiency and clinical performance. The BION1 implants are now being used to electrically exercise paralyzed and weak muscles to prevent or reverse disuse atrophy. This modular, wireless approach to interfacing with the peripheral nervous system should facilitate the development of progressively more complex systems required to address a growing range of clinical applications, leading ultimately to synthesizing complete voluntary functions such as reach and grasp.
Collapse
Affiliation(s)
- G E Loeb
- A.E. Mann Institute for Biomedical Engineering, University of Southern California, 1042 West 36th Place, Room B-12, Los Angeles, CA 90089-1112, USA.
| | | | | | | |
Collapse
|
43
|
Arabi K, Sawan MA. Electronic design of a multichannel programmable implant for neuromuscular electrical stimulation. IEEE TRANSACTIONS ON REHABILITATION ENGINEERING : A PUBLICATION OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY 1999; 7:204-14. [PMID: 10391591 DOI: 10.1109/86.769411] [Citation(s) in RCA: 56] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
An advanced stimulator for neuromuscular stimulation of spinal cord injured patients has been developed. The stimulator is externally controlled and powered by a single encoded radio frequency carrier and has four independently controlled bipolar stimulation channels. It offers a wide range of reprogrammability and flexibility, and can be used in many neuromuscular electrical stimulation applications. The implant system is adaptable to patient's needs and to future developments in stimulation algorithms by reprogramming the stimulator. The stimulator is capable of generating a wide range of stimulation waveforms and stimulation patterns and therefore is very suitable for selective nerve stimulation techniques. The reliability of the implant has been increased by using a forward error detection and correction communication protocol and by designing the chip for structural testability based on scan test approach. Implemented testability scheme makes it possible to verify the complete functionality of the implant before and after implantation. The stimulators architecture is designed to be modular and therefore its different blocks can be reused as standard building blocks in the design and implementation of other neuromuscular prostheses. Design for low-power techniques have also been employed to reduce power consumption of the electronic circuitry.
Collapse
Affiliation(s)
- K Arabi
- Department of Electrical Engineering, Ecole de Technologie Supérieure, Montreal, P.Q., Canada
| | | |
Collapse
|
44
|
Cameron T, Richmond FJ, Loeb GE. Effects of regional stimulation using a miniature stimulator implanted in feline posterior biceps femoris. IEEE Trans Biomed Eng 1998; 45:1036-43. [PMID: 9691578 DOI: 10.1109/10.704872] [Citation(s) in RCA: 38] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
The effects of placement of a miniature implantable stimulator on motor unit recruitment were examined in the posterior head of cat biceps femoris. The implantable stimulator (13-mm long x 2-mm diameter) was injected either proximally near the main nerve branch, or distally near the muscle insertion, through a 12-gauge hypodermic needle. Glycogen-depletion methods were used to map the distribution of fibers activated by electrical stimulation. Muscle fibers were found to be depleted at most or all proximodistal levels of the muscle, but the density of depleted fibers varied transversely according to the stimulus strength and proximity of the device to the nerve-entry site. Thus, muscle cross sections often had a "patchy" appearance produced because different proportions of depleted fibers intermingled with undepleted fibers in different parts of the cross section. In other preparations, the force of muscle contraction was measured when stimuli of varying strengths were delivered by the stimulator positioned at the same proximal or distal sites within the muscle. Devices placed close to the nerve-entry site produced the greatest forces. Those placed more distally produced less force. As stimulus current and/or pulse width increased, muscle force increased, often in steps, until a maximum was reached, which was usually limited by the compliance voltage of the device to less than the force produced by whole nerve stimulation.
Collapse
Affiliation(s)
- T Cameron
- Advanced Neuromodulation Systems, Inc., Allen, Texas 75002, USA.
| | | | | |
Collapse
|
45
|
Cameron T, Liinamaa TL, Loeb GE, Richmond FJ. Long-term biocompatibility of a miniature stimulator implanted in feline hind limb muscles. IEEE Trans Biomed Eng 1998; 45:1024-35. [PMID: 9691577 DOI: 10.1109/10.704871] [Citation(s) in RCA: 29] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/09/2022]
Abstract
Chronic foreign-body responses and muscular changes were examined following the implantation of active miniature stimulators into the hind limb muscles of cats for periods of up to three months. The radio-frequency (RF)-powered stimulators were injected into muscles through a 12-gauge hypodermic needle. The tissue responses around the active stimulators were compared histologically to those provoked by passive devices, broken glass, silicone tubing, polyester suture material coated with polybutylate, and two of the internal components of the stimulator (ferrite, integrated circuit chip). Active and passive stimulators produced similar, benign foreign-body reactions that resulted in an essentially identical fibrous capsule over time. The responses were similar to those produced by the internal components and the suture material, and were more modest than those produced by the broken glass. The capsule did not appear to interfere with the functionality of active devices because thresholds measured during the post-implantation survival period did not change significantly over time. Unexpectedly, the severity of the reaction differed significantly amongst the various target muscles. Medial gastrocnemius exhibited the most severe response, whereas tibialis anterior had the least reaction.
Collapse
Affiliation(s)
- T Cameron
- Advanced Neuromodulation Systems, Inc., Allen, Texas 75002, USA.
| | | | | | | |
Collapse
|
46
|
Shah MR, Phillips RP, Normann RA. A study of printed spiral coils for neuroprosthetic transcranial telemetry applications. IEEE Trans Biomed Eng 1998; 45:867-76. [PMID: 9644895 DOI: 10.1109/10.686794] [Citation(s) in RCA: 24] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/08/2022]
Abstract
We have explored the use of printed spiral coils (PSC's) for neuroprosthetic transcranial telemetry applications. We fabricated two-dimensional PSC's on a thin (25 microns) polyimide substrate using copper (35 microns) as a conducting material. All the coils had a fixed inner diameter of 1.0 cm. We fabricated two sets of coils. One set of coils consisted of 2- to 5-turn circular and square spiral coils and had different trace widths (W), different spacings (S) between adjacent traces, and different outer diameters. The other set of coils consisted of 5-turn circular spiral coils and had fixed inner and outer diameters but different W to S ratios. We measured loss resistances (Rs and Rp) and quality factors (Q) of these coils at different resonating frequencies in the range of 5-40 MHz. Over this frequency range, we observed that for fixed inner and outer diameters, the coil with the largest W achieved the lowest Rs and the highest Rp and Q. These electrical properties and the fact that these coils can conform to the complex convoluted cortical surface suggest that a PSC [15] can provide a viable alternative to a conventional wire-wound coil for neuroprosthetic transcranial telemetry applications.
Collapse
Affiliation(s)
- M R Shah
- Department of Bioengineering, University of Utah, Salt Lake City 84112, USA
| | | | | |
Collapse
|
47
|
Ziaie B, Nardin MD, Coghlan AR, Najafi K. A single-channel implantable microstimulator for functional neuromuscular stimulation. IEEE Trans Biomed Eng 1997; 44:909-20. [PMID: 9311160 DOI: 10.1109/10.634643] [Citation(s) in RCA: 97] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/05/2023]
Abstract
This paper describes a single-channel implantable microstimulator for functional neuromuscular stimulation. This device measures 2 x 2 x 10 mm3 and can be inserted into paralyzed muscle groups by expulsion from a hypodermic needle. Power and data to the device are supplied from outside by RF telemetry using an amplitude-modulated 2-MHz RF carrier generated using a high-efficiency class-E transmitter. The transmitted signal carries a 5-b address which selects one of the 32 possible microstimulators. The selected device then delivers up to 2 microC of charge store in a tantalum chip capacitor for up to 200 microseconds (10 mA) into loads of < 800 omega through a high-current thin-film iridium-oxide (IrOx) electrode (approximately 0.3 mm2 in area). A bi-CMOS receiver circuitry is used to: generate two regulated voltage supplies (4.5 and 9 V), recover a 2-MHz clock from the carrier, demodulate the address code, and activate the output current delivery circuitry upon the reception of an external command. The overall power dissipation of the receiver circuitry is 45-55 mW. The implant is hermetically packaged using a custom-made glass capsule.
Collapse
Affiliation(s)
- B Ziaie
- Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor 48109-2122 USA.
| | | | | | | |
Collapse
|
48
|
Cameron T, Loeb GE, Peck RA, Schulman JH, Strojnik P, Troyk PR. Micromodular implants to provide electrical stimulation of paralyzed muscles and limbs. IEEE Trans Biomed Eng 1997; 44:781-90. [PMID: 9282470 DOI: 10.1109/10.623047] [Citation(s) in RCA: 102] [Impact Index Per Article: 3.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/05/2023]
Abstract
We describe the design, fabrication, and output capabilities of a microminiature electrical stimulator that can be injected in or near nerves and muscles. Each single-channel microstimulator consists of a cylindrical glass capsule with hermetically sealed electrodes in either end (2-mm diameter x 13-mm overall length). Power and digital control data can be transmitted to multiple implants (256 unique addresses) via a 2-MHz RF field created by an external AM oscillator and inductive coil. In vitro testing demonstrated accurate control of output pulsewidth (3-258 microseconds in 1-microseconds steps) and current (0-30 mA in two linear ranges of 16 steps each, up to 8.5 V available compliance voltage). Microstimulators were used successfully for chronic stimulation in hindlimb muscles of cats. Design and fabrication issues affecting yield and reliability of the packaging and electronics are discussed.
Collapse
Affiliation(s)
- T Cameron
- Queen's University, Biomedical Engineering Unit, Kingston, Ont. Canada.
| | | | | | | | | | | |
Collapse
|
49
|
Prochazka A, Gauthier M, Wieler M, Kenwell Z. The bionic glove: an electrical stimulator garment that provides controlled grasp and hand opening in quadriplegia. Arch Phys Med Rehabil 1997; 78:608-14. [PMID: 9196468 DOI: 10.1016/s0003-9993(97)90426-3] [Citation(s) in RCA: 172] [Impact Index Per Article: 6.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/04/2023]
Abstract
OBJECTIVE This report describes the operation of the Bionic Glove, a new functional electrical stimulation (FES) device designed to improve the function of the paralyzed hand after spinal cord injury (SCI) or stroke. DESIGN Signals from a sensor in the glove detecting voluntary wrist movement are used to control FES of muscles either to produce hand-grasp or to open the hand. When the glove is donned, conductive areas on its inside surface automatically make contact with self-adhesive electrodes on the skin. SETTING AND PATIENTS This report concerns nine people with SCI who have used the device in their daily lives for up to a year or more. Measurements were made at clinics in Edmonton, Miami, and Chicago as part of a multicenter clinical trial. OUTCOME MEASURES AND RESULTS The mean peak force of tenodesis grasp in the nine subjects increased from 2.6N (passive) to 11.3N (glove active). Active force was significantly greater than passive grasp force even when muscles were fatigued after repetitive grasp-release cycles. Most manual tasks improved significantly with the use of the glove, as judged by the number of tasks completed in a minute or the subjects' qualitative ratings of task difficulty. CONCLUSION The Bionic Glove can provide significant improvement of hand function in people with C6-C7 SCI.
Collapse
Affiliation(s)
- A Prochazka
- Division of Neuroscience, University of Alberta, Edmonton, Canada
| | | | | | | |
Collapse
|
50
|
Zealear DL, Rainey CL, Herzon GD, Netterville JL, Ossoff RH. Electrical pacing of the paralyzed human larynx. Ann Otol Rhinol Laryngol 1996; 105:689-93. [PMID: 8800054 DOI: 10.1177/000348949610500904] [Citation(s) in RCA: 59] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/02/2023]
Abstract
This study represents the first attempt to electrically pace the paralyzed human larynx. The goal was to determine if electrical stimulation of the posterior cricoarytenoid muscle could produce functional abduction of the vocal fold in pace with inspiration. An external apparatus was used to sense inspiration and reanimate the unilaterally paralyzed larynx of a thyroplasty patient. Stimuli were delivered through a needle electrode to locate and pace the abductor muscle. The magnitude of electrically induced abduction was comparable to spontaneous movement on the normal side. The abduction was appropriately timed with inspiration: this finding demonstrated that this simple pacing system could effectively modulate stimulation with patient respiration.
Collapse
Affiliation(s)
- D L Zealear
- Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University, Nashville, Tennessee
| | | | | | | | | |
Collapse
|